Cargando…
Development and Testing of a Compact Autorefractor Based on Double-Pass Imaging
Autorefraction is an objective way to determine the refractive error of the eye, without the need for feedback by the patient or a well-educated practitioner. To make refractive measurements more accessible in the background of the growing prevalence of myopia, a compact autorefractor was built, con...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9823743/ https://www.ncbi.nlm.nih.gov/pubmed/36616960 http://dx.doi.org/10.3390/s23010362 |
Sumario: | Autorefraction is an objective way to determine the refractive error of the eye, without the need for feedback by the patient or a well-educated practitioner. To make refractive measurements more accessible in the background of the growing prevalence of myopia, a compact autorefractor was built, containing only few optical components and relying on double-pass imaging and the physical properties of the point-spread function and digital image processing instead. A method was developed to analyze spherical defocus as well as the defocus and angle of astigmatism. The device was tested using calibrator eye models in a range of ± 15 D spherical defocus and [Formula: see text] D astigmatic defocus. Reliable results could be achieved across the whole measurement range, with only a small increase in deviation toward high values of refractive errors, showing the feasibility of a PSF-based approach for a compact and low-cost solution for objective measurements of refractive error. |
---|