Surface Analysis of Ti-Alloy Micro-Grooved 12/14 Tapers Assembled to Non-Sleeved and Sleeved Ceramic Heads: A Comparative Study of Retrieved Hip Prostheses

Ti6Al4V titanium alloy (Ti-alloy) sleeved ceramic heads have become widely used in revision surgery when the hip stem is left in situ. This solution guarantees a new junction between the bore of the ceramic head and the Ti-alloy sleeve, regardless of any possible, slight surface damage to the Ti-all...

Descripción completa

Detalles Bibliográficos
Autores principales: Martelli, Andrea, Erani, Paolo, Pazzagli, Nicola, Cannillo, Valeria, Baleani, Massimiliano
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9920329/
https://www.ncbi.nlm.nih.gov/pubmed/36770073
http://dx.doi.org/10.3390/ma16031067
_version_ 1784887043122266112
author Martelli, Andrea
Erani, Paolo
Pazzagli, Nicola
Cannillo, Valeria
Baleani, Massimiliano
author_facet Martelli, Andrea
Erani, Paolo
Pazzagli, Nicola
Cannillo, Valeria
Baleani, Massimiliano
author_sort Martelli, Andrea
collection PubMed
description Ti6Al4V titanium alloy (Ti-alloy) sleeved ceramic heads have become widely used in revision surgery when the hip stem is left in situ. This solution guarantees a new junction between the bore of the ceramic head and the Ti-alloy sleeve, regardless of any possible, slight surface damage to the Ti-alloy taper of the stem. However, this solution introduces an additional Ti-alloy/Ti-alloy interface pairing, which is potentially susceptible to mechanically assisted crevice corrosion. This study evaluated both qualitatively and quantitatively the damage that occurred in vivo on Ti-alloy micro-grooved 12/14 tapers of (i) primary implants with non-sleeved ceramic heads (Group 1), (ii) secondary implants with non-sleeved ceramic heads (Group 2), and (iii) secondary implants with sleeved ceramic heads (Group 3). A total of 45 explants—15 for each group, including short-, medium- and long-neck heads—underwent optical evaluation for surface damage (Goldberg scoring), surface roughness analysis, and SEM/EDX analysis. The Goldberg scores did not reveal different patterns in the tapers’ surface damage; surface damage was classified as absent or mild (surface damage score ≤2) in 94%, another 94%, and 92% of the analysed regions for Group 1, Group 2, and Group 3, respectively. Small but significant differences in morphological changes occurred in the tapers of the three groups: reductions no greater than a few percentage points in median values of roughness parameters were found in Group 1 and Group 2, while negligible changes were found in Group 3. SEM/EDX analysis revealed little (i.e., a slight increase in the oxygen content) to undetectable changes in the chemical composition on the Ti-alloy surface independently of the group. These results suggest that the Ti-alloy/Ti-alloy sleeve/taper junction is only mildly susceptible to mechanically assisted crevice corrosion. Assembling a sleeved ceramic head, with variable neck lengths up to a “long-neck”, to a Ti-alloy micro-grooved 12/14 taper of a stem left in situ does not seem to increase the risk of revision due to trunnionosis, as long as junction stability (i.e., the proper seating of the sleeved ceramic head on the 12/14 taper) is achieved intraoperatively.
format Online
Article
Text
id pubmed-9920329
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-99203292023-02-12 Surface Analysis of Ti-Alloy Micro-Grooved 12/14 Tapers Assembled to Non-Sleeved and Sleeved Ceramic Heads: A Comparative Study of Retrieved Hip Prostheses Martelli, Andrea Erani, Paolo Pazzagli, Nicola Cannillo, Valeria Baleani, Massimiliano Materials (Basel) Article Ti6Al4V titanium alloy (Ti-alloy) sleeved ceramic heads have become widely used in revision surgery when the hip stem is left in situ. This solution guarantees a new junction between the bore of the ceramic head and the Ti-alloy sleeve, regardless of any possible, slight surface damage to the Ti-alloy taper of the stem. However, this solution introduces an additional Ti-alloy/Ti-alloy interface pairing, which is potentially susceptible