Influence of different air CT numbers for IVDT on the dose distribution in TomoTherapy MVCT

This study aims to evaluate the effect of different air computed tomography (CT) numbers of the image value density table (IVDT) on the retrospective dose calculation of head‐and‐neck (HN) radiotherapy using TomoTherapy megavoltage CT (MVCT) images. The CT numbers of the inside and outside air and e...

Descripción completa

Detalles Bibliográficos
Autores principales: Tsunemine, Shogo, Ozawa, Shuichi, Nakao, Minoru, Miura, Hideharu, Saito, Akito, Kawahara, Daisuke, Onishi, Yasuhiko, Onishi, Takashi, Hashiguchi, Taiki, Matsumoto, Yoshihisa, Maruta, Tsutomu, Murakami, Yuji, Nagata, Yasushi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9924119/
https://www.ncbi.nlm.nih.gov/pubmed/36316723
http://dx.doi.org/10.1002/acm2.13835
_version_ 1784887829495545856
author Tsunemine, Shogo
Ozawa, Shuichi
Nakao, Minoru
Miura, Hideharu
Saito, Akito
Kawahara, Daisuke
Onishi, Yasuhiko
Onishi, Takashi
Hashiguchi, Taiki
Matsumoto, Yoshihisa
Maruta, Tsutomu
Murakami, Yuji
Nagata, Yasushi
author_facet Tsunemine, Shogo
Ozawa, Shuichi
Nakao, Minoru
Miura, Hideharu
Saito, Akito
Kawahara, Daisuke
Onishi, Yasuhiko
Onishi, Takashi
Hashiguchi, Taiki
Matsumoto, Yoshihisa
Maruta, Tsutomu
Murakami, Yuji
Nagata, Yasushi
author_sort Tsunemine, Shogo
collection PubMed
description This study aims to evaluate the effect of different air computed tomography (CT) numbers of the image value density table (IVDT) on the retrospective dose calculation of head‐and‐neck (HN) radiotherapy using TomoTherapy megavoltage CT (MVCT) images. The CT numbers of the inside and outside air and each tissue‐equivalent plug of the “Cheese” phantom were obtained from TomoTherapy MVCT. Two IVDTs with different air CT numbers were created and applied to MVCT images of the HN anthropomorphic phantom and recalculated by Planned Adaptive to verify dose distribution. We defined the recalculation dose with MVCT images using both inside and outside air of the IVDT as [Formula: see text] and [Formula: see text] , respectively. Treatment planning doses calculated on kVCT images were compared with those calculated on MVCT images using two different IVDT tables, namely, [Formula: see text] and [Formula: see text]. The difference between average MVCT numbers ±1 standard deviation on inside and outside air of the calibration phantom was 65 ± 36 HU. This difference in MVCT number of air exceeded the recommendation lung tolerance for dose calculation error of 2%. The dose differences between the planning target volume (PTV): D (98%), D (50%), D (2%) and the organ at risk (OAR): D (max), D (mean) recalculated by [Formula: see text] and [Formula: see text] using MVCT images were a maximum of 0.7% and 1.2%, respectively. Recalculated doses to the PTV and OAR with MVCT showed that [Formula: see text] was 0.5%–0.7% closer to the kVCT treatment planning dose than [Formula: see text]. This study showed that [Formula: see text] was more accurate than [Formula: see text] in recalculating the dose HN cases of MVCT using TomoTherapy.
format Online
Article
Text
id pubmed-9924119
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-99241192023-02-14 Influence of different air CT numbers for IVDT on the dose distribution in TomoTherapy MVCT Tsunemine, Shogo Ozawa, Shuichi Nakao, Minoru Miura, Hideharu Saito, Akito Kawahara, Daisuke Onishi, Yasuhiko Onishi, Takashi Hashiguchi, Taiki Matsumoto, Yoshihisa Maruta, Tsutomu Murakami, Yuji Nagata, Yasushi J Appl Clin Med Phys Radiation Oncology Physics This study aims to evaluate the effect of different air computed tomography (CT) numbers of the image value density table (IVDT) on the retrospective dose calculation of head‐and‐neck (HN) radiotherapy using TomoTherapy megavoltage CT (MVCT) images. The CT numbers of the inside and outside air and each tissue‐equivalent plug of the “Cheese” phantom were obtained from TomoTherapy MVCT. Two IVDTs with different air CT numbers were created and applied to MVCT images of the HN anthropomorphic phantom and recalculated by Planned Adaptive to verify dose distribution. We defined the recalculation dose with MVCT images using both inside and outside air of the IVDT as [Formula: see text] and [Formula: see text] , respectively. Treatment planning doses calculated on kVCT images were compared with those calculated on MVCT images using two different IVDT tables, namely, [Formula: see text] and [Formula: see text]. The difference between average MVCT numbers ±1 standard deviation on inside and outside air of the calibration phantom was 65 ± 36 HU. This difference in MVCT number of air exceeded the recommendation lung tolerance for dose calculation error of 2%. The dose differences between the planning target volume (PTV): D (98%), D (50%), D (2%) and the organ at risk (OAR): D (max), D (mean) recalculated by [Formula: see text] and [Formula: see text] using MVCT images were a maximum of 0.7% and 1.2%, respectively. Recalculated doses to the PTV and OAR with MVCT showed that [Formula: see text] was 0.5%–0.7% closer to the kVCT treatment planning dose than [Formula: see text]. This study showed that [Formula: see text] was more accurate than [Formula: see text] in recalculating the dose HN cases of MVCT using TomoTherapy. John Wiley and Sons Inc. 2022-10-31 /pmc/articles/PMC9924119/ /pubmed/36316723 http://dx.doi.org/10.1002/acm2.13835 Text en © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Tsunemine, Shogo
Ozawa, Shuichi
Nakao, Minoru
Miura, Hideharu
Saito, Akito
Kawahara, Daisuke
Onishi, Yasuhiko
Onishi, Takashi
Hashiguchi, Taiki
Matsumoto, Yoshihisa
Maruta, Tsutomu
Murakami, Yuji
Nagata, Yasushi
Influence of different air CT numbers for IVDT on the dose distribution in TomoTherapy MVCT
title Influence of different air CT numbers for IVDT on the dose distribution in TomoTherapy MVCT
title_full Influence of different air CT numbers for IVDT on the dose distribution in TomoTherapy MVCT
title_fullStr Influence of different air CT numbers for IVDT on the dose distribution in TomoTherapy MVCT
title_full_unstemmed Influence of different air CT numbers for IVDT on the dose distribution in TomoTherapy MVCT
title_short Influence of different air CT numbers for IVDT on the dose distribution in TomoTherapy MVCT
title_sort influence of different air ct numbers for ivdt on the dose distribution in tomotherapy mvct
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9924119/
https://www.ncbi.nlm.nih.gov/pubmed/36316723
http://dx.doi.org/10.1002/acm2.13835
work_keys_str_mv AT tsunemineshogo influenceofdifferentairctnumbersforivdtonthedosedistributionintomotherapymvct
AT ozawashuichi influenceofdifferentairctnumbersforivdtonthedosedistributionintomotherapymvct
AT nakaominoru influenceofdifferentairctnumbersforivdtonthedosedistributionintomotherapymvct
AT miurahideharu influenceofdifferentairctnumbersforivdtonthedosedistributionintomotherapymvct
AT saitoakito influenceofdifferentairctnumbersforivdtonthedosedistributionintomotherapymvct
AT kawaharadaisuke influenceofdifferentairctnumbersforivdtonthedosedistributionintomotherapymvct
AT onishiyasuhiko influenceofdifferentairctnumbersforivdtonthedosedistributionintomotherapymvct
AT onishitakashi influenceofdifferentairctnumbersforivdtonthedosedistributionintomotherapymvct
AT hashiguchitaiki influenceofdifferentairctnumbersforivdtonthedosedistributionintomotherapymvct
AT matsumotoyoshihisa influenceofdifferentairctnumbersforivdtonthedosedistributionintomotherapymvct
AT marutatsutomu influenceofdifferentairctnumbersforivdtonthedosedistributionintomotherapymvct
AT murakamiyuji influenceofdifferentairctnumbersforivdtonthedosedistributionintomotherapymvct
AT nagatayasushi influenceofdifferentairctnumbersforivdtonthedosedistributionintomotherapymvct