Cargando…
A Custom Device to Couple Proximal and Distal Femoral Stems: A Case Report and Literature Review
Complex periprosthetic fractures may occur with existing ipsilateral hip and knee implants. These cases are challenging with few options. We report on a custom coupler used to connect a hinged distal femoral replacement to an existing revision hip stem. This option was preferred to a total replaceme...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9873500/ https://www.ncbi.nlm.nih.gov/pubmed/36713938 http://dx.doi.org/10.1016/j.artd.2022.101089 |
Sumario: | Complex periprosthetic fractures may occur with existing ipsilateral hip and knee implants. These cases are challenging with few options. We report on a custom coupler used to connect a hinged distal femoral replacement to an existing revision hip stem. This option was preferred to a total replacement of the femur. The custom coupler maintained the anatomic attachments of the proximal femur, preserved abductor strength during ambulation, and mitigated the morbidity which may arise from a total femur replacement. At 15 months postoperatively, the patient reported no pain, did not demonstrate an antalgic gait pattern, and had satisfactory range of motion at the hip and knee. The literature displays clinical and biomechanical efficacy for coupling devices although complications and device failure are quite variable. |
---|