Dose-efficient assessment of trabecular microstructure using ultra-high-resolution photon-counting CT

Photon-counting (PC) detectors for clinical computed tomography (CT) may offer improved imaging capabilities compared to conventional energy-integrating (EI) detectors, e.g. superior spatial resolution and detective efficiency. We here investigate if PCCT can reduce the administered dose in examinat...

Descripción completa

Detalles Bibliográficos
Autores principales: Peña, Jaime A., Klein, Laura, Maier, Joscha, Damm, Timo, Schlemmer, Heinz-Peter, Engelke, Klaus, Glüer, Claus-Christian, Kachelrieß, Marc, Sawall, Stefan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9948845/
https://www.ncbi.nlm.nih.gov/pubmed/35597742
http://dx.doi.org/10.1016/j.zemedi.2022.04.001
_version_ 1784892865620475904
author Peña, Jaime A.
Klein, Laura
Maier, Joscha
Damm, Timo
Schlemmer, Heinz-Peter
Engelke, Klaus
Glüer, Claus-Christian
Kachelrieß, Marc
Sawall, Stefan
author_facet Peña, Jaime A.
Klein, Laura
Maier, Joscha
Damm, Timo
Schlemmer, Heinz-Peter
Engelke, Klaus
Glüer, Claus-Christian
Kachelrieß, Marc
Sawall, Stefan
author_sort Peña, Jaime A.
collection PubMed
description Photon-counting (PC) detectors for clinical computed tomography (CT) may offer improved imaging capabilities compared to conventional energy-integrating (EI) detectors, e.g. superior spatial resolution and detective efficiency. We here investigate if PCCT can reduce the administered dose in examinations aimed at quantifying trabecular bone microstructure. Five human vertebral bodies were scanned three times in an abdomen phantom (QRM, Germany) using an experimental dual-source CT (Somatom CounT, Siemens Healthineers, Germany) housing an EI detector (0.60 mm pixel size at the iso-center) and a PC detector (0.25 mm pixel size). A tube voltage of 120 kV was used. Tube current-time product for EICT was 355 mAs (23.8 mGy CTDI(32 cm)). Dose-matched UHR-PCCT (UHRdm, 23.8 mGy) and noise-matched acquisitions (UHRnm, 10.5 mGy) were performed and reconstructed to a voxel size of 0.156 mm using a sharp kernel. Measurements of bone mineral density (BMD) and trabecular separation (Tb.Sp) and Tb.Sp percentiles reflecting the different scales of the trabecular interspacing were performed and compared to a gold-standard measurement using a peripheral CT device (XtremeCT, SCANCO Medical, Switzerland) with an isotropic voxel size of 0.082 mm and 6.6 mGy CTDI(10 cm). The image noise was quantified and the relative error with respect to the gold-standard along with the agreement between CT protocols using Lin’s concordance correlation coefficient (r(CCC)) were calculated. The Mean ± StdDev of the measured image noise levels in EICT was 109.6 ± 3.9 HU. UHRdm acquisitions (same dose as EICT) showed a significantly lower noise level of 78.6 ± 4.6 HU (p = 0.0122). UHRnm (44% dose of EICT) showed a noise level of 115.8 ± 3.7 HU, very similar to EICT at the same spatial resolution. For BMD the overall Mean ± StdDev for EI, UHRdm and UHRnm were 114.8 ± 28.6 mgHA/cm(3), 121.6 ± 28.8 mgHA/cm(3) and 121.5 ± 28.6 mgHA/cm(3), respectively, compared to 123.1 ± 25.5 mgHA/cm(3) for XtremeCT. For Tb.Sp these values were 1.86 ± 0.54 mm, 1.80 ± 0.56 mm and 1.84 ± 0.52 mm, respectively, compared to 1.66 ± 0.48 mm for XtremeCT. The ranking of the vertebrae with regard to Tb.Sp data was maintained throughout all Tb.Sp percentiles and among the CT protocols and the gold-standard. The agreement between