Cargando…
Processing Validation Metrics of Syva Enzyme-Multiplied Immunoassay Technique (EMIT) Methotrexate Assay for Beckman Coulter System
Background: High-dose methotrexate (HDMTX), defined as a dose greater than 500 mg/m(2), is used to treat a variety of cancers; and though safe, it can cause major toxicity. Syva enzyme-multiplied immunoassay technique (EMIT) methotrexate (MTX) assay (Gurgaon, India: Siemens Healthcare Diagnostics Lt...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Cureus
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9939913/ https://www.ncbi.nlm.nih.gov/pubmed/36814748 http://dx.doi.org/10.7759/cureus.34025 |
Sumario: | Background: High-dose methotrexate (HDMTX), defined as a dose greater than 500 mg/m(2), is used to treat a variety of cancers; and though safe, it can cause major toxicity. Syva enzyme-multiplied immunoassay technique (EMIT) methotrexate (MTX) assay (Gurgaon, India: Siemens Healthcare Diagnostics Ltd.) uses a homogeneous enzyme immunoassay method. Low-end precision performances are very important for laboratory methods, especially when their results have clinical significance at these levels. Methodology: A total of 25 replicates (five replicates per run, for five runs) were analyzed for profiling. Precision, accuracy, linearity, limit of blank, limit of detection, and limit of quantification were determined using existing guidelines. Imprecision profile and limit of quantitation (LoQ) at 10% were determined by fitting data with hyperbolic regression. Results: The coefficient of variation percentage (CV%) for low, mid, and high-level internal quality control (IQC) was 1.25%, 3.45%, and 1.55%, respectively. Similarly, estimated bias was -4.58%, -3.54%, -7.21% for each level. The assay linearity was maintained from a range of 0.041-1.993 mmol/L with an R(2) of 0.959. The limit of detection was estimated to be 0.07 mmol/L. Conclusion: Syva EMIT MTX assay can be precisely and accurately used to measure low levels of serum methotrexate at levels lower than claimed by the manufacturer, aiding in the monitoring of toxicity in patients. |
---|