Evaluation of the Temple Touch Pro™ noninvasive core-temperature monitoring system in 100 adults under general anesthesia: a prospective comparison with esophageal temperature

Perioperative hypothermia is still common and has relevant complication for the patient. An effective perioperative thermal management requires essentially an accurate method to measure core temperature. So far, only one study has investigated the new Temple Touch Pro™ (Medisim Ltd., Beit-Shemesh, I...

Descripción completa

Detalles Bibliográficos
Autores principales: Bräuer, Anselm, Fazliu, Albulena, Brandes, Ivo F., Vollnhals, Falk, Grote, Rolf, Menzel, Matthias
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Netherlands 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9852212/
https://www.ncbi.nlm.nih.gov/pubmed/35377051
http://dx.doi.org/10.1007/s10877-022-00851-z
_version_ 1784872565375762432
author Bräuer, Anselm
Fazliu, Albulena
Brandes, Ivo F.
Vollnhals, Falk
Grote, Rolf
Menzel, Matthias
author_facet Bräuer, Anselm
Fazliu, Albulena
Brandes, Ivo F.
Vollnhals, Falk
Grote, Rolf
Menzel, Matthias
author_sort Bräuer, Anselm
collection PubMed
description Perioperative hypothermia is still common and has relevant complication for the patient. An effective perioperative thermal management requires essentially an accurate method to measure core temperature. So far, only one study has investigated the new Temple Touch Pro™ (Medisim Ltd., Beit-Shemesh, Israel). during anesthesia Therefore, we assessed the agreement between the Temple Touch Pro™ thermometer (TTP) and distal esophageal temperature (T(Eso)) in a second study. After approval by the local ethics committee we studied 100 adult patients undergoing surgery with general anesthesia. Before induction of anesthesia the TTP sensor unit was attached to the skin above the temporal artery. After induction of anesthesia an esophageal temperature probe was placed in the distal esophagus. Recordings started 10 min after placement of the esophageal temperature probe to allow adequate warming of the probes. Pairs of temperature values were documented in five-minute intervals until emergence of anesthesia. Accuracy of the two methods was assessed by Bland-Altman comparisons of differences with multiple measurements. Core temperatures obtained with the TTP in adults showed a mean bias of -0.04 °C with 95% limits of agreement within − 0.99 °C to + 0.91 °C compared to an esophageal temperature probe. We consider the TTP as a reasonable tool for perioperative temperature monitoring. It is not accurate enough to be used as a reference method in scientific studies, but may be a useful tool especially for conscious patients undergoing neuraxial anesthesia or regional anesthesia with sedation. Trial registration This study was registered in the German Clinical Trials Register (DRKS-ID: 00024050), day of registration 12/01/2021.
format Online
Article
Text
id pubmed-9852212
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Springer Netherlands
record_format MEDLINE/PubMed
spelling pubmed-98522122023-01-21 Evaluation of the Temple Touch Pro™ noninvasive core-temperature monitoring system in 100 adults under general anesthesia: a prospective comparison with esophageal temperature Bräuer, Anselm Fazliu, Albulena Brandes, Ivo F. Vollnhals, Falk Grote, Rolf Menzel, Matthias J Clin Monit Comput Original Research Perioperative hypothermia is still common and has relevant complication for the patient. An effective perioperative thermal management requires essentially an accurate method to measure core temperature. So far, only one study has investigated the new Temple Touch Pro™ (Medisim Ltd., Beit-Shemesh, Israel). during anesthesia Therefore, we assessed the agreement between the Temple Touch Pro™ thermometer (TTP) and distal esophageal temperature (T(Eso)) in a second study. After approval by the local ethics committee we studied 100 adult patients undergoing surgery with general anesthesia. Before induction of anesthesia the TTP sensor unit was attached to the skin above the temporal artery. After induction of anesthesia an esophageal temperature probe was placed in the distal esophagus. Recordings started 10 min after placement of the esophageal temperature probe to allow adequate warming of the probes. Pairs of temperature values were documented in five-minute intervals until emergence of anesthesia. Accuracy of the two methods was assessed by Bland-Altman comparisons of differences with multiple measurements. Core temperatures obtained with the TTP in adults showed a mean bias of -0.04 °C with 95% limits of agreement within − 0.99 °C to + 0.91 °C compared to an esophageal temperature probe. We consider the TTP as a reasonable tool for perioperative temperature monitoring. It is not accurate enough to be used as a reference method in scientific studies, but may be a useful tool especially for conscious patients undergoing neuraxial anesthesia or regional anesthesia with sedation. Trial registration This study was registered in the German Clinical Trials Register (DRKS-ID: 00024050), day of registration 12/01/2021. Springer Netherlands 2022-04-04 2023 /pmc/articles/PMC9852212/ /pubmed/35377051 http://dx.doi.org/10.1007/s10877-022-00851-z Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Research
Bräuer, Anselm
Fazliu, Albulena
Brandes, Ivo F.
Vollnhals, Falk
Grote, Rolf
Menzel, Matthias
Evaluation of the Temple Touch Pro™ noninvasive core-temperature monitoring system in 100 adults under general anesthesia: a prospective comparison with esophageal temperature
title Evaluation of the Temple Touch Pro™ noninvasive core-temperature monitoring system in 100 adults under general anesthesia: a prospective comparison with esophageal temperature
title_full Evaluation of the Temple Touch Pro™ noninvasive core-temperature monitoring system in 100 adults under general anesthesia: a prospective comparison with esophageal temperature
title_fullStr Evaluation of the Temple Touch Pro™ noninvasive core-temperature monitoring system in 100 adults under general anesthesia: a prospective comparison with esophageal temperature
title_full_unstemmed Evaluation of the Temple Touch Pro™ noninvasive core-temperature monitoring system in 100 adults under general anesthesia: a prospective comparison with esophageal temperature
title_short Evaluation of the Temple Touch Pro™ noninvasive core-temperature monitoring system in 100 adults under general anesthesia: a prospective comparison with esophageal temperature
title_sort evaluation of the temple touch pro™ noninvasive core-temperature monitoring system in 100 adults under general anesthesia: a prospective comparison with esophageal temperature
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9852212/
https://www.ncbi.nlm.nih.gov/pubmed/35377051
http://dx.doi.org/10.1007/s10877-022-00851-z
work_keys_str_mv AT braueranselm evaluationofthetempletouchprononinvasivecoretemperaturemonitoringsystemin100adultsundergeneralanesthesiaaprospectivecomparisonwithesophagealtemperature
AT fazliualbulena evaluationofthetempletouchprononinvasivecoretemperaturemonitoringsystemin100adultsundergeneralanesthesiaaprospectivecomparisonwithesophagealtemperature
AT brandesivof evaluationofthetempletouchprononinvasivecoretemperaturemonitoringsystemin100adultsundergeneralanesthesiaaprospectivecomparisonwithesophagealtemperature
AT vollnhalsfalk evaluationofthetempletouchprononinvasivecoretemperaturemonitoringsystemin100adultsundergeneralanesthesiaaprospectivecomparisonwithesophagealtemperature
AT groterolf evaluationofthetempletouchprononinvasivecoretemperaturemonitoringsystemin100adultsundergeneralanesthesiaaprospectivecomparisonwithesophagealtemperature
AT menzelmatthias evaluationofthetempletouchprononinvasivecoretemperaturemonitoringsystemin100adultsundergeneralanesthesiaaprospectivecomparisonwithesophagealtemperature