protocols was very good for all comparisons: UHRnm vs. EICT (BMD r(CCC) = 0.97; Tb.Sp r(CCC) = 0.998), UHRnm vs. UHRdm (BMD r(CCC) = 0.998; Tb.Sp r(CCC) = 0.993) and UHRdm vs. EICT (BMD r(CCC) = 0.97; Tb.Sp r(CCC) = 0.991). Consequently, the relative RMS-errors from linear regressions against the gold-standard for EICT, UHRdm and UHRnm were very similar for BMD (7.1%, 5.2% and 5.4%) and for Tb.Sp (3.3%, 3.3% and 2.9%), with a much lower radiation dose for UHRnm. Short-term reproducibility for BMD measurements was similar and below 0.2% for all protocols, but for Tb.Sp showed better results for UHR (about 1/3 of the level for EICT). In conclusion, CT with UHR-PC detectors demonstrated lower image noise and better reproducibility for assessments of bone microstructure at similar dose levels. For UHRnm, radiation exposure levels could be reduced by 56% without deterioration of performance levels in the assessment of bone mineral density and bone microstructure.
format Online
Article
Text
id pubmed-9948845
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-99488452023-02-23 Dose-efficient assessment of trabecular microstructure using ultra-high-resolution photon-counting CT Peña, Jaime A. Klein, Laura Maier, Joscha Damm, Timo Schlemmer, Heinz-Peter Engelke, Klaus Glüer, Claus-Christian Kachelrieß, Marc Sawall, Stefan Z Med Phys Original Paper Photon-counting (PC) detectors for clinical computed tomography (CT) may offer improved imaging capabilities compared to conventional energy-integrating (EI) detectors, e.g. superior spatial resolution and detective efficiency. We here investigate if PCCT can reduce the administered dose in examinations aimed at quantifying trabecular bone microstructure. Five human vertebral bodies were scanned three times in an abdomen phantom (QRM, Germany) using an experimental dual-source CT (Somatom CounT, Siemens Healthineers, Germany) housing an EI detector (0.60 mm pixel size at the iso-center) and a PC detector (0.25 mm pixel size). A tube voltage of 120 kV was used. Tube current-time product for EICT was 355 mAs (23.8 mGy CTDI(32 cm)). Dose-matched UHR-PCCT (UHRdm, 23.8 mGy) and noise-matched acquisitions (UHRnm, 10.5 mGy) were performed and reconstructed to a voxel size of 0.156 mm using a sharp kernel. Measurements of bone mineral density (BMD) and trabecular separation (Tb.Sp) and Tb.Sp percentiles reflecting the different scales of the trabecular interspacing were performed and compared to a gold-standard measurement using a peripheral CT device (XtremeCT, SCANCO Medical, Switzerland) with an isotropic voxel size of 0.082 mm and 6.6 mGy CTDI(10 cm). The image noise was quantified and the relative error with respect to the gold-standard along with the agreement between CT protocols using Lin’s concordance correlation coefficient (r(CCC)) were calculated. The Mean ± StdDev of the measured image noise levels in EICT was 109.6 ± 3.9 HU. UHRdm acquisitions (same dose as EICT) showed a significantly lower noise level of 78.6 ± 4.6 HU (p = 0.0122). UHRnm (44% dose of EICT) showed a noise level of 115.8 ± 3.7 HU, very similar to EICT at the same spatial resolution. For BMD the overall Mean ± StdDev for EI, UHRdm and UHRnm were 114.8 ± 28.6 mgHA/cm(3), 121.6 ± 28.8 mgHA/cm(3) and 121.5 ± 28.6 mgHA/cm(3), respectively, compared to 123.1 ± 25.5 mgHA/cm(3) for XtremeCT. For Tb.Sp these values were 1.86 ± 0.54 mm, 1.80 ± 0.56 mm and 1.84 ± 0.52 mm, respectively, compared to 1.66 ± 0.48 mm for