A mixed-methods evaluation on the efficacy and perceptions of needleless connector disinfectants
OBJECTIVE: Optimizing needleless connector hub disinfection practice is a key strategy in central-line–associated bloodstream infection (CLABSI) prevention. In this mixed-methods evaluation, 3 products with varying scrub times were tested for experimental disinfection followed by a qualitative nursi...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Cambridge University Press
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9929708/ https://www.ncbi.nlm.nih.gov/pubmed/35387702 http://dx.doi.org/10.1017/ice.2022.72 |
Sumario: | OBJECTIVE: Optimizing needleless connector hub disinfection practice is a key strategy in central-line–associated bloodstream infection (CLABSI) prevention. In this mixed-methods evaluation, 3 products with varying scrub times were tested for experimental disinfection followed by a qualitative nursing assessment of each. METHODS: Needleless connectors were inoculated with varying concentrations of Staphylococcus epidermidis, Pseudomonas aeruginosa, and Staphylococcus aureus followed by disinfection with a 70% isopropyl alcohol (IPA) wipe (a 15-second scrub time and a 15-second dry time), a 70% IPA cap (a 10-second scrub time and a 5-second dry time), or a 3.15% chlorhexidine gluconate with 70% IPA (CHG/IPA) wipe (a 5-second scrub time and a 5-second dry time). Cultures of needleless connectors were obtained after disinfection to quantify bacterial reduction. This was followed by surveying a convenience sample of nursing staff with intensive care unit assignments at an academic tertiary hospital on use of each product. RESULTS: All products reduced overall bacterial burden when compared to sterile water controls, however the IPA and CHG/IPA wipes were superior to the IPA caps when product efficacy was compared. Nursing staff noted improved compliance with CHG/IPA wipes compared with the IPA wipes and the IPA caps, with many preferring the lesser scrub and dry times required for disinfection. CONCLUSION: Achieving adequate bacterial disinfection of needleless connectors while maximizing healthcare staff compliance with scrub and dry times may be best achieved with a combination CHG/IPA wipe. |
---|