Cargando…

A solution for global hygienic challenges regarding the application of heater-cooler systems in cardiac surgery

71.759 surgical procedures were performed in 2019 with the aid of cardiopulmonary bypass in Germany. To adjust the patient’s body temperature on extracorporeal circulation, the application of a heater-cooler unit (HCU) is mandatory. However, in case of insufficient sanitisation of HCU, life-threaten...

Descripción completa

Detalles Bibliográficos
Autores principales: Foltan, Maik, Nikisch, André, Dembianny, Jan, Miano, Anna-Lena, Heinze, Johannes, Klar, Dennis, Göbölös, Laszlo, Lehle, Karla, Schmid, Christof
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9841820/
https://www.ncbi.nlm.nih.gov/pubmed/34256647
http://dx.doi.org/10.1177/02676591211033483
_version_ 1784869975345856512
author Foltan, Maik
Nikisch, André
Dembianny, Jan
Miano, Anna-Lena
Heinze, Johannes
Klar, Dennis
Göbölös, Laszlo
Lehle, Karla
Schmid, Christof
author_facet Foltan, Maik
Nikisch, André
Dembianny, Jan
Miano, Anna-Lena
Heinze, Johannes
Klar, Dennis
Göbölös, Laszlo
Lehle, Karla
Schmid, Christof
author_sort Foltan, Maik
collection PubMed
description 71.759 surgical procedures were performed in 2019 with the aid of cardiopulmonary bypass in Germany. To adjust the patient’s body temperature on extracorporeal circulation, the application of a heater-cooler unit (HCU) is mandatory. However, in case of insufficient sanitisation of HCU, life-threatening infections can be transmitted by the device to the patients, including Legionella bacteria, Mycobacterium chimaera, Pseudomonas aeruginosa. To avoid disease transmission, as a requirement for safe medical practice established by regulatory authorities, HCUs must be regularly disinfected by hazardous chemicals posing a danger for both handling humans and the environment. Therefore, to comply with regulations, HCU manufacturers have introduced both timely and financially extensive sanitisation procedures. Our paper describes a novel, effective and easy to handle disinfection method for the above problematics without utilising hazardous chemicals. The method’s technical principle is electrolysis, resulting in drinking water quality regarding the analysed germs in the worldwide most commonly utilised heater-cooler device. The main aim of the study was to prove the efficacy and reliability of the device cleansing process. Furthermore, the economic impact of the novel method was evaluated. Therefore, we have undertaken 60 microbiological sampling series between December 2019 and November 2020 from a conventional HCU (3T LivaNova, Germany). During the total investigational period, no contamination with Pseudomonas aeruginosa or Legionellae could have been demonstrated in the HCU. The extreme slow-growing nontuberculous M. chimaera was detected only in one sample obtained from diamond electrode cleansed HCU water, and source of contamination was promptly eliminated by a simple technical modification of the device test-site. Additionally, the diamond electrode application is beneficial for eliminating potentially hazardous cleansing material from the process, which may affect otherwise both patients operated on cardiopulmonary bypass and the perfusionists.
format Online
Article
Text
id pubmed-9841820
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-98418202023-01-17 A solution for global hygienic challenges regarding the application of heater-cooler systems in cardiac surgery Foltan, Maik Nikisch, André Dembianny, Jan Miano, Anna-Lena Heinze, Johannes Klar, Dennis Göbölös, Laszlo Lehle, Karla Schmid, Christof Perfusion Original Papers 71.759 surgical procedures were performed in 2019 with the aid of cardiopulmonary bypass in Germany. To adjust the patient’s body temperature on extracorporeal circulation, the application of a heater-cooler unit (HCU) is mandatory. However, in case of insufficient sanitisation of HCU, life-threatening infections can be transmitted by the device to the patients, including Legionella bacteria, Mycobacterium chimaera, Pseudomonas aeruginosa. To avoid disease transmission, as a requirement for safe medical practice established by regulatory authorities, HCUs must be regularly disinfected by hazardous chemicals posing a danger for both handling humans and the environment. Therefore, to comply with regulations, HCU manufacturers have introduced both timely and financially extensive sanitisation procedures. Our paper describes a novel, effective and easy to handle disinfection method for the above problematics without utilising hazardous chemicals. The method’s technical principle is electrolysis, resulting in drinking water quality regarding the analysed germs in the worldwide most commonly utilised heater-cooler device. The main aim of the study was to prove the efficacy and reliability of the device cleansing process. Furthermore, the economic impact of the novel method was evaluated. Therefore, we have undertaken 60 microbiological sampling series between December 2019 and November 2020 from a conventional HCU (3T LivaNova, Germany). During the total investigational period, no contamination with Pseudomonas aeruginosa or Legionellae could have been demonstrated in the HCU. The extreme slow-growing nontuberculous M. chimaera was detected only in one sample obtained from diamond electrode cleansed HCU water, and source of contamination was promptly eliminated by a simple technical modification of the device test-site. Additionally, the diamond electrode application is beneficial for eliminating potentially hazardous cleansing material from the process, which may affect otherwise both patients operated on cardiopulmonary bypass and the perfusionists. SAGE Publications 2021-07-13 2023-01 /pmc/articles/PMC9841820/ /pubmed/34256647 http://dx.doi.org/10.1177/02676591211033483 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by-nc/4.0/This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access page (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Original Papers
Foltan, Maik
Nikisch, André
Dembianny, Jan
Miano, Anna-Lena
Heinze, Johannes
Klar, Dennis
Göbölös, Laszlo
Lehle, Karla
Schmid, Christof
A solution for global hygienic challenges regarding the application of heater-cooler systems in cardiac surgery
title A solution for global hygienic challenges regarding the application of heater-cooler systems in cardiac surgery
title_full A solution for global hygienic challenges regarding the application of heater-cooler systems in cardiac surgery
title_fullStr A solution for global hygienic challenges regarding the application of heater-cooler systems in cardiac surgery
title_full_unstemmed A solution for global hygienic challenges regarding the application of heater-cooler systems in cardiac surgery
title_short A solution for global hygienic challenges regarding the application of heater-cooler systems in cardiac surgery
title_sort solution for global hygienic challenges regarding the application of heater-cooler systems in cardiac surgery
topic Original Papers
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9841820/
https://www.ncbi.nlm.nih.gov/pubmed/34256647
http://dx.doi.org/10.1177/02676591211033483
work_keys_str_mv AT foltanmaik asolutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT nikischandre asolutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT dembiannyjan asolutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT mianoannalena asolutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT heinzejohannes asolutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT klardennis asolutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT goboloslaszlo asolutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT lehlekarla asolutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT schmidchristof asolutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT foltanmaik solutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT nikischandre solutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT dembiannyjan solutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT mianoannalena solutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT heinzejohannes solutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT klardennis solutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT goboloslaszlo solutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT lehlekarla solutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery
AT schmidchristof solutionforglobalhygienicchallengesregardingtheapplicationofheatercoolersystemsincardiacsurgery