Visual and quantitative assessment of hip implant-related metal artifacts at low field MRI: a phantom study comparing a 0.55-T system with 1.5-T and 3-T systems

BACKGROUND: To investigate hip implant-related metal artifacts on a 0.55-T system compared with 1.5-T and 3-T systems. METHODS: Total hip arthroplasty made of three different alloys were evaluated in a water phantom at 0.55, 1.5, and 3 T using routine protocols. Visually assessment (VA) was performe...

Descripción completa

Detalles Bibliográficos
Autores principales: Breit, Hanns-Christian, Vosshenrich, Jan, Clauss, Martin, Weikert, Thomas J., Stieltjes, Bram, Kovacs, Balázs K., Bach, Michael, Harder, Dorothee
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Vienna 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9905379/
https://www.ncbi.nlm.nih.gov/pubmed/36750494
http://dx.doi.org/10.1186/s41747-023-00320-5
Descripción
Sumario:BACKGROUND: To investigate hip implant-related metal artifacts on a 0.55-T system compared with 1.5-T and 3-T systems. METHODS: Total hip arthroplasty made of three different alloys were evaluated in a water phantom at 0.55, 1.5, and 3 T using routine protocols. Visually assessment (VA) was performed by three readers using a Likert scale from 0 (no artifacts) to 6 (extremely severe artifacts). Quantitative assessment (QA) was performed using the coefficient of variation (CoV) and the fraction of voxels within a threshold of the mean signal intensity compared to an automatically defined region of interest (FVwT). Agreement was evaluated using intra/inter-class correlation coefficient (ICC). RESULTS: Interreader agreement of VA was strong-to-moderate (ICC 0.74−0.82). At all field strengths (0.55-T/1.5-T/3-T), artifacts were assigned a lower score for titanium (Ti) alloys (2.44/2.9/2.7) than for stainless steel (Fe-Cr) (4.1/3.9/5.1) and cobalt-chromium (Co-Cr) alloys (4.1/4.1/5.2) (p < 0.001 for both). Artifacts were lower for 0.55-T and 1.5-T than for 3-T systems, for all implants (p ≤ 0.049). A strong VA-to-QA correlation was found (r = 0.81; p < 0.001); CoV was lower for Ti alloys than for Fe-Cr and Co-Cr alloys at all field strengths. The FVwT showed a negative correlation with VA (-0.68 < r < -0.84; p < 0.001). CONCLUSIONS: Artifact intensity was lowest for Ti alloys at 0.55 T. For other alloys, it was similar at 0.55 T and 1.5 T, higher at 3 T. Despite an inferior gradient system and a larger bore width, the 0.55-T system showed the same artifact intensity of the 1.5-T system.