System‐dependent image distortion related to gantry positions of a 0.35 T MRgRT: Characterization and the corresponding correction

PURPOSE: MR‐guided radiotherapy with high accuracy treatment planning requires addressing MR imaging artifacts that originate from system imperfections. This work presents the characterization and corresponding correction of gantry‐related imaging distortions including geometric distortion and isoce...

Descripción completa

Detalles Bibliográficos
Autores principales: Marasini, Shanti, Quinn, Benjamin, Cole, Mike, Flores, Rocco, Kim, Taeho
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9903949/
https://www.ncbi.nlm.nih.gov/pubmed/36354747
http://dx.doi.org/10.1002/acm2.13826
_version_ 1784883542152445952
author Marasini, Shanti
Quinn, Benjamin
Cole, Mike
Flores, Rocco
Kim, Taeho
author_facet Marasini, Shanti
Quinn, Benjamin
Cole, Mike
Flores, Rocco
Kim, Taeho
author_sort Marasini, Shanti
collection PubMed
description PURPOSE: MR‐guided radiotherapy with high accuracy treatment planning requires addressing MR imaging artifacts that originate from system imperfections. This work presents the characterization and corresponding correction of gantry‐related imaging distortions including geometric distortion and isocenter shift in a 0.35 T magnetic resonance imaging (MRI)‐guided radiotherapy (MRgRT) system using distortion vector fields (DVFs). METHODS: Two phantoms, the magnetic resonance imaging distortion in 3D (MRID(3D)) phantom and the Fluke phantom, along with a human volunteer were imaged at different gantry angles on a 0.35 T MR‐Linac. The geometric distortion and isocenter shift were characterized for both phantom images. DVFs with a field of view extended beyond the physical boundary of the MRID(3D) phantom were extracted from images taken at 30° gantry angle increments, with vendor‐provided distortion correction turned on and off (DstOff). These extended DVFs were then applied to the relevant phantom images to correct their geometric distortions and isocenter shift at the respective gantry angles. The extended DVFs produced from the MRID(3D) phantom were also applied to Fluke phantom and human MR images at their respective gantry angles. The resampled images were evaluated using structural similarity index measure (SSIM) comparison with the vendor corrected images from the MRgRT system. RESULTS: Geometric distortion with “mean (± SD) distortion” of 3.2 ± 0.02, 2.9 ± 0.02, and 1.8 ± 0.01 mm and isocenter shift (±SD) of 0.49 ± 0.3, 0.05 ± 0.2, and 0.01 ± 0.03 mm were present in the DstOff MRID(3D) phantom images in right–left (RL), anterior–posterior (AP), and superior–inferior (SI) directions, respectively. After resampling the originally acquired images by applying extended DVFs, the distortion was corrected to 0.18 ± 0.02, 0.09 ± 0.01, 0.15 ± 0.01 mm, and isocenter shift was corrected to 0.14 ± 0.05, −0.02 ± 0.04, and −0.07 ± 0.05 mm in RL, AP, and SI directions, respectively. The Fluke phantom average geometric distortion with “mean (± SD) distortion” of 2.7 ± 0.1 mm was corrected to 0.2 ± 0. 1 mm and the average isocenter shift (± SD) of 0.51 ± 0.2 mm, and 0.05 ± 0.03 was corrected to −0.08 ± 0.03 mm, and −0.05 ± 0.01 in RL and AP directions, respectively. SSIM (mean ± SD) of the original images to resampled images was increased from 0.49 ± 0.02 to 0.78 ± 0.01, 0.45 ± 0.02 to 0.75 ± 0.01, and 0.86 ± 0.25 to 0.98 ± 0.08 for MRID(3D) phantom, Fluke phantom, and human MR images, respectively, for all the gantry angles compared to the vendor corrected images. CONCLUSION: The gantry‐related MR imaging distortion including geometric distortion and isocenter shift was characterized and a corresponding correction was demonstrated using extended DVFs on 0.35 T MRgRT system. The characterized gantry‐related isocenter shift can be combined with geometric distortion correction to provide a technique for the correction of the full system‐dependent distortion in an MRgRT system.
format Online
Article
Text
id pubmed-9903949
