Cargando…

Clinical Evaluation of BioFire COVID-19 Test, BioFire Respiratory Panel 2.1, and Cepheid Xpert Xpress SARS-CoV-2 Assays for Sample-to-Answer Detection of SARS-CoV-2

Background: Due to the extreme infectivity of SARS-CoV-2, sample-to-answer SARS-CoV-2 reverse transcription (RT) polymerase chain reaction (PCR) assays are urgently needed in order to facilitate infectious disease surveillance and control. The purpose of this study was to evaluate three sample-to-an...

Descripción completa

Detalles Bibliográficos
Autores principales: Park, Joonhong, Kim, So Yeon, Lee, Jaehyeon, Hong, Ki Ho
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9859140/
https://www.ncbi.nlm.nih.gov/pubmed/36672974
http://dx.doi.org/10.3390/genes14010233
_version_ 1784874280879652864
author Park, Joonhong
Kim, So Yeon
Lee, Jaehyeon
Hong, Ki Ho
author_facet Park, Joonhong
Kim, So Yeon
Lee, Jaehyeon
Hong, Ki Ho
author_sort Park, Joonhong
collection PubMed
description Background: Due to the extreme infectivity of SARS-CoV-2, sample-to-answer SARS-CoV-2 reverse transcription (RT) polymerase chain reaction (PCR) assays are urgently needed in order to facilitate infectious disease surveillance and control. The purpose of this study was to evaluate three sample-to-answer SARS-CoV-2 RT-PCR assays—BioFire COVID-19 Test, BioFire RP 2.1, and Cepheid Xpert Xpress SARS-CoV-2—using clinical samples. Methods: A total of 77 leftover nasopharyngeal swab (NP) swabs (36 positives and 41 negatives) confirmed by reference SARS-CoV-2 RT real-time (q) PCR assay were collected. The clinical sample concordance, as specified by their respective emergency use authorizations (EUAs), in comparison to the reference SARS-CoV-2 RT-qPCR assay, was assessed. Results: The results showed that all three sample-to-answer SARS-CoV-2 RT-PCR assays provided perfectly concordant results consistent with the reference SARS-CoV-2 RT-qPCR assay. The BioFire COVID-19 Test exhibited the best turnaround time (TAT) compared to the other assays, regardless of the test results, using one-way analysis of variance followed by Scheffe’s post hoc test (p < 0.001). The Xpert Xpress SARS-CoV-2 showed a shorter average TAT (mean ± standard deviation, 49.9 ± 3.1 min) in the positive samples compared to that (55.7 ± 2.5 min) of the negative samples. Conclusions: Our evaluation demonstrates that the BioFire COVID-19 Test, BioFire RP 2.1, and Cepheid Xpert Xpress SARS-CoV-2 assays compare favorably to the reference SARS-CoV-2 RT-qPCR assay, along with a 100% concordance in assay results for clinical samples and an acceptable analytical performance at their guaranteed limits of detection. The addition of a widely used simultaneous sample-to-answer SARS-CoV-2 RT-PCR assay will contribute to the number of medical laboratories able to test for COVID-19.
format Online
Article
Text
id pubmed-9859140
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-98591402023-01-21 Clinical Evaluation of BioFire COVID-19 Test, BioFire Respiratory Panel 2.1, and Cepheid Xpert Xpress SARS-CoV-2 Assays for Sample-to-Answer Detection of SARS-CoV-2 Park, Joonhong Kim, So Yeon Lee, Jaehyeon Hong, Ki Ho Genes (Basel) Article Background: Due to the extreme infectivity of SARS-CoV-2, sample-to-answer SARS-CoV-2 reverse transcription (RT) polymerase chain reaction (PCR) assays are urgently needed in order to facilitate infectious disease surveillance and control. The purpose of this study was to evaluate three sample-to-answer SARS-CoV-2 RT-PCR assays—BioFire COVID-19 Test, BioFire RP 2.1, and Cepheid Xpert Xpress SARS-CoV-2—using clinical samples. Methods: