Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department?

Background and Objectives: Hemolysis is reported to be present in up to 10% of blood gas specimens in the central lab; however, few data on the incidence of hemolysis using a point-of-care testing (POCT) blood gas analysis are available in the setting of the emergency department. The aims of this st...

Descripción completa

Detalles Bibliográficos
Autores principales: Nigro, Marianna, Valli, Gabriele, Marchionne, Maria Luisa, Sattarinia, Dario, Silvestrini, Fabiana, De Pietro, Daniele, Fazzini, Simone, Roselli, Giorgia, Spallino, Andrea, Praticò, Valentina, Mirante, Enrico, Castaldo, Ersilia, Pugliese, Francesco Rocco, Cicchini, Claudia, Ancona, Carlo, De Marco, Francesca, Ruggieri, Maria Pia, Di Somma, Salvatore
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9861078/
https://www.ncbi.nlm.nih.gov/pubmed/36676689
http://dx.doi.org/10.3390/medicina59010066
_version_ 1784874751487901696
author Nigro, Marianna
Valli, Gabriele
Marchionne, Maria Luisa
Sattarinia, Dario
Silvestrini, Fabiana
De Pietro, Daniele
Fazzini, Simone
Roselli, Giorgia
Spallino, Andrea
Praticò, Valentina
Mirante, Enrico
Castaldo, Ersilia
Pugliese, Francesco Rocco
Cicchini, Claudia
Ancona, Carlo
De Marco, Francesca
Ruggieri, Maria Pia
Di Somma, Salvatore
author_facet Nigro, Marianna
Valli, Gabriele
Marchionne, Maria Luisa
Sattarinia, Dario
Silvestrini, Fabiana
De Pietro, Daniele
Fazzini, Simone
Roselli, Giorgia
Spallino, Andrea
Praticò, Valentina
Mirante, Enrico
Castaldo, Ersilia
Pugliese, Francesco Rocco
Cicchini, Claudia
Ancona, Carlo
De Marco, Francesca
Ruggieri, Maria Pia
Di Somma, Salvatore
author_sort Nigro, Marianna
collection PubMed
description Background and Objectives: Hemolysis is reported to be present in up to 10% of blood gas specimens in the central lab; however, few data on the incidence of hemolysis using a point-of-care testing (POCT) blood gas analysis are available in the setting of the emergency department. The aims of this study were: (1) to analyze the prevalence of hemolysis in blood gas samples collected in the ED using a POCT device; and (2) to evaluate the impact of hemolysis on blood sample results and its clinical consequences. Materials and Methods: We collected 525 consecutive POCT arterial blood gas samples using syringes with electrolyte-balanced heparin within 3 different EDs in the metropolitan area of Rome. Immediately after the collection, the blood samples were checked for the presence of hemolysis with a POCT instrument (i.e., HEMCHECK, H-10 (®)). The samples were then subsequently processed for blood gasses, and an electrolytes analysis by a second operator blinded for the hemolysis results. A venous blood sample was simultaneously collected, analyzed for it’s potassium value, and used as a reference. Results: Of the samples, 472 were considered for the statistics, while 53 were excluded due to the high percentage of hemolysis due to operator fault in carrying out the measurement. The final mean hemolysis per operator was 12% (±13% SD), and the total final hemolysis was 14.4%.Potassium (K(+)) was significantly higher in the hemolyzed group compared with the non-hemolyzed sample (4.60 ± 0.11 vs. 3.99 ± 0.03 mEq/L; p < 0.001), and there were differences between arterial potassium versus venous potassium (D(a-v) K+, 0.29 ± 0.06 vs.−0.19 ± 0.02 mEq/L, p < 0.01). A Bland–Altman analysis confirmed that hemolysis significantly overestimated blood potassium level. Conclusion: Almost 12% of POCT blood gas analysis samples performed in the ED could be hemolyzed, and the presence of this hemolysis is not routinely detected. This could cause an error in the interpretation of the results, leading to the consideration of potassium concentrations being below the lower limit within the normal limits and also leading to the diagnosis of false hyperkalemia, which would have potential clinical consequences in therapeutic decision-making in the ED. The routine use of a POCT hemolysis detector could help prevent any misdiagnoses.
format Online
Article
Text
id pubmed-9861078
