Cargando…

Towards Unbiased Fluorophore Counting in Superresolution Fluorescence Microscopy

With the advent of fluorescence superresolution microscopy, nano-sized structures can be imaged with a previously unprecedented accuracy. Therefore, it is rapidly gaining importance as an analytical tool in the life sciences and beyond. However, the images obtained so far lack an absolute scale in t...

Descripción completa

Detalles Bibliográficos
Autores principales: Laitenberger, Oskar, Aspelmeier, Timo, Staudt, Thomas, Geisler, Claudia, Munk, Axel, Egner, Alexander
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9921631/
https://www.ncbi.nlm.nih.gov/pubmed/36770420
http://dx.doi.org/10.3390/nano13030459
Descripción
Sumario:With the advent of fluorescence superresolution microscopy, nano-sized structures can be imaged with a previously unprecedented accuracy. Therefore, it is rapidly gaining importance as an analytical tool in the life sciences and beyond. However, the images obtained so far lack an absolute scale in terms of fluorophore numbers. Here, we use, for the first time, a detailed statistical model of the temporal imaging process which relies on a hidden Markov model operating on two timescales. This allows us to extract this information from the raw data without additional calibration measurements. We show this on the basis of added data from experiments on single Alexa 647 molecules as well as GSDIM/dSTORM measurements on DNA origami structures with a known number of labeling positions.