Cargando…
Ultra-fast, high spatial resolution single-pulse scintillation imaging of synchrocyclotron pencil beam scanning proton delivery
Objective. To demonstrates the ability of an ultra-fast imaging system to measure high resolution spatial and temporal beam characteristics of a synchrocyclotron proton pencil beam scanning (PBS) system. Approach. An ultra-fast (1 kHz frame rate), intensified CMOS camera was triggered by a scintilla...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
IOP Publishing
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9935801/ https://www.ncbi.nlm.nih.gov/pubmed/36716492 http://dx.doi.org/10.1088/1361-6560/acb753 |
_version_ | 1784890091007639552 |
---|---|
author | Clark, Megan Ding, Xuanfeng Zhao, Lewei Pogue, Brian Gladstone, David Rahman, Mahbubur Zhang, Rongxiao Bruza, Petr |
author_facet | Clark, Megan Ding, Xuanfeng Zhao, Lewei Pogue, Brian Gladstone, David Rahman, Mahbubur Zhang, Rongxiao Bruza, Petr |
author_sort | Clark, Megan |
collection | PubMed |
description | Objective. To demonstrates the ability of an ultra-fast imaging system to measure high resolution spatial and temporal beam characteristics of a synchrocyclotron proton pencil beam scanning (PBS) system. Approach. An ultra-fast (1 kHz frame rate), intensified CMOS camera was triggered by a scintillation sheet coupled to a remote trigger unit for beam on detection. The camera was calibrated using the linear (R (2) > 0.9922) dose response of a single spot beam to varying currents. Film taken for the single spot beam was used to produce a scintillation intensity to absolute dose calibration. Main results. Spatial alignment was confirmed with the film, where the x and y-profiles of the single spot cumulative image agreed within 1 mm. A sample brain patient plan was analyzed to demonstrate dose and temporal accuracy for a clinically-relevant plan, through agreement within 1 mm to the planned and delivered spot locations. The cumulative dose agreed with the planned dose with a gamma passing rate of 97.5% (2 mm/3%, 10% dose threshold). Significance. This is the first system able to capture single-pulse spatial and temporal information for the unique pulse structure of a synchrocyclotron PBS systems at conventional dose rates, enabled by the ultra-fast sampling frame rate of this camera. This study indicates that, with continued camera development and testing, target applications in clinical and FLASH proton beam characterization and validation are possible. |
format | Online Article Text |
id | pubmed-9935801 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | IOP Publishing |
record_format | MEDLINE/PubMed |
spelling | pubmed-99358012023-02-18 Ultra-fast, high spatial resolution single-pulse scintillation imaging of synchrocyclotron pencil beam scanning proton delivery Clark, Megan Ding, Xuanfeng Zhao, Lewei Pogue, Brian Gladstone, David Rahman, Mahbubur Zhang, Rongxiao Bruza, Petr Phys Med Biol Paper Objective. To demonstrates the ability of an ultra-fast imaging system to measure high resolution spatial and temporal beam characteristics of a synchrocyclotron proton pencil beam scanning (PBS) system. Approach. An ultra-fast (1 kHz frame rate), intensified CMOS camera was triggered by a scintillation sheet coupled to a remote trigger unit for beam on detection. The camera was calibrated using the linear (R (2) > 0.9922) dose response of a single spot beam to varying currents. Film taken for the single spot beam was used to produce a scintillation intensity to absolute dose calibration. Main results. Spatial alignment was confirmed with the film, where the x and y-profiles of the single spot cumulative image agreed within 1 mm. A sample brain patient plan was analyzed to demonstrate dose and temporal accuracy for a clinically-relevant plan, through agreement within 1 mm to the planned and delivered spot locations. The cumulative dose agreed with the planned dose with a gamma passing rate of 97.5% (2 mm/3%, 10% dose threshold). Significance. This is the first system able to capture single-pulse spatial and temporal information for the unique pulse structure of a synchrocyclotron PBS systems at conventional dose rates, enabled by the ultra-fast sampling frame rate of this camera. This study indicates that, with continued camera development and testing, target applications in clinical and FLASH proton beam characterization and validation are possible. IOP Publishing 2023-02-21 2023-02-17 /pmc/articles/PMC9935801/ /pubmed/36716492 http://dx.doi.org/10.1088/1361-6560/acb753 Text en © 2023 The Author(s). Published on behalf of Institute of Physics and Engineering in Medicine by IOP Publishing Ltd https://creativecommons.org/licenses/by/4.0/Original content from this work may be used under the terms of the Creative Commons Attribution 4.0 licence (https://creativecommons.org/licenses/by/4.0/) . Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. |
spellingShingle | Paper Clark, Megan Ding, Xuanfeng Zhao, Lewei Pogue, Brian Gladstone, David Rahman, Mahbubur Zhang, Rongxiao Bruza, Petr Ultra-fast, high spatial resolution single-pulse scintillation imaging of synchrocyclotron pencil beam scanning proton delivery |
title | Ultra-fast, high spatial resolution single-pulse scintillation imaging of synchrocyclotron pencil beam scanning proton delivery |
title_full | Ultra-fast, high spatial resolution single-pulse scintillation imaging of synchrocyclotron pencil beam scanning proton delivery |
title_fullStr | Ultra-fast, high spatial resolution single-pulse scintillation imaging of synchrocyclotron pencil beam scanning proton delivery |
title_full_unstemmed | Ultra-fast, high spatial resolution single-pulse scintillation imaging of synchrocyclotron pencil beam scanning proton delivery |
title_short | Ultra-fast, high spatial resolution single-pulse scintillation imaging of synchrocyclotron pencil beam scanning proton delivery |
title_sort | ultra-fast, high spatial resolution single-pulse scintillation imaging of synchrocyclotron pencil beam scanning proton delivery |
topic | Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9935801/ https://www.ncbi.nlm.nih.gov/pubmed/36716492 http://dx.doi.org/10.1088/1361-6560/acb753 |
work_keys_str_mv | AT clarkmegan ultrafasthighspatialresolutionsinglepulsescintillationimagingofsynchrocyclotronpencilbeamscanningprotondelivery AT dingxuanfeng ultrafasthighspatialresolutionsinglepulsescintillationimagingofsynchrocyclotronpencilbeamscanningprotondelivery AT zhaolewei ultrafasthighspatialresolutionsinglepulsescintillationimagingofsynchrocyclotronpencilbeamscanningprotondelivery AT poguebrian ultrafasthighspatialresolutionsinglepulsescintillationimagingofsynchrocyclotronpencilbeamscanningprotondelivery AT gladstonedavid ultrafasthighspatialresolutionsinglepulsescintillationimagingofsynchrocyclotronpencilbeamscanningprotondelivery AT rahmanmahbubur ultrafasthighspatialresolutionsinglepulsescintillationimagingofsynchrocyclotronpencilbeamscanningprotondelivery AT zhangrongxiao ultrafasthighspatialresolutionsinglepulsescintillationimagingofsynchrocyclotronpencilbeamscanningprotondelivery AT bruzapetr ultrafasthighspatialresolutionsinglepulsescintillationimagingofsynchrocyclotronpencilbeamscanningprotondelivery |