XtremeCT. The ranking of the vertebrae with regard to Tb.Sp data was maintained throughout all Tb.Sp percentiles and among the CT protocols and the gold-standard. The agreement between protocols was very good for all comparisons: UHRnm vs. EICT (BMD r(CCC) = 0.97; Tb.Sp r(CCC) = 0.998), UHRnm vs. UHRdm (BMD r(CCC) = 0.998; Tb.Sp r(CCC) = 0.993) and UHRdm vs. EICT (BMD r(CCC) = 0.97; Tb.Sp r(CCC) = 0.991). Consequently, the relative RMS-errors from linear regressions against the gold-standard for EICT, UHRdm and UHRnm were very similar for BMD (7.1%, 5.2% and 5.4%) and for Tb.Sp (3.3%, 3.3% and 2.9%), with a much lower radiation dose for UHRnm. Short-term reproducibility for BMD measurements was similar and below 0.2% for all protocols, but for Tb.Sp showed better results for UHR (about 1/3 of the level for EICT). In conclusion, CT with UHR-PC detectors demonstrated lower image noise and better reproducibility for assessments of bone microstructure at similar dose levels. For UHRnm, radiation exposure levels could be reduced by 56% without deterioration of performance levels in the assessment of bone mineral density and bone microstructure. Elsevier 2022-05-18 /pmc/articles/PMC9948845/ /pubmed/35597742 http://dx.doi.org/10.1016/j.zemedi.2022.04.001 Text en © 2022 Published by Elsevier GmbH on behalf of DGMP, ÖGMP and SSRMP. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original Paper
Peña, Jaime A.
Klein, Laura
Maier, Joscha
Damm, Timo
Schlemmer, Heinz-Peter
Engelke, Klaus
Glüer, Claus-Christian
Kachelrieß, Marc
Sawall, Stefan
Dose-efficient assessment of trabecular microstructure using ultra-high-resolution photon-counting CT
title Dose-efficient assessment of trabecular microstructure using ultra-high-resolution photon-counting CT
title_full Dose-efficient assessment of trabecular microstructure using ultra-high-resolution photon-counting CT
title_fullStr Dose-efficient assessment of trabecular microstructure using ultra-high-resolution photon-counting CT
title_full_unstemmed Dose-efficient assessment of trabecular microstructure using ultra-high-resolution photon-counting CT
title_short Dose-efficient assessment of trabecular microstructure using ultra-high-resolution photon-counting CT
title_sort dose-efficient assessment of trabecular microstructure using ultra-high-resolution photon-counting ct
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9948845/
https://www.ncbi.nlm.nih.gov/pubmed/35597742
http://dx.doi.org/10.1016/j.zemedi.2022.04.001
work_keys_str_mv AT penajaimea doseefficientassessmentoftrabecularmicrostructureusingultrahighresolutionphotoncountingct
AT kleinlaura doseefficientassessmentoftrabecularmicrostructureusingultrahighresolutionphotoncountingct
AT maierjoscha doseefficientassessmentoftrabecularmicrostructureusingultrahighresolutionphotoncountingct
AT dammtimo doseefficientassessmentoftrabecularmicrostructureusingultrahighresolutionphotoncountingct
AT schlemmerheinzpeter doseefficientassessmentoftrabecularmicrostructureusingultrahighresolutionphotoncountingct
AT engelkeklaus doseefficientassessmentoftrabecularmicrostructureusingultrahighresolutionphotoncountingct
AT gluerclauschristian doseefficientassessmentoftrabecularmicrostructureusingultrahighresolutionphotoncountingct
AT kachelrießmarc doseefficientassessmentoftrabecularmicrostructureusingultrahighresolutionphotoncountingct
AT sawallstefan doseefficientassessmentoftrabecularmicrostructureusingultrahighresolutionphotoncountingct