Cargando…

Accelerating 3D MTC-BOOST in patients with congenital heart disease using a joint multi-scale variational neural network reconstruction

PURPOSE: Free-breathing Magnetization Transfer Contrast Bright blOOd phase SensiTive (MTC-BOOST) is a prototype balanced-Steady-State Free Precession sequence for 3D whole-heart imaging, that employs the endogenous magnetisation transfer contrast mechanism. This achieves reduction of flow and off-re...

Descripción completa

Detalles Bibliográficos
Autores principales: Fotaki, Anastasia, Fuin, Niccolo, Nordio, Giovanna, Velasco Jimeno, Carlos, Qi, Haikun, Emmanuel, Yaso, Pushparajah, Kuberan, Botnar, René M., Prieto, Claudia
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9826869/
https://www.ncbi.nlm.nih.gov/pubmed/35772584
http://dx.doi.org/10.1016/j.mri.2022.06.012
_version_ 1784866950495600640
author Fotaki, Anastasia
Fuin, Niccolo
Nordio, Giovanna
Velasco Jimeno, Carlos
Qi, Haikun
Emmanuel, Yaso
Pushparajah, Kuberan
Botnar, René M.
Prieto, Claudia
author_facet Fotaki, Anastasia
Fuin, Niccolo
Nordio, Giovanna
Velasco Jimeno, Carlos
Qi, Haikun
Emmanuel, Yaso
Pushparajah, Kuberan
Botnar, René M.
Prieto, Claudia
author_sort Fotaki, Anastasia
collection PubMed
description PURPOSE: Free-breathing Magnetization Transfer Contrast Bright blOOd phase SensiTive (MTC-BOOST) is a prototype balanced-Steady-State Free Precession sequence for 3D whole-heart imaging, that employs the endogenous magnetisation transfer contrast mechanism. This achieves reduction of flow and off-resonance artefacts, that often arise with the clinical T2prepared balanced-Steady-State Free Precession sequence, enabling high quality, contrast-agent free imaging of the thoracic cardiovascular anatomy. Fully-sampled MTC-BOOST acquisition requires long scan times (~10–24 min) and therefore acceleration is needed to permit its clinical incorporation. The aim of this study is to enable and clinically validate the 5-fold accelerated MTC-BOOST acquisition with joint Multi-Scale Variational Neural Network (jMS-VNN) reconstruction. METHODS: Thirty-six patients underwent free-breathing, 3D whole-heart imaging with the MTC-BOOST sequence, which is combined with variable density spiral-like Cartesian sampling and 2D image navigators for translational motion estimation. This sequence acquires two differently weighted bright-blood volumes in an interleaved fashion, which are then joined in a phase sensitive inversion recovery reconstruction to obtain a complementary fully co-registered black-blood volume. Data from eighteen patients were used for training, whereas data from the remaining eighteen patients were used for testing/evaluation. The proposed deep-learning based approach adopts a supervised multi-scale variational neural network for joint reconstruction of the two differently weighted bright-blood volumes acquired with the 5-fold accelerated MTC-BOOST. The two contrast images are stacked as different channels in the network to exploit the shared information. The proposed approach is compared to the fully-sampled MTC-BOOST and 5-fold undersampled MTC-BOOST acquisition with Compressed Sensing (CS) reconstruction in terms of scan/reconstruction time and bright-blood image quality. Comparison against conventional 2-fold undersampled T2-prepared 3D bright-blood whole-heart clinical sequence (T2prep-3DWH) is also included. RESULTS: Acquisition time was 3.0 ± 1.0 min for the 5-fold accelerated MTC-BOOST versus 9.0 ± 1.1 min for the fully-sampled MTC-BOOST and 11.1 ± 2.6 min for the T2prep-3DWH (p < 0.001 and p < 0.001, respectively). Reconstruction time was significantly lower with the jMS-VNN method compared to CS (10 ± 0.5 min vs 20 ± 2 s, p < 0.001). Image quality was higher for the proposed 5-fold undersampled jMS-VNN versus conventional CS, comparable or higher to the corresponding T2prep-3DWH dataset and similar to the fully-sampled MTC-BOOST. CONCLUSION: The proposed 5-fold accelerated jMS-VNN MTC-BOOST framework provides efficient 3D whole-heart bright-blood imaging in fast acquisition and reconstruction time with concomitant reduction of flow and off-resonance artefacts, that are frequently encountered with the clinical sequence. Image quality of the cardiac anatomy and thoracic vasculature is comparable or superior to the clinical scan and 5-fold CS reconstruction in faster reconstruction time, promising potential clinical adoption.
format Online
Article
Text
id pubmed-9826869
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-98268692023-01-10 Accelerating 3D MTC-BOOST in patients with congenital heart disease using a joint multi-scale variational neural network reconstruction Fotaki, Anastasia Fuin, Niccolo Nordio, Giovanna Velasco Jimeno, Carlos Qi, Haikun Emmanuel, Yaso Pushparajah, Kuberan Botnar, René M. Prieto, Claudia Magn Reson Imaging Original Contribution PURPOSE: Free-breathing Magnetization Transfer Contrast Bright blOOd phase SensiTive (MTC-BOOST) is a prototype balanced-Steady-State Free Precession sequence for 3D whole-heart imaging, that employs the endogenous magnetisation transfer contrast mechanism. This achieves reduction of flow and off-resonance artefacts, that often arise with the clinical T2prepared balanced-Steady-State Free Precession sequence, enabling high quality, contrast-agent free imaging of the thoracic cardiovascular anatomy. Fully-sampled MTC-BOOST acquisition requires long scan times (~10–24 min) and