Assessment of the clinical and analytical performance of the Aptima SARS-CoV-2 assay using the VALCOR protocol
BACKGROUND: The COVID-19 pandemic highlighted the importance of diagnostic testing against curbing the spread of SARS-CoV-2. The urgent need and scale for diagnostic tools resulted in manufacturers of SARS-CoV-2 assays receiving emergency authorization that lacked robust analytical or clinical evalu...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9951132/ https://www.ncbi.nlm.nih.gov/pubmed/36829164 http://dx.doi.org/10.1186/s12985-023-01986-4 |
Sumario: | BACKGROUND: The COVID-19 pandemic highlighted the importance of diagnostic testing against curbing the spread of SARS-CoV-2. The urgent need and scale for diagnostic tools resulted in manufacturers of SARS-CoV-2 assays receiving emergency authorization that lacked robust analytical or clinical evaluation. As it is highly likely that testing for SARS-CoV-2 will continue to play a central role in public health, the performance characteristics of assays should be evaluated to ensure reliable diagnostic outcomes are achieved. METHODS: VALCOR or “VALidation of SARS-CORona Virus-2 assays” is a study protocol designed to set up a framework for test validation of SARS-CoV-2 virus assays. Using clinical samples collated from VALCOR, the performance of Aptima SARS-CoV-2 assay was assessed against a standard comparator assay. Diagnostic test parameters such as sensitivity, specificity and overall per cent agreement were calculated for the clinical performance of Aptima SARS-CoV-2 assay. RESULTS: A total of 180 clinical samples were tested with an addition of 40 diluted clinical specimens to determine the limit of detection. When compared to the standard comparator assay Aptima had a sensitivity of 100.0% [95% CI 95.9–100.0] and specificity of 96.7% [95% CI 90.8–99.3]. The overall percent agreement was 98.3% with an excellent Cohen’s coefficient of κ = 0.967 [95% CI 0.929–1.000]. For the limit of detection, Aptima was able to detect all of the diluted clinical samples. CONCLUSION: In conclusion. validation of Aptima SARS-CoV-2 assay using clinical samples collated through the VALCOR protocol showed excellent test performance. Additionally, Aptima demonstrated high analytical sensitivity by detecting all diluted clinical samples corresponding to a low limit of detection. |
---|