Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets

Short implants were introduced to reduce morbidity, treatment duration, and complex bone regeneration interventions in atrophic jaws and to improve patient-reported outcomes. This study aimed to determine the insertion torque (IT), removal torque (RT), and resonance frequency analysis (RFA) values o...

Descripción completa

Detalles Bibliográficos
Autores principales: Comuzzi, Luca, Tumedei, Margherita, Romasco, Tea, Petrini, Morena, Afrashtehfar, Kelvin I., Inchingolo, Francesco, Piattelli, Adriano, Di Pietro, Natalia
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9866818/
https://www.ncbi.nlm.nih.gov/pubmed/36662057
http://dx.doi.org/10.3390/jfb14010010
_version_ 1784876186280656896
author Comuzzi, Luca
Tumedei, Margherita
Romasco, Tea
Petrini, Morena
Afrashtehfar, Kelvin I.
Inchingolo, Francesco
Piattelli, Adriano
Di Pietro, Natalia
author_facet Comuzzi, Luca
Tumedei, Margherita
Romasco, Tea
Petrini, Morena
Afrashtehfar, Kelvin I.
Inchingolo, Francesco
Piattelli, Adriano
Di Pietro, Natalia
author_sort Comuzzi, Luca
collection PubMed
description Short implants were introduced to reduce morbidity, treatment duration, and complex bone regeneration interventions in atrophic jaws and to improve patient-reported outcomes. This study aimed to determine the insertion torque (IT), removal torque (RT), and resonance frequency analysis (RFA) values of ultrashort (3 mm length), short (7 mm length), and standard implants (10 mm length) inserted in 1-, 2-, 3-, and 4-mm thickness polyurethane sheets with densities of 10, 20, and 30 pounds per cubic foot (PCF). Standard-length implants were the gold standard (control). Overall, short-length implant IT values were higher or similar to the control in most experimental conditions. Those inserted into a 3 mm/30 PCF lamina showed the highest IT values, whereas 5 mm diameter ultrashort-length implants inserted into 2 and 3 mm/20 PCF laminas were higher than other implants. RT values followed the same trend and RFA values were more appreciable in short- and standard-length implants in all the scenarios. However, ultrashort-length implants reached a primary stability comparable to that of standard implants in lower thicknesses. In conclusion, although further studies are needed to corroborate this in vitro model with preclinical and clinical studies, our data shed light on short- and ultrashort-length implants geometries to a potential application in critical atrophy of the posterior jaws.
format Online
Article
Text
id pubmed-9866818
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-98668182023-01-22 Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets Comuzzi, Luca Tumedei, Margherita Romasco, Tea Petrini, Morena Afrashtehfar, Kelvin I. Inchingolo, Francesco Piattelli, Adriano Di Pietro, Natalia J Funct Biomater Article Short implants were introduced to reduce morbidity, treatment duration, and complex bone regeneration interventions in atrophic jaws and to improve patient-reported outcomes. This study aimed to determine the insertion torque (IT), removal torque (RT), and resonance frequency analysis (RFA) values of ultrashort (3 mm length), short (7 mm length), and standard implants (10 mm length) inserted in 1-, 2-, 3-, and 4-mm thickness polyurethane sheets with densities of 10, 20, and 30 pounds per cubic foot (PCF). Standard-length implants were the gold standard (control). Overall, short-length implant IT values were higher or similar to the control in most experimental conditions. Those inserted into a 3 mm/30 PCF lamina showed the highest IT values, whereas 5 mm diameter ultrashort-length implants inserted into 2 and 3 mm/20 PCF laminas were higher than other implants. RT values followed the same trend and RFA values were more appreciable in short- and standard-length implants in all the scenarios. However, ultrashort-length implants reached a primary stability comparable to that of standard implants in lower thicknesses. In conclusion, although further studies are needed to corroborate this in vitro model with preclinical and clinical studies, our data shed light on short- and ultrashort-length implants geometries to a potential application in critical atrophy of the posterior jaws. MDPI 2022-12-23 /pmc/articles/PMC9866818/ /pubmed/36662057 http://dx.doi.org/10.3390/jfb14010010 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Comuzzi, Luca
Tumedei, Margherita
Romasco, Tea
Petrini, Morena
Afrashtehfar, Kelvin I.
Inchingolo, Francesco
Piattelli, Adriano
Di Pietro, Natalia
Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets
title Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets
title_full Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets
title_fullStr Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets
title_full_unstemmed Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets
title_short Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets
title_sort insertion torque, removal torque, and resonance frequency analysis values of ultrashort, short, and standard dental implants: an in vitro study on polyurethane foam sheets
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9866818/
https://www.ncbi.nlm.nih.gov/pubmed/36662057
http://dx.doi.org/10.3390/jfb14010010
work_keys_str_mv AT comuzziluca insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets
AT tumedeimargherita insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets
AT romascotea insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets
AT petrinimorena insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets
AT afrashtehfarkelvini insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets
AT inchingolofrancesco insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets
AT piattelliadriano insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets
AT dipietronatalia insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets