Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets
Short implants were introduced to reduce morbidity, treatment duration, and complex bone regeneration interventions in atrophic jaws and to improve patient-reported outcomes. This study aimed to determine the insertion torque (IT), removal torque (RT), and resonance frequency analysis (RFA) values o...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9866818/ https://www.ncbi.nlm.nih.gov/pubmed/36662057 http://dx.doi.org/10.3390/jfb14010010 |
_version_ | 1784876186280656896 |
---|---|
author | Comuzzi, Luca Tumedei, Margherita Romasco, Tea Petrini, Morena Afrashtehfar, Kelvin I. Inchingolo, Francesco Piattelli, Adriano Di Pietro, Natalia |
author_facet | Comuzzi, Luca Tumedei, Margherita Romasco, Tea Petrini, Morena Afrashtehfar, Kelvin I. Inchingolo, Francesco Piattelli, Adriano Di Pietro, Natalia |
author_sort | Comuzzi, Luca |
collection | PubMed |
description | Short implants were introduced to reduce morbidity, treatment duration, and complex bone regeneration interventions in atrophic jaws and to improve patient-reported outcomes. This study aimed to determine the insertion torque (IT), removal torque (RT), and resonance frequency analysis (RFA) values of ultrashort (3 mm length), short (7 mm length), and standard implants (10 mm length) inserted in 1-, 2-, 3-, and 4-mm thickness polyurethane sheets with densities of 10, 20, and 30 pounds per cubic foot (PCF). Standard-length implants were the gold standard (control). Overall, short-length implant IT values were higher or similar to the control in most experimental conditions. Those inserted into a 3 mm/30 PCF lamina showed the highest IT values, whereas 5 mm diameter ultrashort-length implants inserted into 2 and 3 mm/20 PCF laminas were higher than other implants. RT values followed the same trend and RFA values were more appreciable in short- and standard-length implants in all the scenarios. However, ultrashort-length implants reached a primary stability comparable to that of standard implants in lower thicknesses. In conclusion, although further studies are needed to corroborate this in vitro model with preclinical and clinical studies, our data shed light on short- and ultrashort-length implants geometries to a potential application in critical atrophy of the posterior jaws. |
format | Online Article Text |
id | pubmed-9866818 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-98668182023-01-22 Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets Comuzzi, Luca Tumedei, Margherita Romasco, Tea Petrini, Morena Afrashtehfar, Kelvin I. Inchingolo, Francesco Piattelli, Adriano Di Pietro, Natalia J Funct Biomater Article Short implants were introduced to reduce morbidity, treatment duration, and complex bone regeneration interventions in atrophic jaws and to improve patient-reported outcomes. This study aimed to determine the insertion torque (IT), removal torque (RT), and resonance frequency analysis (RFA) values of ultrashort (3 mm length), short (7 mm length), and standard implants (10 mm length) inserted in 1-, 2-, 3-, and 4-mm thickness polyurethane sheets with densities of 10, 20, and 30 pounds per cubic foot (PCF). Standard-length implants were the gold standard (control). Overall, short-length implant IT values were higher or similar to the control in most experimental conditions. Those inserted into a 3 mm/30 PCF lamina showed the highest IT values, whereas 5 mm diameter ultrashort-length implants inserted into 2 and 3 mm/20 PCF laminas were higher than other implants. RT values followed the same trend and RFA values were more appreciable in short- and standard-length implants in all the scenarios. However, ultrashort-length implants reached a primary stability comparable to that of standard implants in lower thicknesses. In conclusion, although further studies are needed to corroborate this in vitro model with preclinical and clinical studies, our data shed light on short- and ultrashort-length implants geometries to a potential application in critical atrophy of the posterior jaws. MDPI 2022-12-23 /pmc/articles/PMC9866818/ /pubmed/36662057 http://dx.doi.org/10.3390/jfb14010010 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Comuzzi, Luca Tumedei, Margherita Romasco, Tea Petrini, Morena Afrashtehfar, Kelvin I. Inchingolo, Francesco Piattelli, Adriano Di Pietro, Natalia Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets |
title | Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets |
title_full | Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets |
title_fullStr | Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets |
title_full_unstemmed | Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets |
title_short | Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets |
title_sort | insertion torque, removal torque, and resonance frequency analysis values of ultrashort, short, and standard dental implants: an in vitro study on polyurethane foam sheets |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9866818/ https://www.ncbi.nlm.nih.gov/pubmed/36662057 http://dx.doi.org/10.3390/jfb14010010 |
work_keys_str_mv | AT comuzziluca insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets AT tumedeimargherita insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets AT romascotea insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets AT petrinimorena insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets AT afrashtehfarkelvini insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets AT inchingolofrancesco insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets AT piattelliadriano insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets AT dipietronatalia insertiontorqueremovaltorqueandresonancefrequencyanalysisvaluesofultrashortshortandstandarddentalimplantsaninvitrostudyonpolyurethanefoamsheets |