Clinical performance evaluation of serum amyloid A module of Mindray BC-7500CS automated hematology analyzer

BACKGROUND: Laboratory detection of high values of serum amyloid A (SAA) is impaired by the hook effect. In response to this problem, Mindray has launched the new generation BC-7500CS automated hematology analyzer with an SAA autodilution (SAA-D) function. The present study aimed to verify the perfo...

Descripción completa

Detalles Bibliográficos
Autores principales: Dai, Jincheng, Zhang, Xiaoyu, Zhou, Jing, Pan, Wei, Yu, Fei
Formato: Online Artículo Texto
Lenguaje:English
Publicado: AME Publishing Company 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9926133/
https://www.ncbi.nlm.nih.gov/pubmed/36798927
http://dx.doi.org/10.21037/tp-22-661
_version_ 1784888211407896576
author Dai, Jincheng
Zhang, Xiaoyu
Zhou, Jing
Pan, Wei
Yu, Fei
author_facet Dai, Jincheng
Zhang, Xiaoyu
Zhou, Jing
Pan, Wei
Yu, Fei
author_sort Dai, Jincheng
collection PubMed
description BACKGROUND: Laboratory detection of high values of serum amyloid A (SAA) is impaired by the hook effect. In response to this problem, Mindray has launched the new generation BC-7500CS automated hematology analyzer with an SAA autodilution (SAA-D) function. The present study aimed to verify the performance of the SAA module. METHODS: Venous whole-blood specimens anticoagulated with EDTA-K2 were randomly collected from outpatients and inpatient of the Children’s Hospital of Nanjing Medical University (CH). Background, repeatability, precision, linear range, intermode comparison, and interference of the SAA module of the Mindray BC-7500CS were evaluated, and the performance of the SAA-D function was verified. RESULTS: The Mindray BC-7500CS showed an SAA background of 0.14 mg/L, well below that claimed by the manufacturer. Repeatability of SAA with standard deviation (SD) <0.6 mg/L and coefficient of variation (CV) <6%, the quality control (QC) precision was less than 8%. The measured value of the linear range was essentially consistent with the theoretical value, and the maximum measured values could reach 1932.38 mg/L. The deviation between whole-blood mode and micro-whole-blood mode was small (r=0.999), and the SAA module displayed high anti-interference ability. In addition, the measured results of specimens with high SAA concentration diluted by SAA-D were close to those after manual dilution (r=0.993). CONCLUSIONS: The SAA module of the Mindray BC-7500CS had excellent performance, and the SAA-D function was highly accurate at measuring specimens with high SAA concentration, enabling reliable SAA detection in the laboratory and clinical practice.
format Online
Article
Text
id pubmed-9926133
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher AME Publishing Company
record_format MEDLINE/PubMed
spelling pubmed-99261332023-02-15 Clinical performance evaluation of serum amyloid A module of Mindray BC-7500CS automated hematology analyzer Dai, Jincheng Zhang, Xiaoyu Zhou, Jing Pan, Wei Yu, Fei Transl Pediatr Original Article BACKGROUND: Laboratory detection of high values of serum amyloid A (SAA) is impaired by the hook effect. In response to this problem, Mindray has launched the new generation BC-7500CS automated hematology analyzer with an SAA autodilution (SAA-D) function. The present study aimed to verify the performance of the SAA module. METHODS: Venous whole-blood specimens anticoagulated with EDTA-K2 were randomly collected from outpatients and inpatient of the Children’s Hospital of Nanjing Medical University (CH). Background, repeatability, precision, linear range, intermode comparison, and interference of the SAA module of the Mindray BC-7500CS were evaluated, and the performance of the SAA-D function was verified. RESULTS: The Mindray BC-7500CS showed an SAA background of 0.14 mg/L, well below that claimed by the manufacturer. Repeatability of SAA with standard deviation (SD) <0.6 mg/L and coefficient of variation (CV) <6%, the quality control (QC) precision was less than 8%. The measured value of the linear range was essentially consistent with the theoretical value, and the maximum measured values could reach 1932.38 mg/L. The deviation between whole-blood mode and micro-whole-blood mode was small (r=0.999), and the SAA module displayed high anti-interference ability. In addition, the measured results of specimens with high SAA concentration diluted by SAA-D were close to those after manual dilution (r=0.993). CONCLUSIONS: The SAA module of the Mindray BC-7500CS had excellent performance, and the SAA-D function was highly accurate at measuring specimens with high SAA concentration, enabling reliable SAA detection in the laboratory and clinical practice. AME Publishing Company 2023-01-16 2023-01-31 /pmc/articles/PMC9926133/ /pubmed/36798927 http://dx.doi.org/10.21037/tp-22-661 Text en 2023 Translational Pediatrics. All rights reserved. https://creativecommons.org/licenses/by-nc-nd/4.0/Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0 (https://creativecommons.org/licenses/by-nc-nd/4.0/) .
spellingShingle Original Article
Dai, Jincheng
Zhang, Xiaoyu
Zhou, Jing
Pan, Wei
Yu, Fei
Clinical performance evaluation of serum amyloid A module of Mindray BC-7500CS automated hematology analyzer
title Clinical performance evaluation of serum amyloid A module of Mindray BC-7500CS automated hematology analyzer
title_full Clinical performance evaluation of serum amyloid A module of Mindray BC-7500CS automated hematology analyzer
title_fullStr Clinical performance evaluation of serum amyloid A module of Mindray BC-7500CS automated hematology analyzer
title_full_unstemmed Clinical performance evaluation of serum amyloid A module of Mindray BC-7500CS automated hematology analyzer
title_short Clinical performance evaluation of serum amyloid A module of Mindray BC-7500CS automated hematology analyzer
title_sort clinical performance evaluation of serum amyloid a module of mindray bc-7500cs automated hematology analyzer
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9926133/
https://www.ncbi.nlm.nih.gov/pubmed/36798927
http://dx.doi.org/10.21037/tp-22-661
work_keys_str_mv AT daijincheng clinicalperformanceevaluationofserumamyloidamoduleofmindraybc7500csautomatedhematologyanalyzer
AT zhangxiaoyu clinicalperformanceevaluationofserumamyloidamoduleofmindraybc7500csautomatedhematologyanalyzer
AT zhoujing clinicalperformanceevaluationofserumamyloidamoduleofmindraybc7500csautomatedhematologyanalyzer
AT panwei clinicalperformanceevaluationofserumamyloidamoduleofmindraybc7500csautomatedhematologyanalyzer
AT yufei clinicalperformanceevaluationofserumamyloidamoduleofmindraybc7500csautomatedhematologyanalyzer