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-99039492023-02-09 System‐dependent image distortion related to gantry positions of a 0.35 T MRgRT: Characterization and the corresponding correction Marasini, Shanti Quinn, Benjamin Cole, Mike Flores, Rocco Kim, Taeho J Appl Clin Med Phys Medical Imaging PURPOSE: MR‐guided radiotherapy with high accuracy treatment planning requires addressing MR imaging artifacts that originate from system imperfections. This work presents the characterization and corresponding correction of gantry‐related imaging distortions including geometric distortion and isocenter shift in a 0.35 T magnetic resonance imaging (MRI)‐guided radiotherapy (MRgRT) system using distortion vector fields (DVFs). METHODS: Two phantoms, the magnetic resonance imaging distortion in 3D (MRID(3D)) phantom and the Fluke phantom, along with a human volunteer were imaged at different gantry angles on a 0.35 T MR‐Linac. The geometric distortion and isocenter shift were characterized for both phantom images. DVFs with a field of view extended beyond the physical boundary of the MRID(3D) phantom were extracted from images taken at 30° gantry angle increments, with vendor‐provided distortion correction turned on and off (DstOff). These extended DVFs were then applied to the relevant phantom images to correct their geometric distortions and isocenter shift at the respective gantry angles. The extended DVFs produced from the MRID(3D) phantom were also applied to Fluke phantom and human MR images at their respective gantry angles. The resampled images were evaluated using structural similarity index measure (SSIM) comparison with the vendor corrected images from the MRgRT system. RESULTS: Geometric distortion with “mean (± SD) distortion” of 3.2 ± 0.02, 2.9 ± 0.02, and 1.8 ± 0.01 mm and isocenter shift (±SD) of 0.49 ± 0.3, 0.05 ± 0.2, and 0.01 ± 0.03 mm were present in the DstOff MRID(3D) phantom images in right–left (RL), anterior–posterior (AP), and superior–inferior (SI) directions, respectively. After resampling the originally acquired images by applying extended DVFs, the distortion was corrected to 0.18 ± 0.02, 0.09 ± 0.01, 0.15 ± 0.01 mm, and isocenter shift was corrected to 0.14 ± 0.05, −0.02 ± 0.04, and −0.07 ± 0.05 mm in RL, AP, and SI directions, respectively. The Fluke phantom average geometric distortion with “mean (± SD) distortion” of 2.7 ± 0.1 mm was corrected to 0.2 ± 0. 1 mm and the average isocenter shift (± SD) of 0.51 ± 0.2 mm, and 0.05 ± 0.03 was corrected to −0.08 ± 0.03 mm, and −0.05 ± 0.01 in RL and AP directions, respectively. SSIM (mean ± SD) of the original images to resampled images was increased from 0.49 ± 0.02 to 0.78 ± 0.01, 0.45 ± 0.02 to 0.75 ± 0.01, and 0.86 ± 0.25 to 0.98 ± 0.08 for MRID(3D) phantom, Fluke phantom, and human MR images, respectively, for all the gantry angles compared to the vendor corrected images. CONCLUSION: The gantry‐related MR imaging distortion including geometric distortion and isocenter shift was characterized and a corresponding correction was demonstrated using extended DVFs on 0.35 T MRgRT system. The characterized gantry‐related isocenter shift can be combined with geometric distortion correction to provide a technique for the correction of the full system‐dependent distortion in an MRgRT system. John Wiley and Sons Inc. 2022-11-10 /pmc/articles/PMC9903949/ /pubmed/36354747 http://dx.doi.org/10.1002/acm2.13826 Text en © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Marasini, Shanti
Quinn, Benjamin
Cole, Mike
Flores, Rocco
Kim, Taeho
System‐dependent image distortion related to gantry positions of a 0.35 T MRgRT: Characterization and the corresponding correction
title System‐dependent image distortion related to gantry positions of a 0.35 T MRgRT: Characterization and the corresponding correction
title_full System‐dependent image distortion related to gantry positions of a 0.35 T MRgRT: Characterization and the corresponding correction
title_fullStr System‐dependent image distortion related to gantry positions of a 0.35 T MRgRT: Characterization and the corresponding correction
title_full_unstemmed System‐dependent image distortion related to gantry positions of a 0.35 T MRgRT: Characterization and the corresponding correction
title_short System‐dependent image distortion related to gantry positions of a 0.35 T MRgRT: Characterization and the corresponding correction
title_sort system‐dependent image distortion related to gantry positions of a 0.35 t mrgrt: characterization and the corresponding correction
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9903949/
https://www.ncbi.nlm.nih.gov/pubmed/36354747
http://dx.doi.org/10.1002/acm2.13826
work_keys_str_mv AT marasinishanti systemdependentimagedistortionrelatedtogantrypositionsofa035tmrgrtcharacterizationandthecorrespondingcorrection
AT quinnbenjamin systemdependentimagedistortionrelatedtogantrypositionsofa035tmrgrtcharacterizationandthecorrespondingcorrection
AT colemike systemdependentimagedistortionrelatedtogantrypositionsofa035tmrgrtcharacterizationandthecorrespondingcorrection
AT floresrocco systemdependentimagedistortionrelatedtogantrypositionsofa035tmrgrtcharacterizationandthecorrespondingcorrection
AT kimtaeho systemdependentimagedistortionrelatedtogantrypositionsofa035tmrgrtcharacterizationandthecorrespondingcorrection