therefore acceleration is needed to permit its clinical incorporation. The aim of this study is to enable and clinically validate the 5-fold accelerated MTC-BOOST acquisition with joint Multi-Scale Variational Neural Network (jMS-VNN) reconstruction. METHODS: Thirty-six patients underwent free-breathing, 3D whole-heart imaging with the MTC-BOOST sequence, which is combined with variable density spiral-like Cartesian sampling and 2D image navigators for translational motion estimation. This sequence acquires two differently weighted bright-blood volumes in an interleaved fashion, which are then joined in a phase sensitive inversion recovery reconstruction to obtain a complementary fully co-registered black-blood volume. Data from eighteen patients were used for training, whereas data from the remaining eighteen patients were used for testing/evaluation. The proposed deep-learning based approach adopts a supervised multi-scale variational neural network for joint reconstruction of the two differently weighted bright-blood volumes acquired with the 5-fold accelerated MTC-BOOST. The two contrast images are stacked as different channels in the network to exploit the shared information. The proposed approach is compared to the fully-sampled MTC-BOOST and 5-fold undersampled MTC-BOOST acquisition with Compressed Sensing (CS) reconstruction in terms of scan/reconstruction time and bright-blood image quality. Comparison against conventional 2-fold undersampled T2-prepared 3D bright-blood whole-heart clinical sequence (T2prep-3DWH) is also included. RESULTS: Acquisition time was 3.0 ± 1.0 min for the 5-fold accelerated MTC-BOOST versus 9.0 ± 1.1 min for the fully-sampled MTC-BOOST and 11.1 ± 2.6 min for the T2prep-3DWH (p < 0.001 and p < 0.001, respectively). Reconstruction time was significantly lower with the jMS-VNN method compared to CS (10 ± 0.5 min vs 20 ± 2 s, p < 0.001). Image quality was higher for the proposed 5-fold undersampled jMS-VNN versus conventional CS, comparable or higher to the corresponding T2prep-3DWH dataset and similar to the fully-sampled MTC-BOOST. CONCLUSION: The proposed 5-fold accelerated jMS-VNN MTC-BOOST framework provides efficient 3D whole-heart bright-blood imaging in fast acquisition and reconstruction time with concomitant reduction of flow and off-resonance artefacts, that are frequently encountered with the clinical sequence. Image quality of the cardiac anatomy and thoracic vasculature is comparable or superior to the clinical scan and 5-fold CS reconstruction in faster reconstruction time, promising potential clinical adoption. Elsevier 2022-10 /pmc/articles/PMC9826869/ /pubmed/35772584 http://dx.doi.org/10.1016/j.mri.2022.06.012 Text en © 2022 The Authors https://creativecommons.org/licenses/by/4.0/This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Original Contribution
Fotaki, Anastasia
Fuin, Niccolo
Nordio, Giovanna
Velasco Jimeno, Carlos
Qi, Haikun
Emmanuel, Yaso
Pushparajah, Kuberan
Botnar, René M.
Prieto, Claudia
Accelerating 3D MTC-BOOST in patients with congenital heart disease using a joint multi-scale variational neural network reconstruction
title Accelerating 3D MTC-BOOST in patients with congenital heart disease using a joint multi-scale variational neural network reconstruction
title_full Accelerating 3D MTC-BOOST in patients with congenital heart disease using a joint multi-scale variational neural network reconstruction
title_fullStr Accelerating 3D MTC-BOOST in patients with congenital heart disease using a joint multi-scale variational neural network reconstruction
title_full_unstemmed Accelerating 3D MTC-BOOST in patients with congenital heart disease using a joint multi-scale variational neural network reconstruction
title_short Accelerating 3D MTC-BOOST in patients with congenital heart disease using a joint multi-scale variational neural network reconstruction
title_sort accelerating 3d mtc-boost in patients with congenital heart disease using a joint multi-scale variational neural network reconstruction
topic Original Contribution
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9826869/
https://www.ncbi.nlm.nih.gov/pubmed/35772584
http://dx.doi.org/10.1016/j.mri.2022.06.012
work_keys_str_mv AT fotakianastasia accelerating3dmtcboostinpatientswithcongenitalheartdiseaseusingajointmultiscalevariationalneuralnetworkreconstruction
AT fuinniccolo accelerating3dmtcboostinpatientswithcongenitalheartdiseaseusingajointmultiscalevariationalneuralnetworkreconstruction
AT nordiogiovanna accelerating3dmtcboostinpatientswithcongenitalheartdiseaseusingajointmultiscalevariationalneuralnetworkreconstruction
AT velascojimenocarlos accelerating3dmtcboostinpatientswithcongenitalheartdiseaseusingajointmultiscalevariationalneuralnetworkreconstruction
AT qihaikun accelerating3dmtcboostinpatientswithcongenitalheartdiseaseusingajointmultiscalevariationalneuralnetworkreconstruction
AT emmanuelyaso accelerating3dmtcboostinpatientswithcongenitalheartdiseaseusingajointmultiscalevariationalneuralnetworkreconstruction
AT pushparajahkuberan accelerating3dmtcboostinpatientswithcongenitalheartdiseaseusingajointmultiscalevariationalneuralnetworkreconstruction
AT botnarrenem accelerating3dmtcboostinpatientswithcongenitalheartdiseaseusingajointmultiscalevariationalneuralnetworkreconstruction
AT prietoclaudia accelerating3dmtcboostinpatientswithcongenitalheartdiseaseusingajointmultiscalevariationalneuralnetworkreconstruction