Phantom-based evaluation of yttrium-90 datasets using biograph vision quadra

PURPOSE: The image quality characteristics of two NEMA phantoms with yttrium-90 ((90)Y) were evaluated on a long axial field-of-view (AFOV) PET/CT. The purpose was to identify the optimized reconstruction setup for the imaging of patients with hepatocellular carcinoma after (90)Y radioembolization....

Descripción completa

Detalles Bibliográficos
Autores principales: Zeimpekis, Konstantinos G., Mercolli, Lorenzo, Conti, Maurizio, Sari, Hasan, Prenosil, George, Shi, Kuangyu, Rominger, Axel
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9931793/
https://www.ncbi.nlm.nih.gov/pubmed/36504278
http://dx.doi.org/10.1007/s00259-022-06074-3
_version_ 1784889311847514112
author Zeimpekis, Konstantinos G.
Mercolli, Lorenzo
Conti, Maurizio
Sari, Hasan
Prenosil, George
Shi, Kuangyu
Rominger, Axel
author_facet Zeimpekis, Konstantinos G.
Mercolli, Lorenzo
Conti, Maurizio
Sari, Hasan
Prenosil, George
Shi, Kuangyu
Rominger, Axel
author_sort Zeimpekis, Konstantinos G.
collection PubMed
description PURPOSE: The image quality characteristics of two NEMA phantoms with yttrium-90 ((90)Y) were evaluated on a long axial field-of-view (AFOV) PET/CT. The purpose was to identify the optimized reconstruction setup for the imaging of patients with hepatocellular carcinoma after (90)Y radioembolization. METHODS: Two NEMA phantoms were used, where one had a 1:10 sphere to background activity concentration ratio and the second had cold background. Reconstruction parameters used are as follows: iterations 2 to 8, Gaussian filter 2- to 6-mm full-width-at-half-maximum, reconstruction matrices 440 × 440 and 220 × 220, high sensitivity (HS), and ultra-high sensitivity (UHS) modes. 50-, 40-, 30-, 20-, 10-, and 5-min acquisitions were reconstructed. The measurements included recovery coefficients (RC), signal-to-noise ratio (SNR), background variability, and lung error which measures the residual error in the corrections. Patient data were reconstructed with 20-, 10-, 5-, and 1-min time frames and evaluated in terms of SNR. RESULTS: The RC for the hot phantom was 0.36, 0.45, 0.53, 0.63, 0.68, and 0.84 for the spheres with diameters of 10, 13, 17, 22, 28, and 37 mm, respectively, for UHS 2 iterations, a 220 × 220 matrix, and 50-min acquisition. The RC values did not differ with acquisition times down to 20 min. The SNR was the highest for 2 iterations, measured 11.7, 16.6, 17.6, 19.4, 21.9, and 27.7 while the background variability was the lowest (27.59, 27.08, 27.36, 26.44, 30.11, and 33.51%). The lung error was 18%. For the patient dataset, the SNR was 19%, 20%, 24%, and 31% higher for 2 iterations compared to 4 iterations for 20-, 10-, 5-, and 1-min time frames, respectively. CONCLUSIONS: This study evaluates the NEMA image quality of a long AFOV PET/CT scanner with (90)Y. It provides high RC for the smallest sphere compared to other standard AFOV scanners at shorter scan times. The maximum patient SNR was for 2 iterations, 20 min, while 5 min delivers images with acceptable SNR.
format Online
Article
Text
id pubmed-9931793
