High-Frequency (30 MHz–6 GHz) Breast Tissue Characterization Stabilized by Suction Force for Intraoperative Tumor Margin Assessment

A gigahertz (GHz) range antenna formed by a coaxial probe has been applied for sensing cancerous breast lesions in the scanning platform with the assistance of a suction tube. The sensor structure was a planar central layer and a metallic sheath of size of 3 cm(2) connected to a network analyzer (ke...

Descripción completa

Detalles Bibliográficos
Autores principales: Mokhtari Dowlatabad, Hadi, Mamdouh, Amir, Yousefpour, Narges, Mahdavi, Reihane, Zandi, Ashkan, Hoseinpour, Parisa, Moosavi-Kiasari, Seyed Mohammad Sadegh, Abbasvandi, Fereshte, Kordehlachin, Yasin, Parniani, Mohammad, Mohammadpour-Aghdam, Karim, Faranoush, Pooya, Foroughi-Gilvaee, Mohammad Reza, Abdolahad, Mohammad
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9857665/
https://www.ncbi.nlm.nih.gov/pubmed/36672989
http://dx.doi.org/10.3390/diagnostics13020179
_version_ 1784873919831867392
author Mokhtari Dowlatabad, Hadi
Mamdouh, Amir
Yousefpour, Narges
Mahdavi, Reihane
Zandi, Ashkan
Hoseinpour, Parisa
Moosavi-Kiasari, Seyed Mohammad Sadegh
Abbasvandi, Fereshte
Kordehlachin, Yasin
Parniani, Mohammad
Mohammadpour-Aghdam, Karim
Faranoush, Pooya
Foroughi-Gilvaee, Mohammad Reza
Abdolahad, Mohammad
author_facet Mokhtari Dowlatabad, Hadi
Mamdouh, Amir
Yousefpour, Narges
Mahdavi, Reihane
Zandi, Ashkan
Hoseinpour, Parisa
Moosavi-Kiasari, Seyed Mohammad Sadegh
Abbasvandi, Fereshte
Kordehlachin, Yasin
Parniani, Mohammad
Mohammadpour-Aghdam, Karim
Faranoush, Pooya
Foroughi-Gilvaee, Mohammad Reza
Abdolahad, Mohammad
author_sort Mokhtari Dowlatabad, Hadi
collection PubMed
description A gigahertz (GHz) range antenna formed by a coaxial probe has been applied for sensing cancerous breast lesions in the scanning platform with the assistance of a suction tube. The sensor structure was a planar central layer and a metallic sheath of size of 3 cm(2) connected to a network analyzer (keySight FieldFox N9918A) with operational bandwidth up to 26.5 GHz. Cancer tumor cells have significantly higher water content (as a dipolar molecule) than normal breast cells, changing their polarization responses and dielectric losses to incoming GHz-based stimulation. Principal component analysis named S(11), related to the dispersion ratio of the input signal, is used as a parameter to identify malignant tumor cells in a mouse model (in vivo) and tumor specimens of breast cancer patients (in vitro) (both central and marginal parts). The results showed that S(11) values in the frequency range from 5 to 6 GHz were significantly higher in cancer-involved breast lesions. Histopathological analysis was the gold standard for achieving the S(11) calibration to distinguish normal from cancerous lesions. Our calibration on tumor specimens presented 82% positive predictive value (PPV), 100% negative predictive value (NPV), and 86% accuracy. Our goal is to apply this system as an in vivo non-invasive tumor margin scanner after further investigations in the future.
format Online
Article
Text
id pubmed-9857665
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-98576652023-01-21 High-Frequency (30 MHz–6 GHz) Breast Tissue Characterization Stabilized by Suction Force for Intraoperative Tumor Margin Assessment Mokhtari Dowlatabad, Hadi Mamdouh, Amir Yousefpour, Narges Mahdavi, Reihane Zandi, Ashkan Hoseinpour, Parisa Moosavi-Kiasari, Seyed Mohammad Sadegh Abbasvandi, Fereshte Kordehlachin, Yasin Parniani, Mohammad Mohammadpour-Aghdam, Karim Faranoush, Pooya Foroughi-Gilvaee, Mohammad Reza Abdolahad, Mohammad Diagnostics (Basel) Article A gigahertz (GHz) range antenna formed by a coaxial probe has been applied for sensing cancerous breast lesions in the scanning platform with the assistance of a suction tube. The sensor structure was a planar central layer and a metallic sheath of size of 3 cm(2) connected to a network analyzer (keySight FieldFox N9918A) with operational bandwidth up to 26.5 GHz. Cancer tumor cells have significantly higher water content (as a dipolar molecule) than normal breast cells, changing their polarization responses and dielectric losses to incoming GHz-based stimulation. Principal component analysis named S(11), related