Evaluation of the Analytical Performances of the Biolabo SOLEA 100 Optical Coagulometer and Comparison with the Stago STA-R MAX Analyser in the Determination of PT, APTT, and Fibrinogen
Introduction. The Biolabo Solea 100 is a fully automated coagulation analyser using an optical system to detect coagulation designed to meet the needs of small- and medium-sized laboratories. This study aimed to evaluate the analytical performance in terms of bias, precision, and interference of the...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9818767/ https://www.ncbi.nlm.nih.gov/pubmed/36611377 http://dx.doi.org/10.3390/diagnostics13010085 |
_version_ | 1784865067139858432 |
---|---|
author | Bellio, Pierangelo De Angelis, Simonetta Piccirilli, Alessandra Di Michele, Giulio Barnabei, Remo Amicosante, Gianfranco Perilli, Mariagrazia Celenza, Giuseppe |
author_facet | Bellio, Pierangelo De Angelis, Simonetta Piccirilli, Alessandra Di Michele, Giulio Barnabei, Remo Amicosante, Gianfranco Perilli, Mariagrazia Celenza, Giuseppe |
author_sort | Bellio, Pierangelo |
collection | PubMed |
description | Introduction. The Biolabo Solea 100 is a fully automated coagulation analyser using an optical system to detect coagulation designed to meet the needs of small- and medium-sized laboratories. This study aimed to evaluate the analytical performance in terms of bias, precision, and interference of the Biolabo Solea 100 coagulometer under routine laboratory conditions. In addition, a comparison was made with Stago STA-R MAX. Materials and Methods. Imprecision and bias were evaluated for activated partial thromboplastin time (APTT), fibrinogen (FIB), and prothrombin time (PT) at the medical decision levels. The results of 200, 181, and 206 plasma samples for APTT, FIB, and PT, respectively, were compared with those obtained by Stago STA-R MAX. In addition, the interference level of bilirubin, haemoglobin, triglycerides, and fractionated heparin was evaluated. Results. Repeatability, intermediate imprecision, bias, and total error are overall below the defined limits of acceptability. Of interest is the high degree of agreement between Solea 100 and STA-R MAX with respect to PT (s), which fits perfectly with the theoretical line of identity (y = 0 + 1.00x). No interferences were found within the limits stated by the manufacturer, with some exceptions for APTT with heparin and APTT and PT for higher bilirubin concentrations. Conclusions. In conclusion, the performance of the Solea 100 optical analyser is satisfactory and adequate for the determination of routine coagulation tests. Moreover, they are perfectly comparable to mechanical systems, such as STA-R MAX and other upper-level analysers, even considering the low interference levels under routine conditions. |
format | Online Article Text |
id | pubmed-9818767 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-98187672023-01-07 Evaluation of the Analytical Performances of the Biolabo SOLEA 100 Optical Coagulometer and Comparison with the Stago STA-R MAX Analyser in the Determination of PT, APTT, and Fibrinogen Bellio, Pierangelo De Angelis, Simonetta Piccirilli, Alessandra Di Michele, Giulio Barnabei, Remo Amicosante, Gianfranco Perilli, Mariagrazia Celenza, Giuseppe Diagnostics (Basel) Article Introduction. The Biolabo Solea 100 is a fully automated coagulation analyser using an optical system to detect coagulation designed to meet the needs of small- and medium-sized laboratories. This study aimed to evaluate the analytical performance in terms of bias, precision, and interference of the Biolabo Solea 100 coagulometer under routine laboratory conditions. In addition, a comparison was made with Stago STA-R MAX. Materials and Methods. Imprecision and bias were evaluated for activated partial thromboplastin time (APTT), fibrinogen (FIB), and prothrombin time (PT) at the medical decision levels. The results of 200, 181, and 206 plasma samples for APTT, FIB, and PT, respectively, were compared with those obtained by Stago STA-R MAX. In addition, the interference level of bilirubin, haemoglobin, triglycerides, and fractionated heparin