Cargando…

Feasibility of a method for low contrast CT image quality assessment using difference detail curves for abdominal scans

This work describes a measurement method for assessing dose-related image-quality of CT scans based on the difference detail curve (DDC) method, and showcases its use in a low contrast setting. The method is based on a phantom consisting of elliptical slices of different sizes into which contrast ob...

Descripción completa

Detalles Bibliográficos
Autores principales: Sommer, Christian, Özden, Ismail, Weyland, Mathias S., Duran, Carolina, Lutters, Gerd, Scheidegger, Stephan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9948851/
https://www.ncbi.nlm.nih.gov/pubmed/35184974
http://dx.doi.org/10.1016/j.zemedi.2022.01.001
_version_ 1784892867019276288
author Sommer, Christian
Özden, Ismail
Weyland, Mathias S.
Duran, Carolina
Lutters, Gerd
Scheidegger, Stephan
author_facet Sommer, Christian
Özden, Ismail
Weyland, Mathias S.
Duran, Carolina
Lutters, Gerd
Scheidegger, Stephan
author_sort Sommer, Christian
collection PubMed
description This work describes a measurement method for assessing dose-related image-quality of CT scans based on the difference detail curve (DDC) method, and showcases its use in a low contrast setting. The method is based on a phantom consisting of elliptical slices of different sizes into which contrast object modules can be inserted. These modules contain contrast objects based on (synthetic) resin mixtures with sucrose (native) or sodium iodine (contrast medium). Mixing ratios are provided to achieve a range of clinically relevant CT-numbers with these materials. The phantom is characterized in terms of contrast accuracy, energy dependency and long-term drift with satisfying results. Contrast accuracy and energy dependency are similar to that of water or soft tissue. Image quality of 655 scans of the phantom acquired at 30 different clinical institutions and with 16 different CT scanner models from 4 manufacturers was assessed by calculating a difference detail curve (DDC) from evaluation of up to 5 human observers using a custom-made software (RadiVates) described in this work. Based on these measurements, inter-observer variability was quantified using a bootstrap method and was shown to be a large contributor to the overall variability. This work demonstrates that assessment of CT image quality is feasible with the aforementioned phantom and DDC method.
format Online
Article
Text
id pubmed-9948851
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-99488512023-02-23 Feasibility of a method for low contrast CT image quality assessment using difference detail curves for abdominal scans Sommer, Christian Özden, Ismail Weyland, Mathias S. Duran, Carolina Lutters, Gerd Scheidegger, Stephan Z Med Phys Original Paper This work describes a measurement method for assessing dose-related image-quality of CT scans based on the difference detail curve (DDC) method, and showcases its use in a low contrast setting. The method is based on a phantom consisting of elliptical slices of different sizes into which contrast object modules can be inserted. These modules contain contrast objects based on (synthetic) resin mixtures with sucrose (native) or sodium iodine (contrast medium). Mixing ratios are provided to achieve a range of clinically relevant CT-numbers with these materials. The phantom is characterized in terms of contrast accuracy, energy dependency and long-term drift with satisfying results. Contrast accuracy and energy dependency are similar to that of water or soft tissue. Image quality of 655 scans of the phantom acquired at 30 different clinical institutions and with 16 different CT scanner models from 4 manufacturers was assessed by calculating a difference detail curve (DDC) from evaluation of up to 5 human observers using a custom-made software (RadiVates) described in this work. Based on these measurements, inter-observer variability was quantified using a bootstrap method and was shown to be a large contributor to the overall variability. This work demonstrates that assessment of CT image quality is feasible with the aforementioned phantom and DDC method. Elsevier 2022-02-17 /pmc/articles/PMC9948851/ /pubmed/35184974 http://dx.doi.org/10.1016/j.zemedi.2022.01.001 Text en © 2021 The Author(s). Published by Elsevier GmbH on behalf of DGMP, ÖGMP and SSRMP. https://creativecommons.org/licenses/by/4.0/This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Original Paper
Sommer, Christian
Özden, Ismail
Weyland, Mathias S.
Duran, Carolina
Lutters, Gerd
Scheidegger, Stephan
Feasibility of a method for low contrast CT image quality assessment using difference detail curves for abdominal scans
title Feasibility of a method for low contrast CT image quality assessment using difference detail curves for abdominal scans
title_full Feasibility of a method for low contrast CT image quality assessment using difference detail curves for abdominal scans
title_fullStr Feasibility of a method for low contrast CT image quality assessment using difference detail curves for abdominal scans
title_full_unstemmed Feasibility of a method for low contrast CT image quality assessment using difference detail curves for abdominal scans
title_short Feasibility of a method for low contrast CT image quality assessment using difference detail curves for abdominal scans
title_sort feasibility of a method for low contrast ct image quality assessment using difference detail curves for abdominal scans
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9948851/
https://www.ncbi.nlm.nih.gov/pubmed/35184974
http://dx.doi.org/10.1016/j.zemedi.2022.01.001
work_keys_str_mv AT sommerchristian feasibilityofamethodforlowcontrastctimagequalityassessmentusingdifferencedetailcurvesforabdominalscans
AT ozdenismail feasibilityofamethodforlowcontrastctimagequalityassessmentusingdifferencedetailcurvesforabdominalscans
AT weylandmathiass feasibilityofamethodforlowcontrastctimagequalityassessmentusingdifferencedetailcurvesforabdominalscans
AT durancarolina feasibilityofamethodforlowcontrastctimagequalityassessmentusingdifferencedetailcurvesforabdominalscans
AT luttersgerd feasibilityofamethodforlowcontrastctimagequalityassessmentusingdifferencedetailcurvesforabdominalscans
AT scheideggerstephan feasibilityofamethodforlowcontrastctimagequalityassessmentusingdifferencedetailcurvesforabdominalscans