to the dispersion ratio of the input signal, is used as a parameter to identify malignant tumor cells in a mouse model (in vivo) and tumor specimens of breast cancer patients (in vitro) (both central and marginal parts). The results showed that S(11) values in the frequency range from 5 to 6 GHz were significantly higher in cancer-involved breast lesions. Histopathological analysis was the gold standard for achieving the S(11) calibration to distinguish normal from cancerous lesions. Our calibration on tumor specimens presented 82% positive predictive value (PPV), 100% negative predictive value (NPV), and 86% accuracy. Our goal is to apply this system as an in vivo non-invasive tumor margin scanner after further investigations in the future. MDPI 2023-01-04 /pmc/articles/PMC9857665/ /pubmed/36672989 http://dx.doi.org/10.3390/diagnostics13020179 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Mokhtari Dowlatabad, Hadi
Mamdouh, Amir
Yousefpour, Narges
Mahdavi, Reihane
Zandi, Ashkan
Hoseinpour, Parisa
Moosavi-Kiasari, Seyed Mohammad Sadegh
Abbasvandi, Fereshte
Kordehlachin, Yasin
Parniani, Mohammad
Mohammadpour-Aghdam, Karim
Faranoush, Pooya
Foroughi-Gilvaee, Mohammad Reza
Abdolahad, Mohammad
High-Frequency (30 MHz–6 GHz) Breast Tissue Characterization Stabilized by Suction Force for Intraoperative Tumor Margin Assessment
title High-Frequency (30 MHz–6 GHz) Breast Tissue Characterization Stabilized by Suction Force for Intraoperative Tumor Margin Assessment
title_full High-Frequency (30 MHz–6 GHz) Breast Tissue Characterization Stabilized by Suction Force for Intraoperative Tumor Margin Assessment
title_fullStr High-Frequency (30 MHz–6 GHz) Breast Tissue Characterization Stabilized by Suction Force for Intraoperative Tumor Margin Assessment
title_full_unstemmed High-Frequency (30 MHz–6 GHz) Breast Tissue Characterization Stabilized by Suction Force for Intraoperative Tumor Margin Assessment
title_short High-Frequency (30 MHz–6 GHz) Breast Tissue Characterization Stabilized by Suction Force for Intraoperative Tumor Margin Assessment
title_sort high-frequency (30 mhz–6 ghz) breast tissue characterization stabilized by suction force for intraoperative tumor margin assessment
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9857665/
https://www.ncbi.nlm.nih.gov/pubmed/36672989
http://dx.doi.org/10.3390/diagnostics13020179
work_keys_str_mv AT mokhtaridowlatabadhadi highfrequency30mhz6ghzbreasttissuecharacterizationstabilizedbysuctionforceforintraoperativetumormarginassessment
AT mamdouhamir highfrequency30mhz6ghzbreasttissuecharacterizationstabilizedbysuctionforceforintraoperativetumormarginassessment
AT yousefpournarges highfrequency30mhz6ghzbreasttissuecharacterizationstabilizedbysuctionforceforintraoperativetumormarginassessment
AT mahdavireihane highfrequency30mhz6ghzbreasttissuecharacterizationstabilizedbysuctionforceforintraoperativetumormarginassessment
AT zandiashkan highfrequency30mhz6ghzbreasttissuecharacterizationstabilizedbysuctionforceforintraoperativetumormarginassessment
AT hoseinpourparisa highfrequency30mhz6ghzbreasttissuecharacterizationstabilizedbysuctionforceforintraoperativetumormarginassessment
AT moosavikiasariseyedmohammadsadegh highfrequency30mhz6ghzbreasttissuecharacterizationstabilizedbysuctionforceforintraoperativetumormarginassessment
AT abbasvandifereshte highfrequency30mhz6ghzbreasttissuecharacterizationstabilizedbysuctionforceforintraoperativetumormarginassessment
AT kordehlachinyasin highfrequency30mhz6ghzbreasttissuecharacterizationstabilizedbysuctionforceforintraoperativetumormarginassessment
AT parnianimohammad highfrequency30mhz6ghzbreasttissuecharacterizationstabilizedbysuctionforceforintraoperativetumormarginassessment
AT mohammadpouraghdamkarim highfrequency30mhz6ghzbreasttissuecharacterizationstabilizedbysuctionforceforintraoperativetumormarginassessment
AT faranoushpooya highfrequency30mhz6ghzbreasttissuecharacterizationstabilizedbysuctionforceforintraoperativetumormarginassessment
AT foroughigilvaeemohammadreza highfrequency30mhz6ghzbreasttissuecharacterizationstabilizedbysuctionforceforintraoperativetumormarginassessment
AT abdolahadmohammad highfrequency30mhz6ghzbreasttissuecharacterizationstabilizedbysuctionforceforintraoperativetumormarginassessment