Comparative analysis of the mechanical limits of resistance in implant/abutment set of a new implant design: An in vitro study

OBJECTIVE: The aim of the present in vitro study was to evaluate the resistance on quasi-static forces and in the fatigue mechanical cycling of a new implant design compared to two other conventional implant designs. MATERIALS AND METHODS: Eighty-eight implants with their respective abutments were t...

Descripción completa

Detalles Bibliográficos
Autores principales: Bianchini, Marco Aurélio, Junior, Nilton De Bortoli, Dedavid, Berenice Anina, De Aza, Piedad N., Gehrke, Sergio Alexandre
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9858081/
https://www.ncbi.nlm.nih.gov/pubmed/36662887
http://dx.doi.org/10.1371/journal.pone.0280684
_version_ 1784874009985286144
author Bianchini, Marco Aurélio
Junior, Nilton De Bortoli
Dedavid, Berenice Anina
De Aza, Piedad N.
Gehrke, Sergio Alexandre
author_facet Bianchini, Marco Aurélio
Junior, Nilton De Bortoli
Dedavid, Berenice Anina
De Aza, Piedad N.
Gehrke, Sergio Alexandre
author_sort Bianchini, Marco Aurélio
collection PubMed
description OBJECTIVE: The aim of the present in vitro study was to evaluate the resistance on quasi-static forces and in the fatigue mechanical cycling of a new implant design compared to two other conventional implant designs. MATERIALS AND METHODS: Eighty-eight implants with their respective abutments were tested and distributed into four groups (n = 22 per group): Morse taper connection implant (MT group), conventional external hexagon implant (EH con group), new Collo implant of external hexagon with the smooth portion out of the bone insertion (EH out group), and new Collo implant of external hexagon with the implant platform inserted to the bone level (EH bl group). All the sets were subjected to quasi-static loading in a universal testing machine, and we measured the maximum resistance force supported by each sample. Another 12 samples from each group were submitted to the cyclic fatigue test at 4 intensities of forces (n = 3 per force): 80%, 60%, 40%, and 20%. The number of cycles supported by each sample at each force intensity was evaluated. RESULTS: The three groups of implants with external hexagon connection had similar maximum strength values of the sets (p > 0.05). Meanwhile, samples from the MT group showed the highest resistance values in comparison to the other three groups (p < 0.05). In the fatigue test, the Collo out group supported a smaller number of cycles that led to the fracture than the other 3 groups proposed at loads of 80%, 60%, and 40%, and only at the load value of 20% all groups had the same performance. CONCLUSIONS: Within the limitations of the present in vitro study, the results showed that the new Collo implant performs better when installed at bone level.
format Online
Article
Text
id pubmed-9858081
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-98580812023-01-21 Comparative analysis of the mechanical limits of resistance in implant/abutment set of a new implant design: An in vitro study Bianchini, Marco Aurélio Junior, Nilton De Bortoli Dedavid, Berenice Anina De Aza, Piedad N. Gehrke, Sergio Alexandre PLoS One Research Article OBJECTIVE: The aim of the present in vitro study was to evaluate the resistance on quasi-static forces and in the fatigue mechanical cycling of a new implant design compared to two other conventional implant designs. MATERIALS AND METHODS: Eighty-eight implants with their respective abutments were tested and distributed into four groups (n = 22 per group): Morse taper connection implant (MT group), conventional external hexagon implant (EH con group), new Collo implant of external hexagon with the smooth portion out of the bone insertion (EH out group), and new Collo implant of external hexagon with the implant platform inserted to the bone level (EH bl group). All the sets were subjected to quasi-static loading in a universal testing machine, and we measured the maximum resistance force supported by each sample. Another 12 samples from each group were submitted to the cyclic fatigue test at 4 intensities of forces (n = 3 per force): 80%, 60%, 40%, and 20%. The number of cycles supported by each sample at each force intensity was evaluated. RESULTS: The three groups of implants with external hexagon connection had similar maximum strength values of the sets (p > 0.05). Meanwhile, samples from the MT group showed the highest resistance values in comparison to the other three groups (p < 0.05). In the fatigue test, the Collo out group supported a smaller number of cycles that led to the fracture than the other 3 groups proposed at loads of 80%, 60%, and 40%, and only at the load value of 20% all groups had the same performance. CONCLUSIONS: Within the limitations of the present in vitro study, the results showed that the new Collo implant performs better when installed at bone level. Public Library of Science 2023-01-20 /pmc/articles/PMC9858081/ /pubmed/36662887 http://dx.doi.org/10.1371/journal.pone.0280684 Text en © 2023 Bianchini et al https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Bianchini, Marco Aurélio
Junior, Nilton De Bortoli
Dedavid, Berenice Anina
De Aza, Piedad N.
Gehrke, Sergio Alexandre
Comparative analysis of the mechanical limits of resistance in implant/abutment set of a new implant design: An in vitro study
title Comparative analysis of the mechanical limits of resistance in implant/abutment set of a new implant design: An in vitro study
title_full Comparative analysis of the mechanical limits of resistance in implant/abutment set of a new implant design: An in vitro study
title_fullStr Comparative analysis of the mechanical limits of resistance in implant/abutment set of a new implant design: An in vitro study
title_full_unstemmed Comparative analysis of the mechanical limits of resistance in implant/abutment set of a new implant design: An in vitro study
title_short Comparative analysis of the mechanical limits of resistance in implant/abutment set of a new implant design: An in vitro study
title_sort comparative analysis of the mechanical limits of resistance in implant/abutment set of a new implant design: an in vitro study
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9858081/
https://www.ncbi.nlm.nih.gov/pubmed/36662887
http://dx.doi.org/10.1371/journal.pone.0280684
work_keys_str_mv AT bianchinimarcoaurelio comparativeanalysisofthemechanicallimitsofresistanceinimplantabutmentsetofanewimplantdesignaninvitrostudy
AT juniorniltondebortoli comparativeanalysisofthemechanicallimitsofresistanceinimplantabutmentsetofanewimplantdesignaninvitrostudy
AT dedavidbereniceanina comparativeanalysisofthemechanicallimitsofresistanceinimplantabutmentsetofanewimplantdesignaninvitrostudy
AT deazapiedadn comparativeanalysisofthemechanicallimitsofresistanceinimplantabutmentsetofanewimplantdesignaninvitrostudy
AT gehrkesergioalexandre comparativeanalysisofthemechanicallimitsofresistanceinimplantabutmentsetofanewimplantdesignaninvitrostudy