Evaluating an advanced double intravenous vasopressor automated system to treat hypotension during spinal anesthesia for cesarean delivery: a randomized controlled trial
BACKGROUND: The optimal treatment of hypotension during spinal anaesthesia is uncertain. A novel double intravenous vasopressor automated (DIVA) system reduces hypotension compared to standard care, and was subsequently modified to an advanced-DIVA (ADIVA) system. The primary objective was to compar...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9878794/ https://www.ncbi.nlm.nih.gov/pubmed/36703120 http://dx.doi.org/10.1186/s12871-023-01992-7 |
_version_ | 1784878565250039808 |
---|---|
author | Tan, Hon Sen Nagarajan, Singaraselvan Chan, Jason Ju In Tan, Chin Wen Sultana, Rehena Sia, Alex Tiong Heng Sng, Ban Leong |
author_facet | Tan, Hon Sen Nagarajan, Singaraselvan Chan, Jason Ju In Tan, Chin Wen Sultana, Rehena Sia, Alex Tiong Heng Sng, Ban Leong |
author_sort | Tan, Hon Sen |
collection | PubMed |
description | BACKGROUND: The optimal treatment of hypotension during spinal anaesthesia is uncertain. A novel double intravenous vasopressor automated (DIVA) system reduces hypotension compared to standard care, and was subsequently modified to an advanced-DIVA (ADIVA) system. The primary objective was to compare ADIVA versus DIVA on incidence of hypotension (systolic BP (SBP) < 80% baseline). METHODS: We conducted a randomized-controlled trial in women undergoing elective cesarean delivery under spinal anesthesia. SBP and heart rate were measured continuously using a Nexfin monitor. ADIVA delivered 25 μg phenylephrine (heart rate > 60 beats.min(−1)) or 2 mg ephedrine (heart rate < 60 beats.min(−1)) at SBP 90 to 110% of baseline, 50 μg phenylephrine or 4 mg ephedrine at SBP 80 to 90%, and 75 μg phenylephrine or 6 mg ephedrine at SBP < 80%. ADIVA calculated the trend of SBP; vasopressors were administered rapidly if SBP trended downward, or 30 s if SBP trended upward. In contrast, DIVA delivered 25 μg phenylephrine or 2 mg ephedrine at SBP 90 to 100% of baseline, and 50 μg phenylephrine or 4 mg ephedrine at SBP < 90%. Boluses were followed by a 10-s lockout. Other outcomes included hypertension (SBP > 120% baseline), vasopressor consumption, clinical outcomes, and performance measures from spinal anesthesia to fetal delivery. RESULTS: We analyzed 94 parturients (ADIVA: n = 46, DIVA: n = 48), with no difference in the incidence of hypotension between ADIVA (78.3%) and DIVA (83.3%, p = 0.677). ADIVA had significantly higher proportion of hypotensive SBP readings, lower phenylephrine consumption and higher umbilical arterial pH. There was no difference in hypertension, bradycardia, ephedrine consumption, intravenous fluid volume, nausea/vomiting, Apgar scores, and umbilical venous pH or lactate. ADIVA maintained SBP higher above baseline with greater fluctuation than DIVA. CONCLUSION: ADIVA was associated with a greater proportion of hypotensive SBP readings, reduced phenylephrine consumption, and increased umbilical arterial pH than DIVA. Further research is needed to determine the optimal method of vasopressor delivery in parturients undergoing cesarean delivery. TRIAL REGISTRATION: This study was registered on Clinicaltrials.gov registry (NCT03620942) on 08/08/2018. |
format | Online Article Text |
id | pubmed-9878794 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-98787942023-01-27 Evaluating an advanced double intravenous vasopressor automated system to treat hypotension during spinal anesthesia for cesarean delivery: a randomized controlled trial Tan, Hon Sen Nagarajan, Singaraselvan Chan, Jason Ju In Tan, Chin Wen Sultana, Rehena Sia, Alex Tiong Heng Sng, Ban Leong BMC Anesthesiol Research BACKGROUND: The optimal treatment of hypotension during spinal anaesthesia is uncertain. A novel double intravenous vasopressor automated (DIVA) system reduces hypotension compared to standard care, and was subsequently modified to an advanced-DIVA (ADIVA) system. The primary objective was to compare ADIVA versus DIVA on incidence of hypotension (systolic BP (SBP) < 80% baseline). METHODS: We conducted a randomized-controlled trial in women undergoing elective cesarean delivery under spinal anesthesia. SBP and heart rate were measured continuously using a Nexfin monitor. ADIVA delivered 25 μg phenylephrine (heart rate > 60 beats.min(−1)) or 2 mg ephedrine (heart rate < 60 beats.min(−1)) at SBP 90 to 110% of baseline, 50 μg