Development of a three‐dimensional dose evaluation method for computed tomography
During a single scan using computed tomography, an X‐ray tube orbits along a 360°‐circular path around the patient. A scan obtained using the half‐cylindrical type phantoms with a radiochromic film sandwiched in between reveals a pixel value map illustrating the two‐dimensional (2D) dose distributio...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9924120/ https://www.ncbi.nlm.nih.gov/pubmed/36610053 http://dx.doi.org/10.1002/acm2.13897 |
Sumario: | During a single scan using computed tomography, an X‐ray tube orbits along a 360°‐circular path around the patient. A scan obtained using the half‐cylindrical type phantoms with a radiochromic film sandwiched in between reveals a pixel value map illustrating the two‐dimensional (2D) dose distribution. A three‐dimensional (3D) dose distribution can be obtained with a 360° rotation of the 2D dose map. This study evaluates the concept and methodology of creating a 3D dose map to develop a phantom with a radiochromic film for obtaining the 3D dose distribution. The coronal and axial plane dose distributions were also evaluated. A single scan computed tomography image obtained using a half‐cylindrical type of acrylic phantom with a sandwiched radiochromic film was studied. The diameters of the phantoms were 10 and 16 cm, and their lengths were 30 cm. A 2D image of the XR‐QA2 film was obtained using an image scanner and image processing software. A red channel image was used to obtain the 3D dose distribution using a computing platform. A pseudo color was applied to the red channel image from which cross‐sectional color images were obtained. Half of the cross‐sectional pixel data were rotated by 360° to generate the data for each axial plane. The image created was saved, and a 3D pixel value map was constructed. The dose measurement procedure for the 3D dose distribution was developed using half‐cylindrical acrylic phantoms with a radiochromic film. |
---|