Cargando…
Removal of Insertable Cardiac Monitor for Cryptogenic Stroke: A Technical Note
Objectives: Insertable cardiac monitors (ICM) allow continuous long-term electrocardiogram monitoring and the detection of paroxysmal atrial fibrillation (PAF) in patients with cryptogenic stroke (CS). Several years have passed since ICM was indicated for CS, and many stroke neurologists will experi...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9865790/ https://www.ncbi.nlm.nih.gov/pubmed/36661910 http://dx.doi.org/10.3390/jcdd10010015 |
_version_ | 1784875926560964608 |
---|---|
author | Egashira, Shuhei Kimura, Naoto Doijiri, Ryosuke |
author_facet | Egashira, Shuhei Kimura, Naoto Doijiri, Ryosuke |
author_sort | Egashira, Shuhei |
collection | PubMed |
description | Objectives: Insertable cardiac monitors (ICM) allow continuous long-term electrocardiogram monitoring and the detection of paroxysmal atrial fibrillation (PAF) in patients with cryptogenic stroke (CS). Several years have passed since ICM was indicated for CS, and many stroke neurologists will experience cases in which ICM removal is required. As a standard protocol, reincision of the wound at the time of implantation has been proposed by ICM brands. However, it may be difficult due to adhesions of subcutaneous tissue, migration of the device from its original position, and the capsule formed around the device. Our objective is to describe simple alternative techniques for successful ICM removal. Materials and Methods: From December 2016 to September 2021, 37 patients with CS underwent ICM removal at our institution. The device was removed through an incision directly above the proximal end of the device, perpendicular to the wound at the time of ICM implantation. The subcutaneous tissue was removed bluntly using forceps along the edges of the proximal end of the device. When a capsule was attached to the device, we cut the capsule with the blade to release the device. Once the device was visible, the proximal end of the device was grasped with forceps, and the device was pulled from the pocket with gentle traction. All patients undergoing ICM removal received a systematic check for wound dehiscence, wound infection, bleeding, and tissue ischemia at an outpatient examination of 1 week. The 37 patients who underwent removal of ICM were retrospectively reviewed in the medical record and analyzed for procedural success, intraoperative complications, and wound course at one week. Results: All patients achieved procedural success. There were no intraoperative complications, wound dehiscence, bleeding, or skin ischemia at one week postoperatively. The reasons for removal were battery depletion in 65%, early removal before battery life after PAF detection in 32%, and exposure to the body surface in 3%. The devices removed were 62% Reveal LINQ (Medtronic, Minneapolis), 30% Confirm Rx (Abbott, Illinois), and 8% BioMonitor 2 (BIOTRONIK, Berlin), indicating that our method is effective regardless of model. Conclusion: We describe a simple technique for ICM removal for CS that is safe, reliable, and potentially effective in wound healing. |
format | Online Article Text |
id | pubmed-9865790 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-98657902023-01-22 Removal of Insertable Cardiac Monitor for Cryptogenic Stroke: A Technical Note Egashira, Shuhei Kimura, Naoto Doijiri, Ryosuke J Cardiovasc Dev Dis Technical Note Objectives: Insertable cardiac monitors (ICM) allow continuous long-term electrocardiogram monitoring and the detection of paroxysmal atrial fibrillation (PAF) in patients with cryptogenic stroke (CS). Several years have passed since ICM was indicated for CS, and many stroke neurologists will experience cases in which ICM removal is required. As a standard protocol, reincision of the wound at the time of implantation has been proposed by ICM brands. However, it may be difficult due to adhesions of subcutaneous tissue, migration of the device from its original position, and the capsule formed around the device. Our objective is to describe simple alternative techniques for successful ICM removal. Materials and Methods: From December 2016 to September 2021, 37 patients with CS underwent ICM removal at our institution. The device was removed through an incision directly above the proximal end of the device, perpendicular to the wound at the time of ICM implantation. The subcutaneous tissue was removed bluntly using forceps along the edges of the proximal end of the device. When a capsule was attached to the device, we cut the capsule with the blade to release the device. Once the device was visible, the proximal end of the device was grasped with forceps, and the device was pulled from the pocket with gentle traction. All patients undergoing ICM removal received a systematic check for wound dehiscence, wound infection, bleeding, and tissue ischemia at an outpatient examination of 1 week. The 37 patients who underwent removal of ICM were retrospectively reviewed in the medical record and analyzed for procedural success, intraoperative complications, and wound course at one week. Results: All patients achieved procedural success. There were no intraoperative complications, wound dehiscence, bleeding, or skin ischemia at one week postoperatively. The reasons for removal were battery depletion in 65%, early removal before battery life after PAF detection in 32%, and exposure to the body surface in 3%. The devices removed were 62% Reveal LINQ (Medtronic, Minneapolis), 30% Confirm Rx (Abbott, Illinois), and 8% BioMonitor 2 (BIOTRONIK, Berlin), indicating that our method is effective regardless of model. Conclusion: We describe a simple technique for ICM removal for CS that is safe, reliable, and potentially effective in wound healing. MDPI 2023-01-01 /pmc/articles/PMC9865790/ /pubmed/36661910 http://dx.doi.org/10.3390/jcdd10010015 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Technical Note Egashira, Shuhei Kimura, Naoto Doijiri, Ryosuke Removal of Insertable Cardiac Monitor for Cryptogenic Stroke: A Technical Note |
title | Removal of Insertable Cardiac Monitor for Cryptogenic Stroke: A Technical Note |
title_full | Removal of Insertable Cardiac Monitor for Cryptogenic Stroke: A Technical Note |
title_fullStr | Removal of Insertable Cardiac Monitor for Cryptogenic Stroke: A Technical Note |
title_full_unstemmed | Removal of Insertable Cardiac Monitor for Cryptogenic Stroke: A Technical Note |
title_short | Removal of Insertable Cardiac Monitor for Cryptogenic Stroke: A Technical Note |
title_sort | removal of insertable cardiac monitor for cryptogenic stroke: a technical note |
topic | Technical Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9865790/ https://www.ncbi.nlm.nih.gov/pubmed/36661910 http://dx.doi.org/10.3390/jcdd10010015 |
work_keys_str_mv | AT egashirashuhei removalofinsertablecardiacmonitorforcryptogenicstrokeatechnicalnote AT kimuranaoto removalofinsertablecardiacmonitorforcryptogenicstrokeatechnicalnote AT doijiriryosuke removalofinsertablecardiacmonitorforcryptogenicstrokeatechnicalnote |