Perioperative Validation of the TensorTip™ MTX Device for Noninvasive Arterial Pressure Measurement: A Method Comparison Study

BACKGROUND: The noninvasive TensorTip™ MTX measures blood pressure by interpreting blood diffusion color of the finger skin. In addition to blood pressure, the device is able to measure various vital signs: heart rate, oxygen saturation, stroke volume, and cardiac output. Studies about accuracy and...

Descripción completa

Detalles Bibliográficos
Autores principales: Servaas, Sjoerd, van Eijk, Lucas T., de Vreede, Silke, Malagon, Ignacio, Slagt, Cornelis
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elmer Press 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9822676/
https://www.ncbi.nlm.nih.gov/pubmed/36660063
http://dx.doi.org/10.14740/cr1438
_version_ 1784866001602478080
author Servaas, Sjoerd
van Eijk, Lucas T.
de Vreede, Silke
Malagon, Ignacio
Slagt, Cornelis
author_facet Servaas, Sjoerd
van Eijk, Lucas T.
de Vreede, Silke
Malagon, Ignacio
Slagt, Cornelis
author_sort Servaas, Sjoerd
collection PubMed
description BACKGROUND: The noninvasive TensorTip™ MTX measures blood pressure by interpreting blood diffusion color of the finger skin. In addition to blood pressure, the device is able to measure various vital signs: heart rate, oxygen saturation, stroke volume, and cardiac output. Studies about accuracy and precision thus far available have only been conducted by the manufacturer. The aim of our study was to investigate the accuracy and precision of the TensorTip MTX in comparison to invasive radial artery blood pressure values. METHODS: Forty-one patients scheduled for elective surgery were enrolled in this study. Placement of the arterial catheter had to be part of the standard of care. Once hemodynamic stable conditions were achieved, blood pressure was measured. Three measurements with the TensorTip MTX were averaged and compared with one invasive blood pressure measurement using Bland-Altman plot and error grid analysis. RESULTS: Systolic, diastolic, and mean blood pressure had a bias of respectively 6.2, -6.9 and 4.4 mm Hg. Corresponding standard deviation were respectively 30.1, 17.0 and 22.2. Calculated percentage errors were 47.6%, 52.9% and 52.3%. Proportional bias was present in all Bland-Altman analyses. Error grid analysis showed 61.0% of systolic blood pressure measurements, and 46.3% of mean blood pressure measurements were in the clinical acceptable zone. CONCLUSIONS: The TensorTip MTX was not able to reliably measure blood pressure compared to blood pressure obtained with an arterial catheter and therefore, the measurement performance is not clinically acceptable. Moreover, a high malfunction rate makes the device unsuitable for use in perioperative period.
format Online
Article
Text
id pubmed-9822676
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Elmer Press
record_format MEDLINE/PubMed
spelling pubmed-98226762023-01-18 Perioperative Validation of the TensorTip™ MTX Device for Noninvasive Arterial Pressure Measurement: A Method Comparison Study Servaas, Sjoerd van Eijk, Lucas T. de Vreede, Silke Malagon, Ignacio Slagt, Cornelis Cardiol Res Original Article BACKGROUND: The noninvasive TensorTip™ MTX measures blood pressure by interpreting blood diffusion color of the finger skin. In addition to blood pressure, the device is able to measure various vital signs: heart rate, oxygen saturation, stroke volume, and cardiac output. Studies about accuracy and precision thus far available have only been conducted by the manufacturer. The aim of our study was to investigate the accuracy and precision of the TensorTip MTX in comparison to invasive radial artery blood pressure values. METHODS: Forty-one patients scheduled for elective surgery were enrolled in this study. Placement of the arterial catheter had to be part of the standard of care. Once hemodynamic stable conditions were achieved, blood pressure was measured. Three measurements with the TensorTip MTX were averaged and compared with one invasive blood pressure measurement using Bland-Altman plot and error grid analysis. RESULTS: Systolic, diastolic, and mean blood pressure had a bias of respectively 6.2, -6.9 and 4.4 mm Hg. Corresponding standard deviation were respectively 30.1, 17.0 and 22.2. Calculated percentage errors were 47.6%, 52.9% and 52.3%. Proportional bias was present in all Bland-Altman analyses. Error grid analysis showed 61.0% of systolic blood pressure measurements, and 46.3% of mean blood pressure measurements were in the clinical acceptable zone. CONCLUSIONS: The TensorTip MTX was not able to reliably measure blood pressure compared to blood pressure obtained with an arterial catheter and therefore, the measurement performance is not clinically acceptable. Moreover, a high malfunction rate makes the device unsuitable for use in perioperative period. Elmer Press 2022-12 2022-12-16 /pmc/articles/PMC9822676/ /pubmed/36660063 http://dx.doi.org/10.14740/cr1438 Text en Copyright 2022, Servaas et al. https://creativecommons.org/licenses/by-nc/4.0/This article is distributed under the terms of the Creative Commons Attribution Non-Commercial 4.0 International License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Servaas, Sjoerd
van Eijk, Lucas T.
de Vreede, Silke
Malagon, Ignacio
Slagt, Cornelis
Perioperative Validation of the TensorTip™ MTX Device for Noninvasive Arterial Pressure Measurement: A Method Comparison Study
title Perioperative Validation of the TensorTip™ MTX Device for Noninvasive Arterial Pressure Measurement: A Method Comparison Study
title_full Perioperative Validation of the TensorTip™ MTX Device for Noninvasive Arterial Pressure Measurement: A Method Comparison Study
title_fullStr Perioperative Validation of the TensorTip™ MTX Device for Noninvasive Arterial Pressure Measurement: A Method Comparison Study
title_full_unstemmed Perioperative Validation of the TensorTip™ MTX Device for Noninvasive Arterial Pressure Measurement: A Method Comparison Study
title_short Perioperative Validation of the TensorTip™ MTX Device for Noninvasive Arterial Pressure Measurement: A Method Comparison Study
title_sort perioperative validation of the tensortip™ mtx device for noninvasive arterial pressure measurement: a method comparison study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9822676/
https://www.ncbi.nlm.nih.gov/pubmed/36660063
http://dx.doi.org/10.14740/cr1438
work_keys_str_mv AT servaassjoerd perioperativevalidationofthetensortipmtxdevicefornoninvasivearterialpressuremeasurementamethodcomparisonstudy
AT vaneijklucast perioperativevalidationofthetensortipmtxdevicefornoninvasivearterialpressuremeasurementamethodcomparisonstudy
AT devreedesilke perioperativevalidationofthetensortipmtxdevicefornoninvasivearterialpressuremeasurementamethodcomparisonstudy
AT malagonignacio perioperativevalidationofthetensortipmtxdevicefornoninvasivearterialpressuremeasurementamethodcomparisonstudy
AT slagtcornelis perioperativevalidationofthetensortipmtxdevicefornoninvasivearterialpressuremeasurementamethodcomparisonstudy