Assessment of Deviations of Implants Installed with Prototyped Surgical Guide and Conventional Guide: In Vitro Study

Objective  The study aimed to assess the angular and linear deviations of implants installed in mannequins aided by surgical guides produced with the techniques of dual tomography (DT), model-based tomography (MT), and nonprototyped guide. Materials and Methods  Implants were installed in mannequins...

Descripción completa

Detalles Bibliográficos
Autores principales: Rodrigues, João Marcelo Meireles, Santos, Pâmela Leticia, Mendonça, Gustavo, Faloni, Ana Paula de Souza, Finoti, Livia Sertori, Margonar, Rogério
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Thieme Medical and Scientific Publishers Pvt. Ltd. 2022
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9949936/
https://www.ncbi.nlm.nih.gov/pubmed/36063845
http://dx.doi.org/10.1055/s-0040-1718791
_version_ 1784893054560239616
author Rodrigues, João Marcelo Meireles
Santos, Pâmela Leticia
Mendonça, Gustavo
Faloni, Ana Paula de Souza
Finoti, Livia Sertori
Margonar, Rogério
author_facet Rodrigues, João Marcelo Meireles
Santos, Pâmela Leticia
Mendonça, Gustavo
Faloni, Ana Paula de Souza
Finoti, Livia Sertori
Margonar, Rogério
author_sort Rodrigues, João Marcelo Meireles
collection PubMed
description Objective  The study aimed to assess the angular and linear deviations of implants installed in mannequins aided by surgical guides produced with the techniques of dual tomography (DT), model-based tomography (MT), and nonprototyped guide. Materials and Methods  Implants were installed in mannequins of a partially edentulous maxilla and divided into three groups: Group C ( n  = 20), implants installed using the conventional technique with flap opening and conventional guide; Group DT ( n  = 20), implants installed using guided surgery with the dual tomography technique; and Group MT ( n  = 20), implants installed using the model-based tomography technique. After implant installation, the mannequin was subjected to a computed tomography (CT) to measure the linear and angular deviations of implant positioning relative to the initial planning on both sides. Results  There was a higher mean angular deviation in group C (4.61 ± 1.21, p ≤ 0.001) than in groups DT (2.13 ± 0.62) and MT (1.87 ± 0.94), which were statistically similar between each other. Similarly, the linear deviations showed group C with the greatest discrepancy in relation to the other groups in the crown (2.17 ± 0.82, p  = 0.007), central (2.2 ± 0.77, p  = 0.004), and apical (2.34 ± 0.8, p  = 0.001) regions. Conclusion  The techniques of DT and MT presented smaller angular and linear deviations than the conventional technique with the nonprototyped guide. There was no difference between the two-guided surgery techniques.
format Online
Article
Text
id pubmed-9949936
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Thieme Medical and Scientific Publishers Pvt. Ltd.
record_format MEDLINE/PubMed
spelling pubmed-99499362023-02-24 Assessment of Deviations of Implants Installed with Prototyped Surgical Guide and Conventional Guide: In Vitro Study Rodrigues, João Marcelo Meireles Santos, Pâmela Leticia Mendonça, Gustavo Faloni, Ana Paula de Souza Finoti, Livia Sertori Margonar, Rogério Eur J Dent Objective  The study aimed to assess the angular and linear deviations of implants installed in mannequins aided by surgical guides produced with the techniques of dual tomography (DT), model-based tomography (MT), and nonprototyped guide. Materials and Methods  Implants were installed in mannequins of a partially edentulous maxilla and divided into three groups: Group C ( n  = 20), implants installed using the conventional technique with flap opening and conventional guide; Group DT ( n  = 20), implants installed using guided surgery with the dual tomography technique; and Group MT ( n  = 20), implants installed using the model-based tomography technique. After implant installation, the mannequin was subjected to a computed tomography (CT) to measure the linear and angular deviations of implant positioning relative to the initial planning on both sides. Results  There was a higher mean angular deviation in group C (4.61 ± 1.21, p ≤ 0.001) than in groups DT (2.13 ± 0.62) and MT (1.87 ± 0.94), which were statistically similar between each other. Similarly, the linear deviations showed group C with the greatest discrepancy in relation to the other groups in the crown (2.17 ± 0.82, p  = 0.007), central (2.2 ± 0.77, p  = 0.004), and apical (2.34 ± 0.8, p  = 0.001) regions. Conclusion  The techniques of DT and MT presented smaller angular and linear deviations than the conventional technique with the nonprototyped guide. There was no difference between the two-guided surgery techniques. Thieme Medical and Scientific Publishers Pvt. Ltd. 2022-09-05 /pmc/articles/PMC9949936/ /pubmed/36063845 http://dx.doi.org/10.1055/s-0040-1718791 Text en The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. ( https://creativecommons.org/licenses/by/4.0/ ) https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Rodrigues, João Marcelo Meireles
Santos, Pâmela Leticia
Mendonça, Gustavo
Faloni, Ana Paula de Souza
Finoti, Livia Sertori
Margonar, Rogério
Assessment of Deviations of Implants Installed with Prototyped Surgical Guide and Conventional Guide: In Vitro Study
title Assessment of Deviations of Implants Installed with Prototyped Surgical Guide and Conventional Guide: In Vitro Study
title_full Assessment of Deviations of Implants Installed with Prototyped Surgical Guide and Conventional Guide: In Vitro Study
title_fullStr Assessment of Deviations of Implants Installed with Prototyped Surgical Guide and Conventional Guide: In Vitro Study
title_full_unstemmed Assessment of Deviations of Implants Installed with Prototyped Surgical Guide and Conventional Guide: In Vitro Study
title_short Assessment of Deviations of Implants Installed with Prototyped Surgical Guide and Conventional Guide: In Vitro Study
title_sort assessment of deviations of implants installed with prototyped surgical guide and conventional guide: in vitro study
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9949936/
https://www.ncbi.nlm.nih.gov/pubmed/36063845
http://dx.doi.org/10.1055/s-0040-1718791
work_keys_str_mv AT rodriguesjoaomarcelomeireles assessmentofdeviationsofimplantsinstalledwithprototypedsurgicalguideandconventionalguideinvitrostudy
AT santospamelaleticia assessmentofdeviationsofimplantsinstalledwithprototypedsurgicalguideandconventionalguideinvitrostudy
AT mendoncagustavo assessmentofdeviationsofimplantsinstalledwithprototypedsurgicalguideandconventionalguideinvitrostudy
AT falonianapauladesouza assessmentofdeviationsofimplantsinstalledwithprototypedsurgicalguideandconventionalguideinvitrostudy
AT finotiliviasertori assessmentofdeviationsofimplantsinstalledwithprototypedsurgicalguideandconventionalguideinvitrostudy
AT margonarrogerio assessmentofdeviationsofimplantsinstalledwithprototypedsurgicalguideandconventionalguideinvitrostudy