Invasive validation of the Antares algorithm for determining central blood pressure based on upper arm oscillometric pulse waves in patients with type 2 diabetes
INTRODUCTION: Antares is a pulse wave analysis (PWA) algorithm designed to allow a non-invasive estimation of central (aortic) blood pressure (cBP) using automated oscillometric blood pressure (BP) devices. Diabetes may affect elastic and muscular arteries differently, resulting in disparate pulse w...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BMJ Publishing Group
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9884863/ https://www.ncbi.nlm.nih.gov/pubmed/36707104 http://dx.doi.org/10.1136/bmjdrc-2022-003119 |
Sumario: | INTRODUCTION: Antares is a pulse wave analysis (PWA) algorithm designed to allow a non-invasive estimation of central (aortic) blood pressure (cBP) using automated oscillometric blood pressure (BP) devices. Diabetes may affect elastic and muscular arteries differently, resulting in disparate pulse wave characteristics in central and peripheral arteries, which may limit the accuracy of PWA devices. The aim of our study was to evaluate the accuracy of Antares for estimating cBP as compared with invasively measured cBP in patients with type 2 diabetes. RESEARCH DESIGN AND METHODS: In this study, consecutive patients undergoing elective coronary angiography were recruited between November 2017 and September 2020. In 119 patients with type 2 diabetes, cBP was measured invasively and simultaneously determined non-invasively using the custo screen 400 device with the integrated Antares algorithm. RESULTS: The mean difference between the estimated and invasively measured cBP was 1.2±6.3 mmHg for central systolic BP (cSBP), 1.0±4.3 mmHg for central mean arterial pressure (cMAP) and 3.6±5.7 mmHg for central diastolic BP (cDBP). High correlations were found between estimated cBP and invasively measured cBP (cSBP: r=0.916; cMAP: r=0.882; cDBP: r=0.791; all p<0.001). CONCLUSIONS: The present study suggests that the Antares algorithm incorporated into the custo screen 400 device can estimate cBP with high accuracy turning a conventional oscillometric BP device into a type II device for the non-invasive estimation of cBP, which is applicable in patients with type 2 diabetes. Integration of Antares into commercially available BP devices could facilitate the introduction of cBP into routine clinical practice as a part of disease and risk management. |
---|