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-98610782023-01-22 Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department? Nigro, Marianna Valli, Gabriele Marchionne, Maria Luisa Sattarinia, Dario Silvestrini, Fabiana De Pietro, Daniele Fazzini, Simone Roselli, Giorgia Spallino, Andrea Praticò, Valentina Mirante, Enrico Castaldo, Ersilia Pugliese, Francesco Rocco Cicchini, Claudia Ancona, Carlo De Marco, Francesca Ruggieri, Maria Pia Di Somma, Salvatore Medicina (Kaunas) Article Background and Objectives: Hemolysis is reported to be present in up to 10% of blood gas specimens in the central lab; however, few data on the incidence of hemolysis using a point-of-care testing (POCT) blood gas analysis are available in the setting of the emergency department. The aims of this study were: (1) to analyze the prevalence of hemolysis in blood gas samples collected in the ED using a POCT device; and (2) to evaluate the impact of hemolysis on blood sample results and its clinical consequences. Materials and Methods: We collected 525 consecutive POCT arterial blood gas samples using syringes with electrolyte-balanced heparin within 3 different EDs in the metropolitan area of Rome. Immediately after the collection, the blood samples were checked for the presence of hemolysis with a POCT instrument (i.e., HEMCHECK, H-10 (®)). The samples were then subsequently processed for blood gasses, and an electrolytes analysis by a second operator blinded for the hemolysis results. A venous blood sample was simultaneously collected, analyzed for it’s potassium value, and used as a reference. Results: Of the samples, 472 were considered for the statistics, while 53 were excluded due to the high percentage of hemolysis due to operator fault in carrying out the measurement. The final mean hemolysis per operator was 12% (±13% SD), and the total final hemolysis was 14.4%.Potassium (K(+)) was significantly higher in the hemolyzed group compared with the non-hemolyzed sample (4.60 ± 0.11 vs. 3.99 ± 0.03 mEq/L; p < 0.001), and there were differences between arterial potassium versus venous potassium (D(a-v) K+, 0.29 ± 0.06 vs.−0.19 ± 0.02 mEq/L, p < 0.01). A Bland–Altman analysis confirmed that hemolysis significantly overestimated blood potassium level. Conclusion: Almost 12% of POCT blood gas analysis samples performed in the ED could be hemolyzed, and the presence of this hemolysis is not routinely detected. This could cause an error in the interpretation of the results, leading to the consideration of potassium concentrations being below the lower limit within the normal limits and also leading to the diagnosis of false hyperkalemia, which would have potential clinical consequences in therapeutic decision-making in the ED. The routine use of a POCT hemolysis detector could help prevent any misdiagnoses. MDPI 2022-12-28 /pmc/articles/PMC9861078/ /pubmed/36676689 http://dx.doi.org/10.3390/medicina59010066 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Nigro, Marianna
Valli, Gabriele
Marchionne, Maria Luisa
Sattarinia, Dario
Silvestrini, Fabiana
De Pietro, Daniele
Fazzini, Simone
Roselli, Giorgia
Spallino, Andrea
Praticò, Valentina
Mirante, Enrico
Castaldo, Ersilia
Pugliese, Francesco Rocco
Cicchini, Claudia
Ancona, Carlo
De Marco, Francesca
Ruggieri, Maria Pia
Di Somma, Salvatore
Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department?
title Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department?
title_full Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department?
title_fullStr Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department?
title_full_unstemmed Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department?
title_short Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department?
title_sort is there a risk of misinterpretation of potassium concentration from undetectable hemolysis using a poct blood gas analyzer in the emergency department?
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9861078/
https://www.ncbi.nlm.nih.gov/pubmed/36676689
http://dx.doi.org/10.3390/medicina59010066
work_keys_str_mv AT nigromarianna isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT valligabriele isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT marchionnemarialuisa isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT sattariniadario isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT silvestrinifabiana isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT depietrodaniele isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT fazzinisimone isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT roselligiorgia isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT spallinoandrea isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT praticovalentina isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT miranteenrico isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT castaldoersilia isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT pugliesefrancescorocco isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT cicchiniclaudia isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT anconacarlo isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT demarcofrancesca isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT ruggierimariapia isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment
AT disommasalvatore isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment