Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department?
Background and Objectives: Hemolysis is reported to be present in up to 10% of blood gas specimens in the central lab; however, few data on the incidence of hemolysis using a point-of-care testing (POCT) blood gas analysis are available in the setting of the emergency department. The aims of this st...
Autores principales: | , , , , , , , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9861078/ https://www.ncbi.nlm.nih.gov/pubmed/36676689 http://dx.doi.org/10.3390/medicina59010066 |
_version_ | 1784874751487901696 |
---|---|
author | Nigro, Marianna Valli, Gabriele Marchionne, Maria Luisa Sattarinia, Dario Silvestrini, Fabiana De Pietro, Daniele Fazzini, Simone Roselli, Giorgia Spallino, Andrea Praticò, Valentina Mirante, Enrico Castaldo, Ersilia Pugliese, Francesco Rocco Cicchini, Claudia Ancona, Carlo De Marco, Francesca Ruggieri, Maria Pia Di Somma, Salvatore |
author_facet | Nigro, Marianna Valli, Gabriele Marchionne, Maria Luisa Sattarinia, Dario Silvestrini, Fabiana De Pietro, Daniele Fazzini, Simone Roselli, Giorgia Spallino, Andrea Praticò, Valentina Mirante, Enrico Castaldo, Ersilia Pugliese, Francesco Rocco Cicchini, Claudia Ancona, Carlo De Marco, Francesca Ruggieri, Maria Pia Di Somma, Salvatore |
author_sort | Nigro, Marianna |
collection | PubMed |
description | Background and Objectives: Hemolysis is reported to be present in up to 10% of blood gas specimens in the central lab; however, few data on the incidence of hemolysis using a point-of-care testing (POCT) blood gas analysis are available in the setting of the emergency department. The aims of this study were: (1) to analyze the prevalence of hemolysis in blood gas samples collected in the ED using a POCT device; and (2) to evaluate the impact of hemolysis on blood sample results and its clinical consequences. Materials and Methods: We collected 525 consecutive POCT arterial blood gas samples using syringes with electrolyte-balanced heparin within 3 different EDs in the metropolitan area of Rome. Immediately after the collection, the blood samples were checked for the presence of hemolysis with a POCT instrument (i.e., HEMCHECK, H-10 (®)). The samples were then subsequently processed for blood gasses, and an electrolytes analysis by a second operator blinded for the hemolysis results. A venous blood sample was simultaneously collected, analyzed for it’s potassium value, and used as a reference. Results: Of the samples, 472 were considered for the statistics, while 53 were excluded due to the high percentage of hemolysis due to operator fault in carrying out the measurement. The final mean hemolysis per operator was 12% (±13% SD), and the total final hemolysis was 14.4%.Potassium (K(+)) was significantly higher in the hemolyzed group compared with the non-hemolyzed sample (4.60 ± 0.11 vs. 3.99 ± 0.03 mEq/L; p < 0.001), and there were differences between arterial potassium versus venous potassium (D(a-v) K+, 0.29 ± 0.06 vs.−0.19 ± 0.02 mEq/L, p < 0.01). A Bland–Altman analysis confirmed that hemolysis significantly overestimated blood potassium level. Conclusion: Almost 12% of POCT blood gas analysis samples performed in the ED could be hemolyzed, and the presence of this hemolysis is not routinely detected. This could cause an error in the interpretation of the results, leading to the consideration of potassium concentrations being below the lower limit within the normal limits and also leading to the diagnosis of false hyperkalemia, which would have potential clinical consequences in therapeutic decision-making in the ED. The routine use of a POCT hemolysis detector could help prevent any misdiagnoses. |
format | Online Article Text |
id | pubmed-9861078 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-98610782023-01-22 Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department? Nigro, Marianna Valli, Gabriele Marchionne, Maria Luisa Sattarinia, Dario Silvestrini, Fabiana De Pietro, Daniele Fazzini, Simone Roselli, Giorgia Spallino, Andrea Praticò, Valentina Mirante, Enrico Castaldo, Ersilia Pugliese, Francesco Rocco Cicchini, Claudia Ancona, Carlo De Marco, Francesca Ruggieri, Maria Pia Di Somma, Salvatore Medicina (Kaunas) Article Background and Objectives: Hemolysis is reported to be present in up to 10% of blood gas specimens in the central lab; however, few data on the incidence of hemolysis using a point-of-care testing (POCT) blood gas analysis are available in the setting of the emergency department. The aims of this study were: (1) to analyze the prevalence of hemolysis in blood gas samples collected in the ED using a POCT device; and (2) to evaluate