Cargando…

Simultaneous Determination of Xanthine and Hypoxanthine Using Polyglycine/rGO-Modified Glassy Carbon Electrode

A novel electrochemical sensor was developed for selective and sensitive determination of xanthine (XT) and hypoxanthine (HX) based on polyglycine (p-Gly) and reduced graphene oxide (rGO) modified glassy carbon electrode (GCE). A mixed dispersion of 7 μL of 5 mM glycine and 1 mg/mL GO was dropped on...

Descripción completa

Detalles Bibliográficos
Autores principales: Wang, Ting, Zhang, Lin, Zhang, Chengyu, Deng, Dongmei, Wang, Dejia, Luo, Liqiang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9922001/
https://www.ncbi.nlm.nih.gov/pubmed/36771122
http://dx.doi.org/10.3390/molecules28031458
Descripción
Sumario:A novel electrochemical sensor was developed for selective and sensitive determination of xanthine (XT) and hypoxanthine (HX) based on polyglycine (p-Gly) and reduced graphene oxide (rGO) modified glassy carbon electrode (GCE). A mixed dispersion of 7 μL of 5 mM glycine and 1 mg/mL GO was dropped on GCE for the fabrication of p-Gly/rGO/GCE, followed by cyclic voltammetric sweeping in 0.1 M phosphate buffer solution within −0.45~1.85 V at a scanning rate of 100 mV·s(−1). The morphological and electrochemical features of p-Gly/rGO/GCE were investigated by scanning electron microscopy and cyclic voltammetry. Under optimal conditions, the linear relationship was acquired for the simultaneous determination of XT and HX in 1–100 μM. The preparation of the electrode was simple and efficient. Additionally, the sensor combined the excellent conductivity of rGO and the polymerization of Gly, demonstrating satisfying simultaneous sensing performance to both XT and HX.