phenylephrine or 4 mg ephedrine at SBP 80 to 90%, and 75 μg phenylephrine or 6 mg ephedrine at SBP < 80%. ADIVA calculated the trend of SBP; vasopressors were administered rapidly if SBP trended downward, or 30 s if SBP trended upward. In contrast, DIVA delivered 25 μg phenylephrine or 2 mg ephedrine at SBP 90 to 100% of baseline, and 50 μg phenylephrine or 4 mg ephedrine at SBP < 90%. Boluses were followed by a 10-s lockout. Other outcomes included hypertension (SBP > 120% baseline), vasopressor consumption, clinical outcomes, and performance measures from spinal anesthesia to fetal delivery. RESULTS: We analyzed 94 parturients (ADIVA: n = 46, DIVA: n = 48), with no difference in the incidence of hypotension between ADIVA (78.3%) and DIVA (83.3%, p = 0.677). ADIVA had significantly higher proportion of hypotensive SBP readings, lower phenylephrine consumption and higher umbilical arterial pH. There was no difference in hypertension, bradycardia, ephedrine consumption, intravenous fluid volume, nausea/vomiting, Apgar scores, and umbilical venous pH or lactate. ADIVA maintained SBP higher above baseline with greater fluctuation than DIVA. CONCLUSION: ADIVA was associated with a greater proportion of hypotensive SBP readings, reduced phenylephrine consumption, and increased umbilical arterial pH than DIVA. Further research is needed to determine the optimal method of vasopressor delivery in parturients undergoing cesarean delivery. TRIAL REGISTRATION: This study was registered on Clinicaltrials.gov registry (NCT03620942) on 08/08/2018. BioMed Central 2023-01-26 /pmc/articles/PMC9878794/ /pubmed/36703120 http://dx.doi.org/10.1186/s12871-023-01992-7 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Tan, Hon Sen Nagarajan, Singaraselvan Chan, Jason Ju In Tan, Chin Wen Sultana, Rehena Sia, Alex Tiong Heng Sng, Ban Leong Evaluating an advanced double intravenous vasopressor automated system to treat hypotension during spinal anesthesia for cesarean delivery: a randomized controlled trial |
title | Evaluating an advanced double intravenous vasopressor automated system to treat hypotension during spinal anesthesia for cesarean delivery: a randomized controlled trial |
title_full | Evaluating an advanced double intravenous vasopressor automated system to treat hypotension during spinal anesthesia for cesarean delivery: a randomized controlled trial |
title_fullStr | Evaluating an advanced double intravenous vasopressor automated system to treat hypotension during spinal anesthesia for cesarean delivery: a randomized controlled trial |
title_full_unstemmed | Evaluating an advanced double intravenous vasopressor automated system to treat hypotension during spinal anesthesia for cesarean delivery: a randomized controlled trial |
title_short | Evaluating an advanced double intravenous vasopressor automated system to treat hypotension during spinal anesthesia for cesarean delivery: a randomized controlled trial |
title_sort | evaluating an advanced double intravenous vasopressor automated system to treat hypotension during spinal anesthesia for cesarean delivery: a randomized controlled trial |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9878794/ https://www.ncbi.nlm.nih.gov/pubmed/36703120 http://dx.doi.org/10.1186/s12871-023-01992-7 |
work_keys_str_mv | AT tanhonsen evaluatinganadvanceddoubleintravenousvasopressorautomatedsystemtotreathypotensionduringspinalanesthesiaforcesareandeliveryarandomizedcontrolledtrial AT nagarajansingaraselvan evaluatinganadvanceddoubleintravenousvasopressorautomatedsystemtotreathypotensionduringspinalanesthesiaforcesareandeliveryarandomizedcontrolledtrial AT chanjasonjuin evaluatinganadvanceddoubleintravenousvasopressorautomatedsystemtotreathypotensionduringspinalanesthesiaforcesareandeliveryarandomizedcontrolledtrial AT tanchinwen evaluatinganadvanceddoubleintravenousvasopressorautomatedsystemtotreathypotensionduringspinalanesthesiaforcesareandeliveryarandomizedcontrolledtrial AT sultanarehena evaluatinganadvanceddoubleintravenousvasopressorautomatedsystemtotreathypotensionduringspinalanesthesiaforcesareandeliveryarandomizedcontrolledtrial AT siaalextiongheng evaluatinganadvanceddoubleintravenousvasopressorautomatedsystemtotreathypotensionduringspinalanesthesiaforcesareandeliveryarandomizedcontrolledtrial AT sngbanleong evaluatinganadvanceddoubleintravenousvasopressorautomatedsystemtotreathypotensionduringspinalanesthesiaforcesareandeliveryarandomizedcontrolledtrial |