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-99317932023-02-17 Phantom-based evaluation of yttrium-90 datasets using biograph vision quadra Zeimpekis, Konstantinos G. Mercolli, Lorenzo Conti, Maurizio Sari, Hasan Prenosil, George Shi, Kuangyu Rominger, Axel Eur J Nucl Med Mol Imaging Original Article PURPOSE: The image quality characteristics of two NEMA phantoms with yttrium-90 ((90)Y) were evaluated on a long axial field-of-view (AFOV) PET/CT. The purpose was to identify the optimized reconstruction setup for the imaging of patients with hepatocellular carcinoma after (90)Y radioembolization. METHODS: Two NEMA phantoms were used, where one had a 1:10 sphere to background activity concentration ratio and the second had cold background. Reconstruction parameters used are as follows: iterations 2 to 8, Gaussian filter 2- to 6-mm full-width-at-half-maximum, reconstruction matrices 440 × 440 and 220 × 220, high sensitivity (HS), and ultra-high sensitivity (UHS) modes. 50-, 40-, 30-, 20-, 10-, and 5-min acquisitions were reconstructed. The measurements included recovery coefficients (RC), signal-to-noise ratio (SNR), background variability, and lung error which measures the residual error in the corrections. Patient data were reconstructed with 20-, 10-, 5-, and 1-min time frames and evaluated in terms of SNR. RESULTS: The RC for the hot phantom was 0.36, 0.45, 0.53, 0.63, 0.68, and 0.84 for the spheres with diameters of 10, 13, 17, 22, 28, and 37 mm, respectively, for UHS 2 iterations, a 220 × 220 matrix, and 50-min acquisition. The RC values did not differ with acquisition times down to 20 min. The SNR was the highest for 2 iterations, measured 11.7, 16.6, 17.6, 19.4, 21.9, and 27.7 while the background variability was the lowest (27.59, 27.08, 27.36, 26.44, 30.11, and 33.51%). The lung error was 18%. For the patient dataset, the SNR was 19%, 20%, 24%, and 31% higher for 2 iterations compared to 4 iterations for 20-, 10-, 5-, and 1-min time frames, respectively. CONCLUSIONS: This study evaluates the NEMA image quality of a long AFOV PET/CT scanner with (90)Y. It provides high RC for the smallest sphere compared to other standard AFOV scanners at shorter scan times. The maximum patient SNR was for 2 iterations, 20 min, while 5 min delivers images with acceptable SNR. Springer Berlin Heidelberg 2022-12-12 2023 /pmc/articles/PMC9931793/ /pubmed/36504278 http://dx.doi.org/10.1007/s00259-022-06074-3 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Article
Zeimpekis, Konstantinos G.
Mercolli, Lorenzo
Conti, Maurizio
Sari, Hasan
Prenosil, George
Shi, Kuangyu
Rominger, Axel
Phantom-based evaluation of yttrium-90 datasets using biograph vision quadra
title Phantom-based evaluation of yttrium-90 datasets using biograph vision quadra
title_full Phantom-based evaluation of yttrium-90 datasets using biograph vision quadra
title_fullStr Phantom-based evaluation of yttrium-90 datasets using biograph vision quadra
title_full_unstemmed Phantom-based evaluation of yttrium-90 datasets using biograph vision quadra
title_short Phantom-based evaluation of yttrium-90 datasets using biograph vision quadra
title_sort phantom-based evaluation of yttrium-90 datasets using biograph vision quadra
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9931793/
https://www.ncbi.nlm.nih.gov/pubmed/36504278
http://dx.doi.org/10.1007/s00259-022-06074-3
work_keys_str_mv AT zeimpekiskonstantinosg phantombasedevaluationofyttrium90datasetsusingbiographvisionquadra
AT mercollilorenzo phantombasedevaluationofyttrium90datasetsusingbiographvisionquadra
AT contimaurizio phantombasedevaluationofyttrium90datasetsusingbiographvisionquadra
AT sarihasan phantombasedevaluationofyttrium90datasetsusingbiographvisionquadra
AT prenosilgeorge phantombasedevaluationofyttrium90datasetsusingbiographvisionquadra
AT shikuangyu phantombasedevaluationofyttrium90datasetsusingbiographvisionquadra
AT romingeraxel phantombasedevaluationofyttrium90datasetsusingbiographvisionquadra