Cargando…
A Wide Dynamic Range Sigma-Delta Modulator for EEG Acquisition Using Randomized DWA and Dynamic-Modulated Scaling-Down Techniques
This paper proposes a wide dynamic range (DR) and high-resolution discrete-time (DT) 2-order 4-bit sigma-delta modulator with a novel dynamic-modulated scaling-down (DM-SD) technology for non-invasive electroencephalogram (EEG) acquisition. The DM-SD technology can expand the input dynamic range and...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9824506/ https://www.ncbi.nlm.nih.gov/pubmed/36616799 http://dx.doi.org/10.3390/s23010201 |
Sumario: | This paper proposes a wide dynamic range (DR) and high-resolution discrete-time (DT) 2-order 4-bit sigma-delta modulator with a novel dynamic-modulated scaling-down (DM-SD) technology for non-invasive electroencephalogram (EEG) acquisition. The DM-SD technology can expand the input dynamic range and suppress large input offsets at the same time. The modulator was designed with 180nm CMOS technology with an area of 0.49 mm [Formula: see text]. We achieve a 118.1 dB SNDR when the input signal is 437.5 Hz and the signal bandwidth is 1500 Hz. Due to the proposed DM-SD technology, the DR is expanded to 126 dB. The power consumption of the whole modulator is 1.6 mW and a 177.8 dB Schreier figure-of-merit (FoMs) is realized. |
---|