Calibration and evaluation of the relative biological effectiveness for carbon-ion radiotherapy in a new relative to a clinically applied treatment planning system

BACKGROUND: The study objective was to validate the relative biological effectiveness (RBE) in RayStation for carbon-ion radiotherapy (CIRT) using the Syngo treatment planning system as reference. METHODS: Local effect model I was established in RayStation (Ray-LEM) with the same parameters as in LE...

Descripción completa

Detalles Bibliográficos
Autores principales: Wang, Weiwei, Huang, Zhijie, Sun, Wei, Wang, Xufei, Zhao, Jingfang, Shen, Hao
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9805684/
https://www.ncbi.nlm.nih.gov/pubmed/36587224
http://dx.doi.org/10.1186/s13014-022-02181-5
Descripción
Sumario:BACKGROUND: The study objective was to validate the relative biological effectiveness (RBE) in RayStation for carbon-ion radiotherapy (CIRT) using the Syngo treatment planning system as reference. METHODS: Local effect model I was established in RayStation (Ray-LEM) with the same parameters as in LEM I in Syngo (Syngo-LEM). Three cube plans covering most of the tumors treated at our center were generated with Syngo-LEM. Ray-LEM re-calculated the Syngo plans and compared the RBEs to the Syngo counterparts. The results showed that RayStation RBE was smaller than Syngo RBE. To ensure that Ray-LEM reproduced Syngo RBE, the observed deviations were used to scale the maximum RBE (RBEmax) in Ray-LEM. After this calibration, we further compared the RayStation RBE to Syngo RBE using additional plans in both homogeneous phantoms and patients, to ensure that the calibrated Ray-LEM reproduced Syngo RBE even with more complex planning features. RESULTS: The calibration increased the RBEmax by 2.3% to raise the Ray-LEM RBE. The target mean RBE deviations in the phantom evaluation plans were median: 0.0 (minimum: − 1.1 to maximum: 0.7) %, and the target mean RBE deviations of the clinical target volumes of 16 patient cases were − 0.4 (− 1.5 to 0.2) %. CONCLUSIONS: The residual RBE difference between RayStation and Syngo was found to be ≤ 1.0%. Thus, we can propose to use RayStation for clinical CIRT treatment planning. However, the potential differences due to the absorbed beam model warrants further exploration. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s13014-022-02181-5.