to mechanically assisted crevice corrosion. This study evaluated both qualitatively and quantitatively the damage that occurred in vivo on Ti-alloy micro-grooved 12/14 tapers of (i) primary implants with non-sleeved ceramic heads (Group 1), (ii) secondary implants with non-sleeved ceramic heads (Group 2), and (iii) secondary implants with sleeved ceramic heads (Group 3). A total of 45 explants—15 for each group, including short-, medium- and long-neck heads—underwent optical evaluation for surface damage (Goldberg scoring), surface roughness analysis, and SEM/EDX analysis. The Goldberg scores did not reveal different patterns in the tapers’ surface damage; surface damage was classified as absent or mild (surface damage score ≤2) in 94%, another 94%, and 92% of the analysed regions for Group 1, Group 2, and Group 3, respectively. Small but significant differences in morphological changes occurred in the tapers of the three groups: reductions no greater than a few percentage points in median values of roughness parameters were found in Group 1 and Group 2, while negligible changes were found in Group 3. SEM/EDX analysis revealed little (i.e., a slight increase in the oxygen content) to undetectable changes in the chemical composition on the Ti-alloy surface independently of the group. These results suggest that the Ti-alloy/Ti-alloy sleeve/taper junction is only mildly susceptible to mechanically assisted crevice corrosion. Assembling a sleeved ceramic head, with variable neck lengths up to a “long-neck”, to a Ti-alloy micro-grooved 12/14 taper of a stem left in situ does not seem to increase the risk of revision due to trunnionosis, as long as junction stability (i.e., the proper seating of the sleeved ceramic head on the 12/14 taper) is achieved intraoperatively. MDPI 2023-01-25 /pmc/articles/PMC9920329/ /pubmed/36770073 http://dx.doi.org/10.3390/ma16031067 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Martelli, Andrea
Erani, Paolo
Pazzagli, Nicola
Cannillo, Valeria
Baleani, Massimiliano
Surface Analysis of Ti-Alloy Micro-Grooved 12/14 Tapers Assembled to Non-Sleeved and Sleeved Ceramic Heads: A Comparative Study of Retrieved Hip Prostheses
title Surface Analysis of Ti-Alloy Micro-Grooved 12/14 Tapers Assembled to Non-Sleeved and Sleeved Ceramic Heads: A Comparative Study of Retrieved Hip Prostheses
title_full Surface Analysis of Ti-Alloy Micro-Grooved 12/14 Tapers Assembled to Non-Sleeved and Sleeved Ceramic Heads: A Comparative Study of Retrieved Hip Prostheses
title_fullStr Surface Analysis of Ti-Alloy Micro-Grooved 12/14 Tapers Assembled to Non-Sleeved and Sleeved Ceramic Heads: A Comparative Study of Retrieved Hip Prostheses
title_full_unstemmed Surface Analysis of Ti-Alloy Micro-Grooved 12/14 Tapers Assembled to Non-Sleeved and Sleeved Ceramic Heads: A Comparative Study of Retrieved Hip Prostheses
title_short Surface Analysis of Ti-Alloy Micro-Grooved 12/14 Tapers Assembled to Non-Sleeved and Sleeved Ceramic Heads: A Comparative Study of Retrieved Hip Prostheses
title_sort surface analysis of ti-alloy micro-grooved 12/14 tapers assembled to non-sleeved and sleeved ceramic heads: a comparative study of retrieved hip prostheses
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9920329/
https://www.ncbi.nlm.nih.gov/pubmed/36770073
http://dx.doi.org/10.3390/ma16031067
work_keys_str_mv AT martelliandrea surfaceanalysisoftialloymicrogrooved1214tapersassembledtononsleevedandsleevedceramicheadsacomparativestudyofretrievedhipprostheses
AT eranipaolo surfaceanalysisoftialloymicrogrooved1214tapersassembledtononsleevedandsleevedceramicheadsacomparativestudyofretrievedhipprostheses
AT pazzaglinicola surfaceanalysisoftialloymicrogrooved1214tapersassembledtononsleevedandsleevedceramicheadsacomparativestudyofretrievedhipprostheses
AT cannillovaleria surfaceanalysisoftialloymicrogrooved1214tapersassembledtononsleevedandsleevedceramicheadsacomparativestudyofretrievedhipprostheses
AT baleanimassimiliano surfaceanalysisoftialloymicrogrooved1214tapersassembledtononsleevedandsleevedceramicheadsacomparativestudyofretrievedhipprostheses