Cargando…

Evaluation of the HemoCue® for blood loss estimation in suction fluid in patients undergoing neurosurgical procedures: A prospective observational study

BACKGROUND AND AIMS: Blood loss in neurosurgical procedures can be rapid and tremendous leading to consequential hemodynamic instability. HemoCue is a portable photometer used for the measurement of blood hemoglobin concentration. Using this point of care device, we contemplated this study to assess...

Descripción completa

Detalles Bibliográficos
Autores principales: Ranjitha, Y. S., Shetty, Anita N., Hemantkumar, Indrani, Gangakhedkar, Gauri R, Choudhari, Ashwini
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9912881/
https://www.ncbi.nlm.nih.gov/pubmed/36778832
http://dx.doi.org/10.4103/joacp.JOACP_638_20
_version_ 1784885300528414720
author Ranjitha, Y. S.
Shetty, Anita N.
Hemantkumar, Indrani
Gangakhedkar, Gauri R
Choudhari, Ashwini
author_facet Ranjitha, Y. S.
Shetty, Anita N.
Hemantkumar, Indrani
Gangakhedkar, Gauri R
Choudhari, Ashwini
author_sort Ranjitha, Y. S.
collection PubMed
description BACKGROUND AND AIMS: Blood loss in neurosurgical procedures can be rapid and tremendous leading to consequential hemodynamic instability. HemoCue is a portable photometer used for the measurement of blood hemoglobin concentration. Using this point of care device, we contemplated this study to assess the reliability of HemoCue for suction hemoglobin determination and calculation of surgical blood loss by comparing with the gold standard laboratory Coulter Counter method in patients undergoing neurosurgical procedures. MATERIAL AND METHODS: This prospective observational study was conducted in 233 patients undergoing elective neurosurgical procedures in the age group of 18 to 60 years and having preoperative hemoglobin (Hb) values above 9 g/dL. Values of preoperative hemoglobin, suction hemoglobin, and volume in the suction container were used to obtain the estimated blood loss. The normality of the data was tested using the Kolmogorov Smirnov test. Bland Altman analysis was used to test the reliability of HemoCue in estimating hemoglobin and blood loss with reference to the gold standard laboratory Coulter Counter automated hematology analyzer. RESULTS: Median blood loss calculated in our study using HemoCue was 554.65 mL with an interquartile range of 336.81 mL to 982.39 mL. Laboratory counter method estimated median blood loss was 533.37 mL with an interquartile range of 335.21 mL to 994.73 mL. The majority of the data obtained and analyzed using the Bland and Altman analysis method were within a 95% confidence interval. CONCLUSION: The HemoCue method is a reliable method and henceforth can be used to estimate blood loss in suction fluid in patients undergoing neurosurgical procedures.
format Online
Article
Text
id pubmed-9912881
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-99128812023-02-11 Evaluation of the HemoCue® for blood loss estimation in suction fluid in patients undergoing neurosurgical procedures: A prospective observational study Ranjitha, Y. S. Shetty, Anita N. Hemantkumar, Indrani Gangakhedkar, Gauri R Choudhari, Ashwini J Anaesthesiol Clin Pharmacol Original Article BACKGROUND AND AIMS: Blood loss in neurosurgical procedures can be rapid and tremendous leading to consequential hemodynamic instability. HemoCue is a portable photometer used for the measurement of blood hemoglobin concentration. Using this point of care device, we contemplated this study to assess the reliability of HemoCue for suction hemoglobin determination and calculation of surgical blood loss by comparing with the gold standard laboratory Coulter Counter method in patients undergoing neurosurgical procedures. MATERIAL AND METHODS: This prospective observational study was conducted in 233 patients undergoing elective neurosurgical procedures in the age group of 18 to 60 years and having preoperative hemoglobin (Hb) values above 9 g/dL. Values of preoperative hemoglobin, suction hemoglobin, and volume in the suction container were used to obtain the estimated blood loss. The normality of the data was tested using the Kolmogorov Smirnov test. Bland Altman analysis was used to test the reliability of HemoCue in estimating hemoglobin and blood loss with reference to the gold standard laboratory Coulter Counter automated hematology analyzer. RESULTS: Median blood loss calculated in our study using HemoCue was 554.65 mL with an interquartile range of 336.81 mL to 982.39 mL. Laboratory counter method estimated median blood loss was 533.37 mL with an interquartile range of 335.21 mL to 994.73 mL. The majority of the data obtained and analyzed using the Bland and Altman analysis method were within a 95% confidence interval. CONCLUSION: The HemoCue method is a reliable method and henceforth can be used to estimate blood loss in suction fluid in patients undergoing neurosurgical procedures. Wolters Kluwer - Medknow 2022 2022-06-15 /pmc/articles/PMC9912881/ /pubmed/36778832 http://dx.doi.org/10.4103/joacp.JOACP_638_20 Text en Copyright: © 2022 Journal of Anaesthesiology Clinical Pharmacology https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Ranjitha, Y. S.
Shetty, Anita N.
Hemantkumar, Indrani
Gangakhedkar, Gauri R
Choudhari, Ashwini
Evaluation of the HemoCue® for blood loss estimation in suction fluid in patients undergoing neurosurgical procedures: A prospective observational study
title Evaluation of the HemoCue® for blood loss estimation in suction fluid in patients undergoing neurosurgical procedures: A prospective observational study
title_full Evaluation of the HemoCue® for blood loss estimation in suction fluid in patients undergoing neurosurgical procedures: A prospective observational study
title_fullStr Evaluation of the HemoCue® for blood loss estimation in suction fluid in patients undergoing neurosurgical procedures: A prospective observational study
title_full_unstemmed Evaluation of the HemoCue® for blood loss estimation in suction fluid in patients undergoing neurosurgical procedures: A prospective observational study
title_short Evaluation of the HemoCue® for blood loss estimation in suction fluid in patients undergoing neurosurgical procedures: A prospective observational study
title_sort evaluation of the hemocue® for blood loss estimation in suction fluid in patients undergoing neurosurgical procedures: a prospective observational study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9912881/
https://www.ncbi.nlm.nih.gov/pubmed/36778832
http://dx.doi.org/10.4103/joacp.JOACP_638_20
work_keys_str_mv AT ranjithays evaluationofthehemocueforbloodlossestimationinsuctionfluidinpatientsundergoingneurosurgicalproceduresaprospectiveobservationalstudy
AT shettyanitan evaluationofthehemocueforbloodlossestimationinsuctionfluidinpatientsundergoingneurosurgicalproceduresaprospectiveobservationalstudy
AT hemantkumarindrani evaluationofthehemocueforbloodlossestimationinsuctionfluidinpatientsundergoingneurosurgicalproceduresaprospectiveobservationalstudy
AT gangakhedkargaurir evaluationofthehemocueforbloodlossestimationinsuctionfluidinpatientsundergoingneurosurgicalproceduresaprospectiveobservationalstudy
AT choudhariashwini evaluationofthehemocueforbloodlossestimationinsuctionfluidinpatientsundergoingneurosurgicalproceduresaprospectiveobservationalstudy