Cargando…

Measurement of Lag-Screw Anteversion With an iPhone During Trochanteric Fracture Surgery

Introduction  A useful way to easily evaluate femoral rotation during surgery for trochanteric fractures is not known. Hence, this pilot study aimed to develop an intraoperative indicator to evaluate anteversion in femoral trochanteric fractures. Material and methods Prospectively, from June 2021 to...

Descripción completa

Detalles Bibliográficos
Autores principales: Kinami, Yo, Hasei, Joe, Fujiwara, Kazuo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9884309/
https://www.ncbi.nlm.nih.gov/pubmed/36721595
http://dx.doi.org/10.7759/cureus.33110
Descripción
Sumario:Introduction  A useful way to easily evaluate femoral rotation during surgery for trochanteric fractures is not known. Hence, this pilot study aimed to develop an intraoperative indicator to evaluate anteversion in femoral trochanteric fractures. Material and methods Prospectively, from June 2021 to January 2022, all patients with femoral trochanteric fractures (Orthopaedic Trauma Association classification: 31A1-3) treated using a cephalo-medullary nail with a lag-screw neck-shaft angle of 125° were included in this study. During surgery, lag-screw anteversion (LS-AV) was measured using the goniometer application in an iPhone with the fractured femur table-top-plane level with the traction table floor. Accuracy was analyzed by comparing axial-projected lag-screw anteversion (AxP-LS-AV) and three-dimensional computed tomography lag-screw anteversion (3DCT-LS-AV) measurements after surgery. Results Fifty patients (14 males and 36 females) were included in the study. The mean age was 87 (range; 69-98) years; the Orthopaedic Trauma Association classifications were A1 (28 patients), A2 (18 patients), and A3 (4 patients). The mean LS-AV was 10.7° ± 6.9°, the mean AxP-LS-AV was 12.8° ± 8.3°, and the mean 3DCT-LS-AV was 13.1° ± 8.6°. The median difference between AxP-LS-AV and 3DCT-LS-AV was 3.0° (range: 0°-12°), and 40 (80%) patients had differences of ≤5° (Bland-Altman plot: inside of limit of agreement = 86%, paired t-test p = 0.7, Pearson correlation coefficient r = 0.817, p <0.001). Conclusion Femur malrotation is defined as a deformity of >15° relative to the normal contralateral limb. Intraoperative LS-AV iPhone measurement on table-top-plane standard had sufficient accuracy as an indicator of anteversion in femoral trochanteric fractures.