Case report: peri-device leakage after percutaneous left atrial appendage occlusion: plug, clip, or amputate?
BACKGROUND: Although peri-device leakage is frequently observed after left atrial appendage occlusion (LAAO), there is no consensus on the optimal management strategy. It is unknown whether additional plugging should be preferred over surgical exclusion of the LAA, as experience with additional plug...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9856255/ https://www.ncbi.nlm.nih.gov/pubmed/36694875 http://dx.doi.org/10.1093/ehjcr/ytac494 |
_version_ | 1784873575401914368 |
---|---|
author | Ramdat Misier, Nawin L Kharbanda, Rohit K van Schaagen, Frank R N de Groot, Natasja M S |
author_facet | Ramdat Misier, Nawin L Kharbanda, Rohit K van Schaagen, Frank R N de Groot, Natasja M S |
author_sort | Ramdat Misier, Nawin L |
collection | PubMed |
description | BACKGROUND: Although peri-device leakage is frequently observed after left atrial appendage occlusion (LAAO), there is no consensus on the optimal management strategy. It is unknown whether additional plugging should be preferred over surgical exclusion of the LAA, as experience with additional plugging is limited. CASE SUMMARY: In this case report, we demonstrate the clinical implications of additional plugging and surgical exclusion in a 65-year-old male patient with peri-device leakage and recurrent thromboembolic events. After the recurrence of paroxysmal atrial fibrillation (AF) and a transient ischaemic attack despite adequate anticoagulation, the patient was opted for re-do pulmonary vein isolation and LAAO with a Watchman device. Due to multiple ischaemic strokes and recurrent AF in combination with significant peri-device leakage, additional plugging with a second device was performed. Post-procedurally, the patient had another ischaemic stroke and persisting peri-device leakage was observed during follow-up. Due to progressive symptoms of AF and patient’s preference to discontinue DOAC, he underwent a Cox MAZE IV procedure, including amputation of the LAA with both devices. Within six months after surgery, the patient experienced two more ischaemic events. In the following two years, the patient remained free of any cerebrovascular accidents or recurrence of AF. DISCUSSION: Additional plugging of peri-device leakage is not always successful in stroke prevention. In combination with recurrent AF, progressive symptoms, contraindication for oral anticoagulation, and patient’s preference, surgical LAA exclusion could be preferred over additional plugging. |
format | Online Article Text |
id | pubmed-9856255 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Oxford University Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-98562552023-01-23 Case report: peri-device leakage after percutaneous left atrial appendage occlusion: plug, clip, or amputate? Ramdat Misier, Nawin L Kharbanda, Rohit K van Schaagen, Frank R N de Groot, Natasja M S Eur Heart J Case Rep Case Report BACKGROUND: Although peri-device leakage is frequently observed after left atrial appendage occlusion (LAAO), there is no consensus on the optimal management strategy. It is unknown whether additional plugging should be preferred over surgical exclusion of the LAA, as experience with additional plugging is limited. CASE SUMMARY: In this case report, we demonstrate the clinical implications of additional plugging and surgical exclusion in a 65-year-old male patient with peri-device leakage and recurrent thromboembolic events. After the recurrence of paroxysmal atrial fibrillation (AF) and a transient ischaemic attack despite adequate anticoagulation, the patient was opted for re-do pulmonary vein isolation and LAAO with a Watchman device. Due to multiple ischaemic strokes and recurrent AF in combination with significant peri-device leakage, additional plugging with a second device was performed. Post-procedurally, the patient had another ischaemic stroke and persisting peri-device leakage was observed during follow-up. Due to progressive symptoms of AF and patient’s preference to discontinue DOAC, he underwent a Cox MAZE IV procedure, including amputation of the LAA with both devices. Within six months after surgery, the patient experienced two more ischaemic events. In the following two years, the patient remained free of any cerebrovascular accidents or recurrence of AF. DISCUSSION: Additional plugging of peri-device leakage is not always successful in stroke prevention. In combination with recurrent AF, progressive symptoms, contraindication for oral anticoagulation, and patient’s preference, surgical LAA exclusion could be preferred over additional plugging. Oxford University Press 2023-01-03 /pmc/articles/PMC9856255/ /pubmed/36694875 http://dx.doi.org/10.1093/ehjcr/ytac494 Text en © The Author(s) 2023. Published by Oxford University Press on behalf of the European Society of Cardiology. https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial License (https://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com |
spellingShingle | Case Report Ramdat Misier, Nawin L Kharbanda, Rohit K van Schaagen, Frank R N de Groot, Natasja M S Case report: peri-device leakage after percutaneous left atrial appendage occlusion: plug, clip, or amputate? |
title | Case report: peri-device leakage after percutaneous left atrial appendage occlusion: plug, clip, or amputate? |
title_full | Case report: peri-device leakage after percutaneous left atrial appendage occlusion: plug, clip, or amputate? |
title_fullStr | Case report: peri-device leakage after percutaneous left atrial appendage occlusion: plug, clip, or amputate? |
title_full_unstemmed | Case report: peri-device leakage after percutaneous left atrial appendage occlusion: plug, clip, or amputate? |
title_short | Case report: peri-device leakage after percutaneous left atrial appendage occlusion: plug, clip, or amputate? |
title_sort | case report: peri-device leakage after percutaneous left atrial appendage occlusion: plug, clip, or amputate? |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9856255/ https://www.ncbi.nlm.nih.gov/pubmed/36694875 http://dx.doi.org/10.1093/ehjcr/ytac494 |
work_keys_str_mv | AT ramdatmisiernawinl casereportperideviceleakageafterpercutaneousleftatrialappendageocclusionplugcliporamputate AT kharbandarohitk casereportperideviceleakageafterpercutaneousleftatrialappendageocclusionplugcliporamputate AT vanschaagenfrankrn casereportperideviceleakageafterpercutaneousleftatrialappendageocclusionplugcliporamputate AT degrootnatasjams casereportperideviceleakageafterpercutaneousleftatrialappendageocclusionplugcliporamputate |