Corneal thickness measurements with the RTVue, Casia-2, and Pentacam devices in patients with mild-to-moderate keratoconus: a comparative study

BACKGROUND: To compare the characteristics of corneal thickness measurements among the RTVue, Casia-2, and Pentacam in patients with mild-to-moderate keratoconus. METHODS: We recruited 46 eyes of 46 patients diagnosed with mild-to-moderate keratoconus at our hospital between January and March 2022....

Descripción completa

Detalles Bibliográficos
Autores principales: Sun, Bingqing, Zhang, Xiaoyu, Sun, Ling, Huang, Yangyi, Tian, Mi, Shen, Yang, Ding, Lan, Zhou, Xingtao
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9878798/
https://www.ncbi.nlm.nih.gov/pubmed/36703165
http://dx.doi.org/10.1186/s12886-023-02767-x
_version_ 1784878566256672768
author Sun, Bingqing
Zhang, Xiaoyu
Sun, Ling
Huang, Yangyi
Tian, Mi
Shen, Yang
Ding, Lan
Zhou, Xingtao
author_facet Sun, Bingqing
Zhang, Xiaoyu
Sun, Ling
Huang, Yangyi
Tian, Mi
Shen, Yang
Ding, Lan
Zhou, Xingtao
author_sort Sun, Bingqing
collection PubMed
description BACKGROUND: To compare the characteristics of corneal thickness measurements among the RTVue, Casia-2, and Pentacam in patients with mild-to-moderate keratoconus. METHODS: We recruited 46 eyes of 46 patients diagnosed with mild-to-moderate keratoconus at our hospital between January and March 2022. The central corneal thickness (CCT) and thinnest corneal thickness (TCT) were measured using two optical coherence tomography (OCT) instruments (RTVue and Casia-2) and the more conventional Pentacam. Differences and correlations between the CCTs and TCTs, based on the device and influencing factors, were explored. RESULTS: The CCTs were highly consistent among the groups (p = 0.434) and correlated with one another (p < 0.001). The TCTs measured by OCTs were thinner than those measured by the Pentacam (p < 0.001); however, all three devices were highly correlated (p < 0.001). The thinnest point location measurements with RTVue and Casia-2 differed significantly from the measurements with the Pentacam. Bland–Altman plots demonstrated a significant agreement between Pentacam and OCTs in TCT measurement (p < 0.001); the 95% limits of agreement were − 3.1 μm to + 33.1 μm for Pentacam and RTVue and − 8.6 μm to + 36.5 μm for Pentacam and Casia-2. RTVue and Casia-2 showed no difference in corneal thickness (p = 0.633) and thinnest point location measurement (p > 0.05). Multivariate analysis identified that the TCT measurement difference between the RTVue and Pentacam was related to the difference between the CCT and TCT (b = 0.490, 95% confidence interval [CI]: 0.033 to 0.948, p = 0.036), whereas the difference between the Casia-2 and Pentacam was related to the anterior radius for curvature (A) grade (b = 3.9, 95% CI: 1.753 to 6.074, p = 0.001), corneal pachymetry at the thinnest (C) grade (b = − 7.875, 95% CI: − 11.404 to − 4.346, p < 0.001), and the difference between the CCT and TCT (b = 0.425, 95% CI: 0.1 to 0.751, p = 0.012). CONCLUSIONS: CCTs in patients with mild-to-moderate keratoconus were similar among all three devices, but the TCTs and the thinnest point locations were not. Furthermore, the TCT measurement differences between the OCT devices and the Pentacam were more pronounced in keratoconus cases with a steeper anterior surface, thicker TCTs, and a larger difference between the CCT and TCT. TRIAL REGISTRATION: Number: 2021118–1. Retrospectively registered: September 01, 2021. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12886-023-02767-x.
format Online
Article
Text
id pubmed-9878798
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-98787982023-01-27 Corneal thickness measurements with the RTVue, Casia-2, and Pentacam devices in patients with mild-to-moderate keratoconus: a comparative study Sun, Bingqing Zhang, Xiaoyu Sun, Ling Huang, Yangyi Tian, Mi Shen, Yang Ding, Lan Zhou, Xingtao BMC Ophthalmol Research BACKGROUND: To compare the characteristics of corneal thickness measurements among the RTVue, Casia-2, and Pentacam in patients with mild-to-moderate keratoconus. METHODS: We recruited 46 eyes of 46 patients diagnosed with mild-to-moderate keratoconus at our hospital between January and March 2022. The central corneal thickness (CCT) and thinnest corneal thickness (TCT) were measured using two optical coherence tomography (OCT) instruments (RTVue and Casia-2) and the more conventional Pentacam. Differences and correlations between the CCTs and TCTs, based on the