Comparing Remote Speckle Plethysmography and Finger-Clip Photoplethysmography with Non-Invasive Finger Arterial Pressure Pulse Waves, Regarding Morphology and Arrival Time
Objective: The goal was to compare Speckle plethysmography (SPG) and Photoplethysmography (PPG) with non-invasive finger Arterial Pressure (fiAP) regarding Pulse Wave Morphology (PWM) and Pulse Arrival Time (PAT). Methods: Healthy volunteers (n = 8) were connected to a Non-Invasive Blood Pressure (N...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9854800/ https://www.ncbi.nlm.nih.gov/pubmed/36671673 http://dx.doi.org/10.3390/bioengineering10010101 |
_version_ | 1784873214170628096 |
---|---|
author | Herranz Olazabal, Jorge Wieringa, Fokko Hermeling, Evelien Van Hoof, Chris |
author_facet | Herranz Olazabal, Jorge Wieringa, Fokko Hermeling, Evelien Van Hoof, Chris |
author_sort | Herranz Olazabal, Jorge |
collection | PubMed |
description | Objective: The goal was to compare Speckle plethysmography (SPG) and Photoplethysmography (PPG) with non-invasive finger Arterial Pressure (fiAP) regarding Pulse Wave Morphology (PWM) and Pulse Arrival Time (PAT). Methods: Healthy volunteers (n = 8) were connected to a Non-Invasive Blood Pressure (NIBP) monitor providing fiAP pulse wave and PPG from a clinical transmission-mode SpO(2) finger clip. Biopac recorded 3-lead ECG. A camera placed at a 25 cm distance recorded a video stream (100 fps) of a finger illuminated by a laser diode at 639 nm. A chest belt (Polar) monitored respiration. All signals were recorded simultaneously during episodes of spontaneous breathing and paced breathing. Analysis: Post-processing was performed in Matlab to obtain SPG and analyze the SPG, PPG and fiAP mean absolute deviations (MADs) on PWM, plus PAT modulation. Results: Across 2599 beats, the average fiAP MAD with PPG was 0.17 (0–1) and with SPG 0.09 (0–1). PAT derived from ECG–fiAP correlated as follows: 0.65 for ECG–SPG and 0.67 for ECG–PPG. Conclusion: Compared to the clinical NIBP monitor fiAP reference, PWM from an experimental camera-derived non-contact reflective-mode SPG setup resembled fiAP significantly better than PPG from a simultaneously recorded clinical transmission-mode finger clip. For PAT values, no significant difference was found between ECG–SPG and ECG–PPG compared to ECG–fiAP. |
format | Online Article Text |
id | pubmed-9854800 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-98548002023-01-21 Comparing Remote Speckle Plethysmography and Finger-Clip Photoplethysmography with Non-Invasive Finger Arterial Pressure Pulse Waves, Regarding Morphology and Arrival Time Herranz Olazabal, Jorge Wieringa, Fokko Hermeling, Evelien Van Hoof, Chris Bioengineering (Basel) Article Objective: The goal was to compare Speckle plethysmography (SPG) and Photoplethysmography (PPG) with non-invasive finger Arterial Pressure (fiAP) regarding Pulse Wave Morphology (PWM) and Pulse Arrival Time (PAT). Methods: Healthy volunteers (n = 8) were connected to a Non-Invasive Blood Pressure (NIBP) monitor providing fiAP pulse wave and PPG from a clinical transmission-mode SpO(2) finger clip. Biopac recorded 3-lead ECG. A camera placed at a 25 cm distance recorded a video stream (100 fps) of a finger illuminated by a laser diode at 639 nm. A chest belt (Polar) monitored respiration. All signals were recorded simultaneously during episodes of spontaneous breathing and paced breathing. Analysis: Post-processing was performed in Matlab to obtain SPG and analyze the SPG, PPG and fiAP mean absolute deviations (MADs) on PWM, plus PAT modulation. Results: Across 2599 beats, the average fiAP MAD with PPG was 0.17 (0–1) and with SPG 0.09 (0–1). PAT derived from ECG–fiAP correlated as follows: 0.65 for ECG–SPG and 0.67 for ECG–PPG. Conclusion: Compared to the clinical NIBP monitor fiAP reference, PWM from an experimental camera-derived non-contact reflective-mode SPG setup resembled fiAP significantly better than PPG from a simultaneously recorded clinical transmission-mode finger clip. For PAT values, no significant difference was found between ECG–SPG and ECG–PPG compared to ECG–fiAP. MDPI 2023-01-11 /pmc/articles/PMC9854800/ /pubmed/36671673 http://dx.doi.org/10.3390/bioengineering10010101 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Herranz Olazabal, Jorge Wieringa, Fokko Hermeling, Evelien Van Hoof, Chris Comparing Remote Speckle Plethysmography and Finger-Clip Photoplethysmography with Non-Invasive Finger Arterial Pressure Pulse Waves, Regarding Morphology and Arrival Time |
title | Comparing Remote Speckle Plethysmography and Finger-Clip Photoplethysmography with Non-Invasive Finger Arterial Pressure Pulse Waves, Regarding Morphology and Arrival Time |
title_full | Comparing Remote Speckle Plethysmography and Finger-Clip Photoplethysmography with Non-Invasive Finger Arterial Pressure Pulse Waves, Regarding Morphology and Arrival Time |
title_fullStr | Comparing Remote Speckle Plethysmography and Finger-Clip Photoplethysmography with Non-Invasive Finger Arterial Pressure Pulse Waves, Regarding Morphology and Arrival Time |
title_full_unstemmed | Comparing Remote Speckle Plethysmography and Finger-Clip Photoplethysmography with Non-Invasive Finger Arterial Pressure Pulse Waves, Regarding Morphology and Arrival Time |
title_short | Comparing Remote Speckle Plethysmography and Finger-Clip Photoplethysmography with Non-Invasive Finger Arterial Pressure Pulse Waves, Regarding Morphology and Arrival Time |
title_sort | comparing remote speckle plethysmography and finger-clip photoplethysmography with non-invasive finger arterial pressure pulse waves, regarding morphology and arrival time |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9854800/ https://www.ncbi.nlm.nih.gov/pubmed/36671673 http://dx.doi.org/10.3390/bioengineering10010101 |
work_keys_str_mv | AT herranzolazabaljorge comparingremotespeckleplethysmographyandfingerclipphotoplethysmographywithnoninvasivefingerarterialpressurepulsewavesregardingmorphologyandarrivaltime AT wieringafokko comparingremotespeckleplethysmographyandfingerclipphotoplethysmographywithnoninvasivefingerarterialpressurepulsewavesregardingmorphologyandarrivaltime AT hermelingevelien comparingremotespeckleplethysmographyandfingerclipphotoplethysmographywithnoninvasivefingerarterialpressurepulsewavesregardingmorphologyandarrivaltime AT vanhoofchris comparingremotespeckleplethysmographyandfingerclipphotoplethysmographywithnoninvasivefingerarterialpressurepulsewavesregardingmorphologyandarrivaltime |