the impact of hemolysis on blood sample results and its clinical consequences. Materials and Methods: We collected 525 consecutive POCT arterial blood gas samples using syringes with electrolyte-balanced heparin within 3 different EDs in the metropolitan area of Rome. Immediately after the collection, the blood samples were checked for the presence of hemolysis with a POCT instrument (i.e., HEMCHECK, H-10 (®)). The samples were then subsequently processed for blood gasses, and an electrolytes analysis by a second operator blinded for the hemolysis results. A venous blood sample was simultaneously collected, analyzed for it’s potassium value, and used as a reference. Results: Of the samples, 472 were considered for the statistics, while 53 were excluded due to the high percentage of hemolysis due to operator fault in carrying out the measurement. The final mean hemolysis per operator was 12% (±13% SD), and the total final hemolysis was 14.4%.Potassium (K(+)) was significantly higher in the hemolyzed group compared with the non-hemolyzed sample (4.60 ± 0.11 vs. 3.99 ± 0.03 mEq/L; p < 0.001), and there were differences between arterial potassium versus venous potassium (D(a-v) K+, 0.29 ± 0.06 vs.−0.19 ± 0.02 mEq/L, p < 0.01). A Bland–Altman analysis confirmed that hemolysis significantly overestimated blood potassium level. Conclusion: Almost 12% of POCT blood gas analysis samples performed in the ED could be hemolyzed, and the presence of this hemolysis is not routinely detected. This could cause an error in the interpretation of the results, leading to the consideration of potassium concentrations being below the lower limit within the normal limits and also leading to the diagnosis of false hyperkalemia, which would have potential clinical consequences in therapeutic decision-making in the ED. The routine use of a POCT hemolysis detector could help prevent any misdiagnoses. MDPI 2022-12-28 /pmc/articles/PMC9861078/ /pubmed/36676689 http://dx.doi.org/10.3390/medicina59010066 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Nigro, Marianna Valli, Gabriele Marchionne, Maria Luisa Sattarinia, Dario Silvestrini, Fabiana De Pietro, Daniele Fazzini, Simone Roselli, Giorgia Spallino, Andrea Praticò, Valentina Mirante, Enrico Castaldo, Ersilia Pugliese, Francesco Rocco Cicchini, Claudia Ancona, Carlo De Marco, Francesca Ruggieri, Maria Pia Di Somma, Salvatore Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department? |
title | Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department? |
title_full | Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department? |
title_fullStr | Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department? |
title_full_unstemmed | Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department? |
title_short | Is There a Risk of Misinterpretation of Potassium Concentration from Undetectable Hemolysis Using a POCT Blood Gas Analyzer in the Emergency Department? |
title_sort | is there a risk of misinterpretation of potassium concentration from undetectable hemolysis using a poct blood gas analyzer in the emergency department? |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9861078/ https://www.ncbi.nlm.nih.gov/pubmed/36676689 http://dx.doi.org/10.3390/medicina59010066 |
work_keys_str_mv | AT nigromarianna isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT valligabriele isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT marchionnemarialuisa isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT sattariniadario isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT silvestrinifabiana isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT depietrodaniele isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT fazzinisimone isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT roselligiorgia isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT spallinoandrea isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT praticovalentina isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT miranteenrico isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT castaldoersilia isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT pugliesefrancescorocco isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT cicchiniclaudia isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT anconacarlo isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT demarcofrancesca isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT ruggierimariapia isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment AT disommasalvatore isthereariskofmisinterpretationofpotassiumconcentrationfromundetectablehemolysisusingapoctbloodgasanalyzerintheemergencydepartment |