Comparative Analysis of 10-2 Test on Advanced Vision Analyzer and Humphrey Perimeter in Glaucoma

PURPOSE: To evaluate diagnostic precision and prove equivalence of 2 devices, Advanced vision analyzer (AVA, Elisar Vision Technology) and Humphrey field analyzer (HFA, Zeiss) for the detection of glaucoma on 10-2 program. DESIGN: Prospective, cross-sectional, observational study. PARTICIPANTS: Thre...

Descripción completa

Detalles Bibliográficos
Autores principales: Narang, Priya, Agarwal, Amar, Agarwal, Ashvin, Narang, Rhea, Sundaramoorthy, Lavanya
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9944553/
https://www.ncbi.nlm.nih.gov/pubmed/36846107
http://dx.doi.org/10.1016/j.xops.2022.100264
_version_ 1784891940909613056
author Narang, Priya
Agarwal, Amar
Agarwal, Ashvin
Narang, Rhea
Sundaramoorthy, Lavanya
author_facet Narang, Priya
Agarwal, Amar
Agarwal, Ashvin
Narang, Rhea
Sundaramoorthy, Lavanya
author_sort Narang, Priya
collection PubMed
description PURPOSE: To evaluate diagnostic precision and prove equivalence of 2 devices, Advanced vision analyzer (AVA, Elisar Vision Technology) and Humphrey field analyzer (HFA, Zeiss) for the detection of glaucoma on 10-2 program. DESIGN: Prospective, cross-sectional, observational study. PARTICIPANTS: Threshold estimates of 1 eye each of 66 patients with glaucoma, 36 control participants, and 10 glaucoma suspects were analyzed on 10-2 test with AVA and HFA. METHODS: Mean sensitivity (MS) values of 68 points and central 16 test points were calculated and compared. Intraclass correlation (ICC), Bland–Altman (BA) plots, linear regression of MS, mean deviation (MD), and pattern standard deviation (PSD) were computed to assess the 10-2 threshold estimate of the devices. Receiver operating characteristic curves were generated for MS and MD values, and the area under the curve (AUC) was compared with assessing diagnostic precision. MAIN OUTCOME MEASURES: Mean sensitivity values of 68 points and central 16 points, AUC for MS and MD values, ICC values, BA plots, and linear-regression analysis. RESULTS: Bland–Altman plot showed significant correlation for MS, MD, and PSD values for both devices. For MS, the overall ICC value was 0.96 (P < 0.001) with a mean bias of 0.0 dB and limits of agreement range of 7.59. The difference in MS values between both devices was −0.4760 ± 1.95 (P > 0.05). The AUC for MS values for AVA was 0.89 and for HFA was 0.92 (P = 0.188); whereas it was similar at 0.88 for MD values (P = 0.799). Advanced vision analyzer and HFA identically discriminated between healthy and patients with glaucoma (P < 0.001), although HFA denoted marginally greater ability (P > 0.05). CONCLUSIONS: Statistical results denote adequate equivalence between AVA and HFA because threshold estimates of AVA strongly correlate with HFA for 10-2 program. FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found after the references.
format Online
Article
Text
id pubmed-9944553
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-99445532023-02-23 Comparative Analysis of 10-2 Test on Advanced Vision Analyzer and Humphrey Perimeter in Glaucoma Narang, Priya Agarwal, Amar Agarwal, Ashvin Narang, Rhea Sundaramoorthy, Lavanya Ophthalmol Sci Original Article PURPOSE: To evaluate diagnostic precision and prove equivalence of 2 devices, Advanced vision analyzer (AVA, Elisar Vision Technology) and Humphrey field analyzer (HFA, Zeiss) for the detection of glaucoma on 10-2 program. DESIGN: Prospective, cross-sectional, observational study. PARTICIPANTS: Threshold estimates of 1 eye each of 66 patients with glaucoma, 36 control participants, and 10 glaucoma suspects were analyzed on 10-2 test with AVA and HFA. METHODS: Mean sensitivity (MS) values of 68 points and central 16 test points were calculated and compared. Intraclass correlation (ICC), Bland–Altman (BA) plots, linear regression of MS, mean deviation (MD), and pattern standard deviation (PSD) were computed to assess the 10-2 threshold estimate of the devices. Receiver operating characteristic curves were generated for MS and MD values, and the area under the curve (AUC) was compared with assessing diagnostic precision. MAIN OUTCOME MEASURES: Mean sensitivity values of 68 points and central 16 points, AUC for MS and MD values, ICC values, BA plots, and linear-regression analysis. RESULTS: Bland–Altman plot showed significant correlation for MS, MD, and PSD values for both devices. For MS, the overall ICC value was 0.96 (P < 0.001) with a mean bias of 0.0 dB and limits of agreement range of 7.59. The difference in MS values between both devices was −0.4760 ± 1.95 (P > 0.05). The AUC for MS values for AVA was 0.89 and for HFA was 0.92 (P = 0.188); whereas it was similar at 0.88 for MD values (P = 0.799). Advanced vision analyzer and HFA identically discriminated between healthy and patients with glaucoma (P < 0.001), although HFA denoted marginally greater ability (P > 0.05). CONCLUSIONS: Statistical results denote adequate equivalence between AVA and HFA because threshold estimates of AVA strongly correlate with HFA for 10-2 program. FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found after the references. Elsevier 2022-12-26 /pmc/articles/PMC9944553/ /pubmed/36846107 http://dx.doi.org/10.1016/j.xops.2022.100264 Text en © 2022 Published by Elsevier Inc. on behalf of American Academy of Ophthalmology. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original Article
Narang, Priya
Agarwal, Amar
Agarwal, Ashvin
Narang, Rhea
Sundaramoorthy, Lavanya
Comparative Analysis of 10-2 Test on Advanced Vision Analyzer and Humphrey Perimeter in Glaucoma
title Comparative Analysis of 10-2 Test on Advanced Vision Analyzer and Humphrey Perimeter in Glaucoma
title_full Comparative Analysis of 10-2 Test on Advanced Vision Analyzer and Humphrey Perimeter in Glaucoma
title_fullStr Comparative Analysis of 10-2 Test on Advanced Vision Analyzer and Humphrey Perimeter in Glaucoma
title_full_unstemmed Comparative Analysis of 10-2 Test on Advanced Vision Analyzer and Humphrey Perimeter in Glaucoma
title_short Comparative Analysis of 10-2 Test on Advanced Vision Analyzer and Humphrey Perimeter in Glaucoma
title_sort comparative analysis of 10-2 test on advanced vision analyzer and humphrey perimeter in glaucoma
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9944553/
https://www.ncbi.nlm.nih.gov/pubmed/36846107
http://dx.doi.org/10.1016/j.xops.2022.100264
work_keys_str_mv AT narangpriya comparativeanalysisof102testonadvancedvisionanalyzerandhumphreyperimeteringlaucoma
AT agarwalamar comparativeanalysisof102testonadvancedvisionanalyzerandhumphreyperimeteringlaucoma
AT agarwalashvin comparativeanalysisof102testonadvancedvisionanalyzerandhumphreyperimeteringlaucoma
AT narangrhea comparativeanalysisof102testonadvancedvisionanalyzerandhumphreyperimeteringlaucoma
AT sundaramoorthylavanya comparativeanalysisof102testonadvancedvisionanalyzerandhumphreyperimeteringlaucoma