A total of 77 leftover nasopharyngeal swab (NP) swabs (36 positives and 41 negatives) confirmed by reference SARS-CoV-2 RT real-time (q) PCR assay were collected. The clinical sample concordance, as specified by their respective emergency use authorizations (EUAs), in comparison to the reference SARS-CoV-2 RT-qPCR assay, was assessed. Results: The results showed that all three sample-to-answer SARS-CoV-2 RT-PCR assays provided perfectly concordant results consistent with the reference SARS-CoV-2 RT-qPCR assay. The BioFire COVID-19 Test exhibited the best turnaround time (TAT) compared to the other assays, regardless of the test results, using one-way analysis of variance followed by Scheffe’s post hoc test (p < 0.001). The Xpert Xpress SARS-CoV-2 showed a shorter average TAT (mean ± standard deviation, 49.9 ± 3.1 min) in the positive samples compared to that (55.7 ± 2.5 min) of the negative samples. Conclusions: Our evaluation demonstrates that the BioFire COVID-19 Test, BioFire RP 2.1, and Cepheid Xpert Xpress SARS-CoV-2 assays compare favorably to the reference SARS-CoV-2 RT-qPCR assay, along with a 100% concordance in assay results for clinical samples and an acceptable analytical performance at their guaranteed limits of detection. The addition of a widely used simultaneous sample-to-answer SARS-CoV-2 RT-PCR assay will contribute to the number of medical laboratories able to test for COVID-19. MDPI 2023-01-16 /pmc/articles/PMC9859140/ /pubmed/36672974 http://dx.doi.org/10.3390/genes14010233 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Park, Joonhong
Kim, So Yeon
Lee, Jaehyeon
Hong, Ki Ho
Clinical Evaluation of BioFire COVID-19 Test, BioFire Respiratory Panel 2.1, and Cepheid Xpert Xpress SARS-CoV-2 Assays for Sample-to-Answer Detection of SARS-CoV-2
title Clinical Evaluation of BioFire COVID-19 Test, BioFire Respiratory Panel 2.1, and Cepheid Xpert Xpress SARS-CoV-2 Assays for Sample-to-Answer Detection of SARS-CoV-2
title_full Clinical Evaluation of BioFire COVID-19 Test, BioFire Respiratory Panel 2.1, and Cepheid Xpert Xpress SARS-CoV-2 Assays for Sample-to-Answer Detection of SARS-CoV-2
title_fullStr Clinical Evaluation of BioFire COVID-19 Test, BioFire Respiratory Panel 2.1, and Cepheid Xpert Xpress SARS-CoV-2 Assays for Sample-to-Answer Detection of SARS-CoV-2
title_full_unstemmed Clinical Evaluation of BioFire COVID-19 Test, BioFire Respiratory Panel 2.1, and Cepheid Xpert Xpress SARS-CoV-2 Assays for Sample-to-Answer Detection of SARS-CoV-2
title_short Clinical Evaluation of BioFire COVID-19 Test, BioFire Respiratory Panel 2.1, and Cepheid Xpert Xpress SARS-CoV-2 Assays for Sample-to-Answer Detection of SARS-CoV-2
title_sort clinical evaluation of biofire covid-19 test, biofire respiratory panel 2.1, and cepheid xpert xpress sars-cov-2 assays for sample-to-answer detection of sars-cov-2
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9859140/
https://www.ncbi.nlm.nih.gov/pubmed/36672974
http://dx.doi.org/10.3390/genes14010233
work_keys_str_mv AT parkjoonhong clinicalevaluationofbiofirecovid19testbiofirerespiratorypanel21andcepheidxpertxpresssarscov2assaysforsampletoanswerdetectionofsarscov2
AT kimsoyeon clinicalevaluationofbiofirecovid19testbiofirerespiratorypanel21andcepheidxpertxpresssarscov2assaysforsampletoanswerdetectionofsarscov2
AT leejaehyeon clinicalevaluationofbiofirecovid19testbiofirerespiratorypanel21andcepheidxpertxpresssarscov2assaysforsampletoanswerdetectionofsarscov2
AT hongkiho clinicalevaluationofbiofirecovid19testbiofirerespiratorypanel21andcepheidxpertxpresssarscov2assaysforsampletoanswerdetectionofsarscov2