Partial volume effect in SPECT & PET imaging and impact on radionuclide dosimetry estimates

OBJECTIVE(S): The spatial resolution of emission tomographic imaging systems can lead to a significant underestimation in the apparent radioactivity concentration in objects of size comparable to the resolution volume of the system. The aim of this study was to investigate the impact of the partial...

Descripción completa

Detalles Bibliográficos
Autores principales: Marquis, H, Willowson, KP, Bailey, DL
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Mashhad University of Medical Sciences 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9803618/
https://www.ncbi.nlm.nih.gov/pubmed/36619190
http://dx.doi.org/10.22038/AOJNMB.2022.63827.1448
_version_ 1784861924849090560
author Marquis, H
Willowson, KP
Bailey, DL
author_facet Marquis, H
Willowson, KP
Bailey, DL
author_sort Marquis, H
collection PubMed
description OBJECTIVE(S): The spatial resolution of emission tomographic imaging systems can lead to a significant underestimation in the apparent radioactivity concentration in objects of size comparable to the resolution volume of the system. The aim of this study was to investigate the impact of the partial volume effect (PVE) on clinical imaging in PET and SPECT with current state-of-the-art instrumentation and the implications that this has for radionuclide dosimetry estimates. METHODS: Using the IEC Image Quality Phantom we have measured the underestimation in observed uptake in objects of various sizes for both PET and SPECT imaging conditions. Both single pixel measures (i.e., SUV(max)) and region of interest mean values were examined over a range of object sizes. We have further examined the impact of the PVE on dosimetry estimates in OLINDA in (177)Lu SPECT imaging based on a subject with multiple somatostatin receptor positive paragangliomas in the head and neck. RESULTS: In PET, single pixel estimates of uptake are affected for objects less than approximately 18 mm in minor axis with existing systems. In SPECT imaging with medium energy collimators (e.g., for (177)Lu imaging), however, the underestimates are far greater, where single pixel estimates in objects less than 2-3×the resolution volume are significantly impacted. In SPECT, region of interest mean values are underestimated in objects less than 10 cm in diameter. In the clinical case example, the dosimetry measured with SPECT ranged from more than 60% underestimate in the largest lesion (28×22 mm in maximal cross-section; 10.2 cc volume) to >99% underestimate in the smallest lesion (4×5 mm; 0.06 cc). CONCLUSION: The partial volume effect remains a significant factor when estimating radionuclide uptake in vivo, especially in small volumes. Accurate estimates of absorbed dose from radionuclide therapy will be particularly challenging until robust solutions to correct for the PVE are found.
format Online
Article
Text
id pubmed-9803618
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Mashhad University of Medical Sciences
record_format MEDLINE/PubMed
spelling pubmed-98036182023-01-07 Partial volume effect in SPECT & PET imaging and impact on radionuclide dosimetry estimates Marquis, H Willowson, KP Bailey, DL Asia Ocean J Nucl Med Biol Original Article OBJECTIVE(S): The spatial resolution of emission tomographic imaging systems can lead to a significant underestimation in the apparent radioactivity concentration in objects of size comparable to the resolution volume of the system. The aim of this study was to investigate the impact of the partial volume effect (PVE) on clinical imaging in PET and SPECT with current state-of-the-art instrumentation and the implications that this has for radionuclide dosimetry estimates. METHODS: Using the IEC Image Quality Phantom we have measured the underestimation in observed uptake in objects of various sizes for both PET and SPECT imaging conditions. Both single pixel measures (i.e., SUV(max)) and region of interest mean values were examined over a range of object sizes. We have further examined the impact of the PVE on dosimetry estimates in OLINDA in (177)Lu SPECT imaging based on a subject with multiple somatostatin receptor positive paragangliomas in the head and neck. RESULTS: In PET, single pixel estimates of uptake are affected for objects less than approximately 18 mm in minor axis with existing systems. In SPECT imaging with medium energy collimators (e.g., for (177)Lu imaging), however, the underestimates are far greater, where single pixel estimates in objects less than 2-3×the resolution volume are significantly impacted. In SPECT, region of interest mean values are underestimated in objects less than 10 cm in diameter. In the clinical case example, the dosimetry measured with SPECT ranged from more than 60% underestimate in the largest lesion (28×22 mm in maximal cross-section; 10.2 cc volume) to >99% underestimate in the smallest lesion (4×5 mm; 0.06 cc). CONCLUSION: The partial volume effect remains a significant factor when estimating radionuclide uptake in vivo, especially in small volumes. Accurate estimates of absorbed dose from radionuclide therapy will be particularly challenging until robust solutions to correct for the PVE are found. Mashhad University of Medical Sciences 2023 /pmc/articles/PMC9803618/ /pubmed/36619190 http://dx.doi.org/10.22038/AOJNMB.2022.63827.1448 Text en https://creativecommons.org/licenses/by/3.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution License, (http://creativecommons.org/licenses/by/3.0/ (https://creativecommons.org/licenses/by/3.0/) ) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Marquis, H
Willowson, KP
Bailey, DL
Partial volume effect in SPECT & PET imaging and impact on radionuclide dosimetry estimates
title Partial volume effect in SPECT & PET imaging and impact on radionuclide dosimetry estimates
title_full Partial volume effect in SPECT & PET imaging and impact on radionuclide dosimetry estimates
title_fullStr Partial volume effect in SPECT & PET imaging and impact on radionuclide dosimetry estimates
title_full_unstemmed Partial volume effect in SPECT & PET imaging and impact on radionuclide dosimetry estimates
title_short Partial volume effect in SPECT & PET imaging and impact on radionuclide dosimetry estimates
title_sort partial volume effect in spect & pet imaging and impact on radionuclide dosimetry estimates
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9803618/
https://www.ncbi.nlm.nih.gov/pubmed/36619190
http://dx.doi.org/10.22038/AOJNMB.2022.63827.1448
work_keys_str_mv AT marquish partialvolumeeffectinspectpetimagingandimpactonradionuclidedosimetryestimates
AT willowsonkp partialvolumeeffectinspectpetimagingandimpactonradionuclidedosimetryestimates
AT baileydl partialvolumeeffectinspectpetimagingandimpactonradionuclidedosimetryestimates