was evaluated. Results. Repeatability, intermediate imprecision, bias, and total error are overall below the defined limits of acceptability. Of interest is the high degree of agreement between Solea 100 and STA-R MAX with respect to PT (s), which fits perfectly with the theoretical line of identity (y = 0 + 1.00x). No interferences were found within the limits stated by the manufacturer, with some exceptions for APTT with heparin and APTT and PT for higher bilirubin concentrations. Conclusions. In conclusion, the performance of the Solea 100 optical analyser is satisfactory and adequate for the determination of routine coagulation tests. Moreover, they are perfectly comparable to mechanical systems, such as STA-R MAX and other upper-level analysers, even considering the low interference levels under routine conditions. MDPI 2022-12-28 /pmc/articles/PMC9818767/ /pubmed/36611377 http://dx.doi.org/10.3390/diagnostics13010085 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Bellio, Pierangelo De Angelis, Simonetta Piccirilli, Alessandra Di Michele, Giulio Barnabei, Remo Amicosante, Gianfranco Perilli, Mariagrazia Celenza, Giuseppe Evaluation of the Analytical Performances of the Biolabo SOLEA 100 Optical Coagulometer and Comparison with the Stago STA-R MAX Analyser in the Determination of PT, APTT, and Fibrinogen |
title | Evaluation of the Analytical Performances of the Biolabo SOLEA 100 Optical Coagulometer and Comparison with the Stago STA-R MAX Analyser in the Determination of PT, APTT, and Fibrinogen |
title_full | Evaluation of the Analytical Performances of the Biolabo SOLEA 100 Optical Coagulometer and Comparison with the Stago STA-R MAX Analyser in the Determination of PT, APTT, and Fibrinogen |
title_fullStr | Evaluation of the Analytical Performances of the Biolabo SOLEA 100 Optical Coagulometer and Comparison with the Stago STA-R MAX Analyser in the Determination of PT, APTT, and Fibrinogen |
title_full_unstemmed | Evaluation of the Analytical Performances of the Biolabo SOLEA 100 Optical Coagulometer and Comparison with the Stago STA-R MAX Analyser in the Determination of PT, APTT, and Fibrinogen |
title_short | Evaluation of the Analytical Performances of the Biolabo SOLEA 100 Optical Coagulometer and Comparison with the Stago STA-R MAX Analyser in the Determination of PT, APTT, and Fibrinogen |
title_sort | evaluation of the analytical performances of the biolabo solea 100 optical coagulometer and comparison with the stago sta-r max analyser in the determination of pt, aptt, and fibrinogen |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9818767/ https://www.ncbi.nlm.nih.gov/pubmed/36611377 http://dx.doi.org/10.3390/diagnostics13010085 |
work_keys_str_mv | AT belliopierangelo evaluationoftheanalyticalperformancesofthebiolabosolea100opticalcoagulometerandcomparisonwiththestagostarmaxanalyserinthedeterminationofptapttandfibrinogen AT deangelissimonetta evaluationoftheanalyticalperformancesofthebiolabosolea100opticalcoagulometerandcomparisonwiththestagostarmaxanalyserinthedeterminationofptapttandfibrinogen AT piccirillialessandra evaluationoftheanalyticalperformancesofthebiolabosolea100opticalcoagulometerandcomparisonwiththestagostarmaxanalyserinthedeterminationofptapttandfibrinogen AT dimichelegiulio evaluationoftheanalyticalperformancesofthebiolabosolea100opticalcoagulometerandcomparisonwiththestagostarmaxanalyserinthedeterminationofptapttandfibrinogen AT barnabeiremo evaluationoftheanalyticalperformancesofthebiolabosolea100opticalcoagulometerandcomparisonwiththestagostarmaxanalyserinthedeterminationofptapttandfibrinogen AT amicosantegianfranco evaluationoftheanalyticalperformancesofthebiolabosolea100opticalcoagulometerandcomparisonwiththestagostarmaxanalyserinthedeterminationofptapttandfibrinogen AT perillimariagrazia evaluationoftheanalyticalperformancesofthebiolabosolea100opticalcoagulometerandcomparisonwiththestagostarmaxanalyserinthedeterminationofptapttandfibrinogen AT celenzagiuseppe evaluationoftheanalyticalperformancesofthebiolabosolea100opticalcoagulometerandcomparisonwiththestagostarmaxanalyserinthedeterminationofptapttandfibrinogen |