Clinical and Technical Validation of Novel Bite Force Measuring Device for Functional Analysis after Mandibular Reconstruction
Bite force measuring devices that are generally suitable for edentulous patients or patients undergoing mandibular reconstruction are missing. This study assesses the validity of a new bite force measuring device (prototype of loadpad(®), novel GmbH) and evaluates its feasibility in patients after s...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9955263/ https://www.ncbi.nlm.nih.gov/pubmed/36832074 http://dx.doi.org/10.3390/diagnostics13040586 |
_version_ | 1784894309384847360 |
---|---|
author | Steffen, Claudius Duda, Katharina Wulsten, Dag Voss, Jan O. Koerdt, Steffen Nahles, Susanne Heiland, Max Checa, Sara Rendenbach, Carsten |
author_facet | Steffen, Claudius Duda, Katharina Wulsten, Dag Voss, Jan O. Koerdt, Steffen Nahles, Susanne Heiland, Max Checa, Sara Rendenbach, Carsten |
author_sort | Steffen, Claudius |
collection | PubMed |
description | Bite force measuring devices that are generally suitable for edentulous patients or patients undergoing mandibular reconstruction are missing. This study assesses the validity of a new bite force measuring device (prototype of loadpad(®), novel GmbH) and evaluates its feasibility in patients after segmental mandibular resection. Accuracy and reproducibility were analyzed with two different protocols using a universal testing machine (Z010 AllroundLine, Zwick/Roell, Ulm, Germany). Four groups were tested to evaluate the impact of silicone layers around the sensor: no silicone (“pure”), 2.0 mm soft silicone (“2-soft”), 7.0 mm soft silicone (“7-soft”) and 2.0 mm hard silicone (“2-hard”). Thereafter, the device was tested in 10 patients prospectively who underwent mandibular reconstruction using a fibula free flap. Average relative deviations of the measured force in relation to the applied load reached 0.77% (“7-soft”) to 5.28% (“2-hard”). Repeated measurements in “2-soft” revealed a mean relative deviation of 2.5% until an applied load of 600 N. Maximum bite force decreased postoperatively by 51.8% to a maximum mean bite force of 131.5 N. The novel device guarantees a high accuracy and degree of reproducibility. Furthermore, it offers new opportunities to quantify perioperative oral function after reconstructive surgery of the mandible also in edentulous patients. |
format | Online Article Text |
id | pubmed-9955263 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-99552632023-02-25 Clinical and Technical Validation of Novel Bite Force Measuring Device for Functional Analysis after Mandibular Reconstruction Steffen, Claudius Duda, Katharina Wulsten, Dag Voss, Jan O. Koerdt, Steffen Nahles, Susanne Heiland, Max Checa, Sara Rendenbach, Carsten Diagnostics (Basel) Article Bite force measuring devices that are generally suitable for edentulous patients or patients undergoing mandibular reconstruction are missing. This study assesses the validity of a new bite force measuring device (prototype of loadpad(®), novel GmbH) and evaluates its feasibility in patients after segmental mandibular resection. Accuracy and reproducibility were analyzed with two different protocols using a universal testing machine (Z010 AllroundLine, Zwick/Roell, Ulm, Germany). Four groups were tested to evaluate the impact of silicone layers around the sensor: no silicone (“pure”), 2.0 mm soft silicone (“2-soft”), 7.0 mm soft silicone (“7-soft”) and 2.0 mm hard silicone (“2-hard”). Thereafter, the device was tested in 10 patients prospectively who underwent mandibular reconstruction using a fibula free flap. Average relative deviations of the measured force in relation to the applied load reached 0.77% (“7-soft”) to 5.28% (“2-hard”). Repeated measurements in “2-soft” revealed a mean relative deviation of 2.5% until an applied load of 600 N. Maximum bite force decreased postoperatively by 51.8% to a maximum mean bite force of 131.5 N. The novel device guarantees a high accuracy and degree of reproducibility. Furthermore, it offers new opportunities to quantify perioperative oral function after reconstructive surgery of the mandible also in edentulous patients. MDPI 2023-02-05 /pmc/articles/PMC9955263/ /pubmed/36832074 http://dx.doi.org/10.3390/diagnostics13040586 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Steffen, Claudius Duda, Katharina Wulsten, Dag Voss, Jan O. Koerdt, Steffen Nahles, Susanne Heiland, Max Checa, Sara Rendenbach, Carsten Clinical and Technical Validation of Novel Bite Force Measuring Device for Functional Analysis after Mandibular Reconstruction |
title | Clinical and Technical Validation of Novel Bite Force Measuring Device for Functional Analysis after Mandibular Reconstruction |
title_full | Clinical and Technical Validation of Novel Bite Force Measuring Device for Functional Analysis after Mandibular Reconstruction |
title_fullStr | Clinical and Technical Validation of Novel Bite Force Measuring Device for Functional Analysis after Mandibular Reconstruction |
title_full_unstemmed | Clinical and Technical Validation of Novel Bite Force Measuring Device for Functional Analysis after Mandibular Reconstruction |
title_short | Clinical and Technical Validation of Novel Bite Force Measuring Device for Functional Analysis after Mandibular Reconstruction |
title_sort | clinical and technical validation of novel bite force measuring device for functional analysis after mandibular reconstruction |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9955263/ https://www.ncbi.nlm.nih.gov/pubmed/36832074 http://dx.doi.org/10.3390/diagnostics13040586 |
work_keys_str_mv | AT steffenclaudius clinicalandtechnicalvalidationofnovelbiteforcemeasuringdeviceforfunctionalanalysisaftermandibularreconstruction AT dudakatharina clinicalandtechnicalvalidationofnovelbiteforcemeasuringdeviceforfunctionalanalysisaftermandibularreconstruction AT wulstendag clinicalandtechnicalvalidationofnovelbiteforcemeasuringdeviceforfunctionalanalysisaftermandibularreconstruction AT vossjano clinicalandtechnicalvalidationofnovelbiteforcemeasuringdeviceforfunctionalanalysisaftermandibularreconstruction AT koerdtsteffen clinicalandtechnicalvalidationofnovelbiteforcemeasuringdeviceforfunctionalanalysisaftermandibularreconstruction AT nahlessusanne clinicalandtechnicalvalidationofnovelbiteforcemeasuringdeviceforfunctionalanalysisaftermandibularreconstruction AT heilandmax clinicalandtechnicalvalidationofnovelbiteforcemeasuringdeviceforfunctionalanalysisaftermandibularreconstruction AT checasara clinicalandtechnicalvalidationofnovelbiteforcemeasuringdeviceforfunctionalanalysisaftermandibularreconstruction AT rendenbachcarsten clinicalandtechnicalvalidationofnovelbiteforcemeasuringdeviceforfunctionalanalysisaftermandibularreconstruction |