device and influencing factors, were explored. RESULTS: The CCTs were highly consistent among the groups (p = 0.434) and correlated with one another (p < 0.001). The TCTs measured by OCTs were thinner than those measured by the Pentacam (p < 0.001); however, all three devices were highly correlated (p < 0.001). The thinnest point location measurements with RTVue and Casia-2 differed significantly from the measurements with the Pentacam. Bland–Altman plots demonstrated a significant agreement between Pentacam and OCTs in TCT measurement (p < 0.001); the 95% limits of agreement were − 3.1 μm to + 33.1 μm for Pentacam and RTVue and − 8.6 μm to + 36.5 μm for Pentacam and Casia-2. RTVue and Casia-2 showed no difference in corneal thickness (p = 0.633) and thinnest point location measurement (p > 0.05). Multivariate analysis identified that the TCT measurement difference between the RTVue and Pentacam was related to the difference between the CCT and TCT (b = 0.490, 95% confidence interval [CI]: 0.033 to 0.948, p = 0.036), whereas the difference between the Casia-2 and Pentacam was related to the anterior radius for curvature (A) grade (b = 3.9, 95% CI: 1.753 to 6.074, p = 0.001), corneal pachymetry at the thinnest (C) grade (b = − 7.875, 95% CI: − 11.404 to − 4.346, p < 0.001), and the difference between the CCT and TCT (b = 0.425, 95% CI: 0.1 to 0.751, p = 0.012). CONCLUSIONS: CCTs in patients with mild-to-moderate keratoconus were similar among all three devices, but the TCTs and the thinnest point locations were not. Furthermore, the TCT measurement differences between the OCT devices and the Pentacam were more pronounced in keratoconus cases with a steeper anterior surface, thicker TCTs, and a larger difference between the CCT and TCT. TRIAL REGISTRATION: Number: 2021118–1. Retrospectively registered: September 01, 2021. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12886-023-02767-x. BioMed Central 2023-01-26 /pmc/articles/PMC9878798/ /pubmed/36703165 http://dx.doi.org/10.1186/s12886-023-02767-x Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Sun, Bingqing
Zhang, Xiaoyu
Sun, Ling
Huang, Yangyi
Tian, Mi
Shen, Yang
Ding, Lan
Zhou, Xingtao
Corneal thickness measurements with the RTVue, Casia-2, and Pentacam devices in patients with mild-to-moderate keratoconus: a comparative study
title Corneal thickness measurements with the RTVue, Casia-2, and Pentacam devices in patients with mild-to-moderate keratoconus: a comparative study
title_full Corneal thickness measurements with the RTVue, Casia-2, and Pentacam devices in patients with mild-to-moderate keratoconus: a comparative study
title_fullStr Corneal thickness measurements with the RTVue, Casia-2, and Pentacam devices in patients with mild-to-moderate keratoconus: a comparative study
title_full_unstemmed Corneal thickness measurements with the RTVue, Casia-2, and Pentacam devices in patients with mild-to-moderate keratoconus: a comparative study
title_short Corneal thickness measurements with the RTVue, Casia-2, and Pentacam devices in patients with mild-to-moderate keratoconus: a comparative study
title_sort corneal thickness measurements with the rtvue, casia-2, and pentacam devices in patients with mild-to-moderate keratoconus: a comparative study
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9878798/
https://www.ncbi.nlm.nih.gov/pubmed/36703165
http://dx.doi.org/10.1186/s12886-023-02767-x
work_keys_str_mv AT sunbingqing cornealthicknessmeasurementswiththertvuecasia2andpentacamdevicesinpatientswithmildtomoderatekeratoconusacomparativestudy
AT zhangxiaoyu cornealthicknessmeasurementswiththertvuecasia2andpentacamdevicesinpatientswithmildtomoderatekeratoconusacomparativestudy
AT sunling cornealthicknessmeasurementswiththertvuecasia2andpentacamdevicesinpatientswithmildtomoderatekeratoconusacomparativestudy
AT huangyangyi cornealthicknessmeasurementswiththertvuecasia2andpentacamdevicesinpatientswithmildtomoderatekeratoconusacomparativestudy
AT tianmi cornealthicknessmeasurementswiththertvuecasia2andpentacamdevicesinpatientswithmildtomoderatekeratoconusacomparativestudy
AT shenyang cornealthicknessmeasurementswiththertvuecasia2andpentacamdevicesinpatientswithmildtomoderatekeratoconusacomparativestudy
AT dinglan cornealthicknessmeasurementswiththertvuecasia2andpentacamdevicesinpatientswithmildtomoderatekeratoconusacomparativestudy
AT zhouxingtao cornealthicknessmeasurementswiththertvuecasia2andpentacamdevicesinpatientswithmildtomoderatekeratoconusacomparativestudy