Characterization of Inorganic Scintillator Detectors for Dosimetry in Image-Guided Small Animal Radiotherapy Platforms

SIMPLE SUMMARY: Dosimetry for preclinical radiotherapy research requires the standardization of dose validation and quality assurance procedures. Except for dose reference measurements, dosimetric quality control of image-guided small animal irradiation platforms is mostly performed with passive det...

Descripción completa

Detalles Bibliográficos
Autores principales: Silvestre Patallo, Ileana, Subiel, Anna, Carter, Rebecca, Flynn, Samuel, Schettino, Giuseppe, Nisbet, Andrew
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9913621/
https://www.ncbi.nlm.nih.gov/pubmed/36765943
http://dx.doi.org/10.3390/cancers15030987
_version_ 1784885470670356480
author Silvestre Patallo, Ileana
Subiel, Anna
Carter, Rebecca
Flynn, Samuel
Schettino, Giuseppe
Nisbet, Andrew
author_facet Silvestre Patallo, Ileana
Subiel, Anna
Carter, Rebecca
Flynn, Samuel
Schettino, Giuseppe
Nisbet, Andrew
author_sort Silvestre Patallo, Ileana
collection PubMed
description SIMPLE SUMMARY: Dosimetry for preclinical radiotherapy research requires the standardization of dose validation and quality assurance procedures. Except for dose reference measurements, dosimetric quality control of image-guided small animal irradiation platforms is mostly performed with passive detectors (alanine and gafchromic films). This results in the inconvenient task of lengthy post-processing. In this paper, we carried out a dosimetric characterization of an active detection system based on inorganic scintillators in medium-energy X-rays. We implemented a cross-calibration framework based on international dosimetric protocols to assess the energy dependence of the detector. Additionally, we determined relative output factors for very small fields and compared them to other measurement systems (EBT3 film and CMOS sensor). We demonstrated the suitability of the inorganic scintillation system for the development of phantom-based end-to-end tests for dose verification in newly implemented preclinical radiotherapy irradiation techniques. ABSTRACT: The purpose of the study was to characterize a detection system based on inorganic scintillators and determine its suitability for dosimetry in preclinical radiation research. Dose rate, linearity, and repeatability of the response (among others) were assessed for medium-energy X-ray beam qualities. The response’s variation with temperature and beam angle incidence was also evaluated. Absorbed dose quality-dependent calibration coefficients, based on a cross-calibration against air kerma secondary standard ionization chambers, were determined. Relative output factors (ROF) for small, collimated fields (≤10 mm × 10 mm) were measured and compared with Gafchromic film and to a CMOS imaging sensor. Independently of the beam quality, the scintillator signal repeatability was adequate and linear with dose. Compared with EBT3 films and CMOS, ROF was within 5% (except for smaller circular fields). We demonstrated that when the detector is cross-calibrated in the user’s beam, it is a useful tool for dosimetry in medium-energy X-rays with small fields delivered by Image-Guided Small Animal Radiotherapy Platforms. It supports the development of procedures for independent “live” dose verification of complex preclinical radiotherapy plans with the possibility to insert the detectors in phantoms.
format Online
Article
Text
id pubmed-9913621
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-99136212023-02-11 Characterization of Inorganic Scintillator Detectors for Dosimetry in Image-Guided Small Animal Radiotherapy Platforms Silvestre Patallo, Ileana Subiel, Anna Carter, Rebecca Flynn, Samuel Schettino, Giuseppe Nisbet, Andrew Cancers (Basel) Article SIMPLE SUMMARY: Dosimetry for preclinical radiotherapy research requires the standardization of dose validation and quality assurance procedures. Except for dose reference measurements, dosimetric quality control of image-guided small animal irradiation platforms is mostly performed with passive detectors (alanine and gafchromic films). This results in the inconvenient task of lengthy post-processing. In this paper, we carried out a dosimetric characterization of an active detection system based on inorganic scintillators in medium-energy X-rays. We implemented a cross-calibration framework based on international dosimetric protocols to assess the energy dependence of the detector. Additionally, we determined relative output factors for very small fields and compared them to other measurement systems (EBT3 film and CMOS sensor). We demonstrated the suitability of the inorganic scintillation system for the development of phantom-based end-to-end tests for dose verification in newly implemented preclinical radiotherapy irradiation techniques. ABSTRACT: The purpose of the study was to characterize a detection system based on inorganic scintillators and determine its suitability for dosimetry in preclinical radiation research. Dose rate, linearity, and repeatability of the response (among others) were assessed for medium-energy X-ray beam qualities. The response’s variation with temperature and beam angle incidence was also evaluated. Absorbed dose quality-dependent calibration coefficients, based on a cross-calibration against air kerma secondary standard ionization chambers, were determined. Relative output factors (ROF) for small, collimated fields (≤10 mm × 10 mm) were measured and compared with Gafchromic film and to a CMOS imaging sensor. Independently of the beam quality, the scintillator signal repeatability was adequate and linear with dose. Compared with EBT3 films and CMOS, ROF was within 5% (except for smaller circular fields). We demonstrated that when the detector is cross-calibrated in the user’s beam, it is a useful tool for dosimetry in medium-energy X-rays with small fields delivered by Image-Guided Small Animal Radiotherapy Platforms. It supports the development of procedures for independent “live” dose verification of complex preclinical radiotherapy plans with the possibility to insert the detectors in phantoms. MDPI 2023-02-03 /pmc/articles/PMC9913621/ /pubmed/36765943 http://dx.doi.org/10.3390/cancers15030987 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Silvestre Patallo, Ileana
Subiel, Anna
Carter, Rebecca
Flynn, Samuel
Schettino, Giuseppe
Nisbet, Andrew
Characterization of Inorganic Scintillator Detectors for Dosimetry in Image-Guided Small Animal Radiotherapy Platforms
title Characterization of Inorganic Scintillator Detectors for Dosimetry in Image-Guided Small Animal Radiotherapy Platforms
title_full Characterization of Inorganic Scintillator Detectors for Dosimetry in Image-Guided Small Animal Radiotherapy Platforms
title_fullStr Characterization of Inorganic Scintillator Detectors for Dosimetry in Image-Guided Small Animal Radiotherapy Platforms
title_full_unstemmed Characterization of Inorganic Scintillator Detectors for Dosimetry in Image-Guided Small Animal Radiotherapy Platforms
title_short Characterization of Inorganic Scintillator Detectors for Dosimetry in Image-Guided Small Animal Radiotherapy Platforms
title_sort characterization of inorganic scintillator detectors for dosimetry in image-guided small animal radiotherapy platforms
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9913621/
https://www.ncbi.nlm.nih.gov/pubmed/36765943
http://dx.doi.org/10.3390/cancers15030987
work_keys_str_mv AT silvestrepatalloileana characterizationofinorganicscintillatordetectorsfordosimetryinimageguidedsmallanimalradiotherapyplatforms
AT subielanna characterizationofinorganicscintillatordetectorsfordosimetryinimageguidedsmallanimalradiotherapyplatforms
AT carterrebecca characterizationofinorganicscintillatordetectorsfordosimetryinimageguidedsmallanimalradiotherapyplatforms
AT flynnsamuel characterizationofinorganicscintillatordetectorsfordosimetryinimageguidedsmallanimalradiotherapyplatforms
AT schettinogiuseppe characterizationofinorganicscintillatordetectorsfordosimetryinimageguidedsmallanimalradiotherapyplatforms
AT nisbetandrew characterizationofinorganicscintillatordetectorsfordosimetryinimageguidedsmallanimalradiotherapyplatforms