Accuracy of Citrate Anticoagulant Amount, Volume, and Concentration in Evacuated Blood Collection Tubes Evaluated with UV Molecular Absorption Spectrometry on a Purified Water Model

Citrate anticoagulant concentration affects the results of coagulation tests. Until now, the end user had no direct insight into the quality of evacuated blood collection tubes. By introducing an easy-to-perform UV spectrometric method for citrate determination on a purified water model, we enabled...

Descripción completa

Detalles Bibliográficos
Autores principales: Gros, Nataša, Klobučar, Tadej, Gaber, Klara
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9860671/
https://www.ncbi.nlm.nih.gov/pubmed/36677544
http://dx.doi.org/10.3390/molecules28020486
_version_ 1784874643432144896
author Gros, Nataša
Klobučar, Tadej
Gaber, Klara
author_facet Gros, Nataša
Klobučar, Tadej
Gaber, Klara
author_sort Gros, Nataša
collection PubMed
description Citrate anticoagulant concentration affects the results of coagulation tests. Until now, the end user had no direct insight into the quality of evacuated blood collection tubes. By introducing an easy-to-perform UV spectrometric method for citrate determination on a purified water model, we enabled the evaluation of (1) the accuracy of the anticoagulant amount added into the tubes by a producer, (2) the accuracy of the volume of anticoagulant solution in the tube at the instant of examination, (3) the anticoagulant concentrations at a draw volume. We examined the Vacuette(®), Greiner BIO-ONE, Vacutube, LT Burnik d.o.o., and BD Vacutainer(®) tubes. The anticoagulant amount added into the tubes during production had a relative bias between 3.2 and 23.0%. The anticoagulant volume deficiency at the instant of examination expressed as a relative bias ranged between −11.6 and −91.1%. The anticoagulant concentration relative bias after the addition of purified water in a volume that equalled a nominal draw volume extended from 9.3 to 25.7%. Draw-volume was mostly compliant during shelf life. Only Vacutube lost water over time. Contamination with potassium, magnesium, or both was observed in all the tubes but did not exceed a 0.21 mmol/L level. This study enables medical laboratories to gain insight into the characteristics of the citrate blood collection tubes as one of the preanalytical variables. In situations that require anticoagulant adjustment for accurate results, this can help make the right decisions. The methodology gives producers additional means of controlling the quality of their production process.
format Online
Article
Text
id pubmed-9860671
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-98606712023-01-22 Accuracy of Citrate Anticoagulant Amount, Volume, and Concentration in Evacuated Blood Collection Tubes Evaluated with UV Molecular Absorption Spectrometry on a Purified Water Model Gros, Nataša Klobučar, Tadej Gaber, Klara Molecules Article Citrate anticoagulant concentration affects the results of coagulation tests. Until now, the end user had no direct insight into the quality of evacuated blood collection tubes. By introducing an easy-to-perform UV spectrometric method for citrate determination on a purified water model, we enabled the evaluation of (1) the accuracy of the anticoagulant amount added into the tubes by a producer, (2) the accuracy of the volume of anticoagulant solution in the tube at the instant of examination, (3) the anticoagulant concentrations at a draw volume. We examined the Vacuette(®), Greiner BIO-ONE, Vacutube, LT Burnik d.o.o., and BD Vacutainer(®) tubes. The anticoagulant amount added into the tubes during production had a relative bias between 3.2 and 23.0%. The anticoagulant volume deficiency at the instant of examination expressed as a relative bias ranged between −11.6 and −91.1%. The anticoagulant concentration relative bias after the addition of purified water in a volume that equalled a nominal draw volume extended from 9.3 to 25.7%. Draw-volume was mostly compliant during shelf life. Only Vacutube lost water over time. Contamination with potassium, magnesium, or both was observed in all the tubes but did not exceed a 0.21 mmol/L level. This study enables medical laboratories to gain insight into the characteristics of the citrate blood collection tubes as one of the preanalytical variables. In situations that require anticoagulant adjustment for accurate results, this can help make the right decisions. The methodology gives producers additional means of controlling the quality of their production process. MDPI 2023-01-04 /pmc/articles/PMC9860671/ /pubmed/36677544 http://dx.doi.org/10.3390/molecules28020486 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Gros, Nataša
Klobučar, Tadej
Gaber, Klara
Accuracy of Citrate Anticoagulant Amount, Volume, and Concentration in Evacuated Blood Collection Tubes Evaluated with UV Molecular Absorption Spectrometry on a Purified Water Model
title Accuracy of Citrate Anticoagulant Amount, Volume, and Concentration in Evacuated Blood Collection Tubes Evaluated with UV Molecular Absorption Spectrometry on a Purified Water Model
title_full Accuracy of Citrate Anticoagulant Amount, Volume, and Concentration in Evacuated Blood Collection Tubes Evaluated with UV Molecular Absorption Spectrometry on a Purified Water Model
title_fullStr Accuracy of Citrate Anticoagulant Amount, Volume, and Concentration in Evacuated Blood Collection Tubes Evaluated with UV Molecular Absorption Spectrometry on a Purified Water Model
title_full_unstemmed Accuracy of Citrate Anticoagulant Amount, Volume, and Concentration in Evacuated Blood Collection Tubes Evaluated with UV Molecular Absorption Spectrometry on a Purified Water Model
title_short Accuracy of Citrate Anticoagulant Amount, Volume, and Concentration in Evacuated Blood Collection Tubes Evaluated with UV Molecular Absorption Spectrometry on a Purified Water Model
title_sort accuracy of citrate anticoagulant amount, volume, and concentration in evacuated blood collection tubes evaluated with uv molecular absorption spectrometry on a purified water model
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9860671/
https://www.ncbi.nlm.nih.gov/pubmed/36677544
http://dx.doi.org/10.3390/molecules28020486
work_keys_str_mv AT grosnatasa accuracyofcitrateanticoagulantamountvolumeandconcentrationinevacuatedbloodcollectiontubesevaluatedwithuvmolecularabsorptionspectrometryonapurifiedwatermodel
AT klobucartadej accuracyofcitrateanticoagulantamountvolumeandconcentrationinevacuatedbloodcollectiontubesevaluatedwithuvmolecularabsorptionspectrometryonapurifiedwatermodel
AT gaberklara accuracyofcitrateanticoagulantamountvolumeandconcentrationinevacuatedbloodcollectiontubesevaluatedwithuvmolecularabsorptionspectrometryonapurifiedwatermodel