Cargando…

Performance Evaluation of STANDARD Q COVID/FLU Ag Combo for Detection of SARS-CoV-2 and Influenza A/B

We evaluated the performance of the STANDARD Q COVID/FLU Ag Combo test (Q Ag combo test) for the detection of SARS-CoV-2, influenza A, and influenza B using a single point-of-care device compared with real-time PCR. A total of 408 individuals, 55 positives with SARS-CoV-2, 90 with influenza A, 68 wi...

Descripción completa

Detalles Bibliográficos
Autores principales: Widyasari, Kristin, Kim, Soomin, Kim, Sunjoo, Lim, Chae Seung
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9818676/
https://www.ncbi.nlm.nih.gov/pubmed/36611324
http://dx.doi.org/10.3390/diagnostics13010032
_version_ 1784865044294533120
author Widyasari, Kristin
Kim, Soomin
Kim, Sunjoo
Lim, Chae Seung
author_facet Widyasari, Kristin
Kim, Soomin
Kim, Sunjoo
Lim, Chae Seung
author_sort Widyasari, Kristin
collection PubMed
description We evaluated the performance of the STANDARD Q COVID/FLU Ag Combo test (Q Ag combo test) for the detection of SARS-CoV-2, influenza A, and influenza B using a single point-of-care device compared with real-time PCR. A total of 408 individuals, 55 positives with SARS-CoV-2, 90 with influenza A, 68 with influenza B, and 195 negatives for all viruses, participated. The Q Ag combo test demonstrated a high level of sensitivity of 92.73% and a specificity of 99.49% for the detection of SARS-CoV-2. When the number of days from symptom onset (DSO) was restricted to 0 < DSO ≤ 6, the sensitivity of the Q Ag combo test to detect SARS-CoV-2 was 100%, and when the Ct value of RdRp was ≤20, the sensitivity to detect SARS-CoV-2 was 93.10%. The Q Ag combo test results also demonstrated a sensitivity of 92.22% and a specificity of 100% for influenza A, a sensitivity of 91.18%, and a specificity of 99.49% for influenza B. The agreement analysis of the Q Ag combo test with the RT-PCR results demonstrated excellent outcomes, making it useful and efficient for the detection of SARS-CoV-2, influenza A, and influenza B.
format Online
Article
Text
id pubmed-9818676
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-98186762023-01-07 Performance Evaluation of STANDARD Q COVID/FLU Ag Combo for Detection of SARS-CoV-2 and Influenza A/B Widyasari, Kristin Kim, Soomin Kim, Sunjoo Lim, Chae Seung Diagnostics (Basel) Article We evaluated the performance of the STANDARD Q COVID/FLU Ag Combo test (Q Ag combo test) for the detection of SARS-CoV-2, influenza A, and influenza B using a single point-of-care device compared with real-time PCR. A total of 408 individuals, 55 positives with SARS-CoV-2, 90 with influenza A, 68 with influenza B, and 195 negatives for all viruses, participated. The Q Ag combo test demonstrated a high level of sensitivity of 92.73% and a specificity of 99.49% for the detection of SARS-CoV-2. When the number of days from symptom onset (DSO) was restricted to 0 < DSO ≤ 6, the sensitivity of the Q Ag combo test to detect SARS-CoV-2 was 100%, and when the Ct value of RdRp was ≤20, the sensitivity to detect SARS-CoV-2 was 93.10%. The Q Ag combo test results also demonstrated a sensitivity of 92.22% and a specificity of 100% for influenza A, a sensitivity of 91.18%, and a specificity of 99.49% for influenza B. The agreement analysis of the Q Ag combo test with the RT-PCR results demonstrated excellent outcomes, making it useful and efficient for the detection of SARS-CoV-2, influenza A, and influenza B. MDPI 2022-12-22 /pmc/articles/PMC9818676/ /pubmed/36611324 http://dx.doi.org/10.3390/diagnostics13010032 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Widyasari, Kristin
Kim, Soomin
Kim, Sunjoo
Lim, Chae Seung
Performance Evaluation of STANDARD Q COVID/FLU Ag Combo for Detection of SARS-CoV-2 and Influenza A/B
title Performance Evaluation of STANDARD Q COVID/FLU Ag Combo for Detection of SARS-CoV-2 and Influenza A/B
title_full Performance Evaluation of STANDARD Q COVID/FLU Ag Combo for Detection of SARS-CoV-2 and Influenza A/B
title_fullStr Performance Evaluation of STANDARD Q COVID/FLU Ag Combo for Detection of SARS-CoV-2 and Influenza A/B
title_full_unstemmed Performance Evaluation of STANDARD Q COVID/FLU Ag Combo for Detection of SARS-CoV-2 and Influenza A/B
title_short Performance Evaluation of STANDARD Q COVID/FLU Ag Combo for Detection of SARS-CoV-2 and Influenza A/B
title_sort performance evaluation of standard q covid/flu ag combo for detection of sars-cov-2 and influenza a/b
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9818676/
https://www.ncbi.nlm.nih.gov/pubmed/36611324
http://dx.doi.org/10.3390/diagnostics13010032
work_keys_str_mv AT widyasarikristin performanceevaluationofstandardqcovidfluagcombofordetectionofsarscov2andinfluenzaab
AT kimsoomin performanceevaluationofstandardqcovidfluagcombofordetectionofsarscov2andinfluenzaab
AT kimsunjoo performanceevaluationofstandardqcovidfluagcombofordetectionofsarscov2andinfluenzaab
AT limchaeseung performanceevaluationofstandardqcovidfluagcombofordetectionofsarscov2andinfluenzaab