Cargando…
Dual beam‐current transformer design for monitoring and reporting of electron ultra‐high dose rate (FLASH) beam parameters
PURPOSE: To investigate the usefulness and effectiveness of a dual beam‐current transformer (BCTs) design to monitor and record the beam dosimetry output and energy of pulsed electron FLASH (eFLASH) beams in real‐time, and to inform on the usefulness of this design for future eFLASH beam control. ME...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9924113/ https://www.ncbi.nlm.nih.gov/pubmed/36601691 http://dx.doi.org/10.1002/acm2.13891 |
_version_ | 1784887827995033600 |
---|---|
author | Liu, Kevin Palmiero, Allison Chopra, Nitish Velasquez, Brett Li, Ziyi Beddar, Sam Schüler, Emil |
author_facet | Liu, Kevin Palmiero, Allison Chopra, Nitish Velasquez, Brett Li, Ziyi Beddar, Sam Schüler, Emil |
author_sort | Liu, Kevin |
collection | PubMed |
description | PURPOSE: To investigate the usefulness and effectiveness of a dual beam‐current transformer (BCTs) design to monitor and record the beam dosimetry output and energy of pulsed electron FLASH (eFLASH) beams in real‐time, and to inform on the usefulness of this design for future eFLASH beam control. METHODS: Two BCTs are integrated into the head of a FLASH Mobetron system, one located after the primary scattering foil and the other downstream of the secondary scattering foil. The response of the BCTs was evaluated individually to monitor beam output as a function of dose, scattering conditions, and ability to capture physical beam parameters such as pulse width (PW), pulse repetition frequency (PRF), and dose per pulse (DPP), and in combination to determine beam energy using the ratio of the lower‐to‐upper BCT signal. RESULTS: A linear relationship was observed between the absorbed dose measured on Gafchromic film and the BCT signals for both the upper and lower BCT (R (2) > 0.99). A linear relationship was also observed in the BCT signals as a function of the number of pulses delivered regardless of the PW, DPP, or PRF (R (2) > 0.99). The lower‐to‐upper BCT ratio was found to correlate strongly with the energy of the eFLASH beam due to differential beam attenuation caused by the secondary scattering foil. The BCTs were also able to provide accurate information about the PW, PRF, energy, and DPP for each individual pulse delivered in real‐time. CONCLUSION: The dual BCT system integrated within the FLASH Mobetron was shown to be a reliable monitoring system able to quantify accelerator performance and capture all essential physical beam parameters on a pulse‐by‐pulse basis, and the ratio between the two BCTs was strongly correlated with beam energy. The fast signal readout and processing enables the BCTs to provide real‐time information on beam output and energy and is proposed as a system suitable for accurate beam monitoring and control of eFLASH beams. |
format | Online Article Text |
id | pubmed-9924113 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-99241132023-02-14 Dual beam‐current transformer design for monitoring and reporting of electron ultra‐high dose rate (FLASH) beam parameters Liu, Kevin Palmiero, Allison Chopra, Nitish Velasquez, Brett Li, Ziyi Beddar, Sam Schüler, Emil J Appl Clin Med Phys Radiation Measurements PURPOSE: To investigate the usefulness and effectiveness of a dual beam‐current transformer (BCTs) design to monitor and record the beam dosimetry output and energy of pulsed electron FLASH (eFLASH) beams in real‐time, and to inform on the usefulness of this design for future eFLASH beam control. METHODS: Two BCTs are integrated into the head of a FLASH Mobetron system, one located after the primary scattering foil and the other downstream of the secondary scattering foil. The response of the BCTs was evaluated individually to monitor beam output as a function of dose, scattering conditions, and ability to capture physical beam parameters such as pulse width (PW), pulse repetition frequency (PRF), and dose per pulse (DPP), and in combination to determine beam energy using the ratio of the lower‐to‐upper BCT signal. RESULTS: A linear relationship was observed between the absorbed dose measured on Gafchromic film and the BCT signals for both the upper and lower BCT (R (2) > 0.99). A linear relationship was also observed in the BCT signals as a function of the number of pulses delivered regardless of the PW, DPP, or PRF (R (2) > 0.99). The lower‐to‐upper BCT ratio was found to correlate strongly with the energy of the eFLASH beam due to differential beam attenuation caused by the secondary scattering foil. The BCTs were also able to provide accurate information about the PW, PRF, energy, and DPP for each individual pulse delivered in real‐time. CONCLUSION: The dual BCT system integrated within the FLASH Mobetron was shown to be a reliable monitoring system able to quantify accelerator performance and capture all essential physical beam parameters on a pulse‐by‐pulse basis, and the ratio between the two BCTs was strongly correlated with beam energy. The fast signal readout and processing enables the BCTs to provide real‐time information on beam output and energy and is proposed as a system suitable for accurate beam monitoring and control of eFLASH beams. John Wiley and Sons Inc. 2023-01-04 /pmc/articles/PMC9924113/ /pubmed/36601691 http://dx.doi.org/10.1002/acm2.13891 Text en © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Measurements Liu, Kevin Palmiero, Allison Chopra, Nitish Velasquez, Brett Li, Ziyi Beddar, Sam Schüler, Emil Dual beam‐current transformer design for monitoring and reporting of electron ultra‐high dose rate (FLASH) beam parameters |
title | Dual beam‐current transformer design for monitoring and reporting of electron ultra‐high dose rate (FLASH) beam parameters |
title_full | Dual beam‐current transformer design for monitoring and reporting of electron ultra‐high dose rate (FLASH) beam parameters |
title_fullStr | Dual beam‐current transformer design for monitoring and reporting of electron ultra‐high dose rate (FLASH) beam parameters |
title_full_unstemmed | Dual beam‐current transformer design for monitoring and reporting of electron ultra‐high dose rate (FLASH) beam parameters |
title_short | Dual beam‐current transformer design for monitoring and reporting of electron ultra‐high dose rate (FLASH) beam parameters |
title_sort | dual beam‐current transformer design for monitoring and reporting of electron ultra‐high dose rate (flash) beam parameters |
topic | Radiation Measurements |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9924113/ https://www.ncbi.nlm.nih.gov/pubmed/36601691 http://dx.doi.org/10.1002/acm2.13891 |
work_keys_str_mv | AT liukevin dualbeamcurrenttransformerdesignformonitoringandreportingofelectronultrahighdoserateflashbeamparameters AT palmieroallison dualbeamcurrenttransformerdesignformonitoringandreportingofelectronultrahighdoserateflashbeamparameters AT chopranitish dualbeamcurrenttransformerdesignformonitoringandreportingofelectronultrahighdoserateflashbeamparameters AT velasquezbrett dualbeamcurrenttransformerdesignformonitoringandreportingofelectronultrahighdoserateflashbeamparameters AT liziyi dualbeamcurrenttransformerdesignformonitoringandreportingofelectronultrahighdoserateflashbeamparameters AT beddarsam dualbeamcurrenttransformerdesignformonitoringandreportingofelectronultrahighdoserateflashbeamparameters AT schuleremil dualbeamcurrenttransformerdesignformonitoringandreportingofelectronultrahighdoserateflashbeamparameters |