Cargando…

Advanced Diffusion-Weighted Imaging Sequences for Breast MRI: Comprehensive Comparison of Improved Sequences and Ultra-High B-Values to Identify the Optimal Combination

This study investigated the image quality and choice of ultra-high b-value of two DWI breast-MRI research applications. The study cohort comprised 40 patients (20 malignant lesions). In addition to s-DWI with two m-b-values (b50 and b800) and three e-b-values (e-b1500, e-b2000, and e-b2500), z-DWI a...

Descripción completa

Detalles Bibliográficos
Autores principales: Hausmann, Daniel, Todorski, Inga, Pindur, Alexandra, Weiland, Elisabeth, Benkert, Thomas, Bosshard, Lars, Prummer, Michael, Kubik-Huch, Rahel A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9955562/
https://www.ncbi.nlm.nih.gov/pubmed/36832095
http://dx.doi.org/10.3390/diagnostics13040607
_version_ 1784894377299017728
author Hausmann, Daniel
Todorski, Inga
Pindur, Alexandra
Weiland, Elisabeth
Benkert, Thomas
Bosshard, Lars
Prummer, Michael
Kubik-Huch, Rahel A.
author_facet Hausmann, Daniel
Todorski, Inga
Pindur, Alexandra
Weiland, Elisabeth
Benkert, Thomas
Bosshard, Lars
Prummer, Michael
Kubik-Huch, Rahel A.
author_sort Hausmann, Daniel
collection PubMed
description This study investigated the image quality and choice of ultra-high b-value of two DWI breast-MRI research applications. The study cohort comprised 40 patients (20 malignant lesions). In addition to s-DWI with two m-b-values (b50 and b800) and three e-b-values (e-b1500, e-b2000, and e-b2500), z-DWI and IR m-b1500 DWI were applied. z-DWI was acquired with the same measured b-values and e-b-values as the standard sequence. For IR m-b1500 DWI, b50 and b1500 were measured, and e-b2000 and e-b2500 were mathematically extrapolated. Three readers used Likert scales to independently analyze all ultra-high b-values (b1500–b2500) for each DWI with regards to scan preference and image quality. ADC values were measured in all 20 lesions. z-DWI was the most preferred (54%), followed by IR m-b1500 DWI (46%). b1500 was significantly preferred over b2000 for z-DWI and IR m-b1500 DWI (p = 0.001 and p = 0.002, respectively). Lesion detection was not significantly different among sequences or b-values (p = 0.174). There were no significant differences in measured ADC values within lesions between s-DWI (ADC: 0.97 [±0.09] × 10(−3) mm(2)/s) and z-DWI (ADC: 0.99 [±0.11] × 10(−3) mm(2)/s; p = 1.000). However, there was a trend toward lower values in IR m-b1500 DWI (ADC: 0.80 [±0.06] × 10(−3) mm(2)/s) than in s-DWI (p = 0.090) and z-DWI (p = 0.110). Overall, image quality was superior and there were fewer image artifacts when using the advanced sequences (z-DWI + IR m-b1500 DWI) compared with s-DWI. Considering scan preferences, we found that the optimal combination was z-DWI with a calculated b1500, especially regarding examination time.
format Online
Article
Text
id pubmed-9955562
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-99555622023-02-25 Advanced Diffusion-Weighted Imaging Sequences for Breast MRI: Comprehensive Comparison of Improved Sequences and Ultra-High B-Values to Identify the Optimal Combination Hausmann, Daniel Todorski, Inga Pindur, Alexandra Weiland, Elisabeth Benkert, Thomas Bosshard, Lars Prummer, Michael Kubik-Huch, Rahel A. Diagnostics (Basel) Article This study investigated the image quality and choice of ultra-high b-value of two DWI breast-MRI research applications. The study cohort comprised 40 patients (20 malignant lesions). In addition to s-DWI with two m-b-values (b50 and b800) and three e-b-values (e-b1500, e-b2000, and e-b2500), z-DWI and IR m-b1500 DWI were applied. z-DWI was acquired with the same measured b-values and e-b-values as the standard sequence. For IR m-b1500 DWI, b50 and b1500 were measured, and e-b2000 and e-b2500 were mathematically extrapolated. Three readers used Likert scales to independently analyze all ultra-high b-values (b1500–b2500) for each DWI with regards to scan preference and image quality. ADC values were measured in all 20 lesions. z-DWI was the most preferred (54%), followed by IR m-b1500 DWI (46%). b1500 was significantly preferred over b2000 for z-DWI and IR m-b1500 DWI (p = 0.001 and p = 0.002, respectively). Lesion detection was not significantly different among sequences or b-values (p = 0.174). There were no significant differences in measured ADC values within lesions between s-DWI (ADC: 0.97 [±0.09] × 10(−3) mm(2)/s) and z-DWI (ADC: 0.99 [±0.11] × 10(−3) mm(2)/s; p = 1.000). However, there was a trend toward lower values in IR m-b1500 DWI (ADC: 0.80 [±0.06] × 10(−3) mm(2)/s) than in s-DWI (p = 0.090) and z-DWI (p = 0.110). Overall, image quality was superior and there were fewer image artifacts when using the advanced sequences (z-DWI + IR m-b1500 DWI) compared with s-DWI. Considering scan preferences, we found that the optimal combination was z-DWI with a calculated b1500, especially regarding examination time. MDPI 2023-02-07 /pmc/articles/PMC9955562/ /pubmed/36832095 http://dx.doi.org/10.3390/diagnostics13040607 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Hausmann, Daniel
Todorski, Inga
Pindur, Alexandra
Weiland, Elisabeth
Benkert, Thomas
Bosshard, Lars
Prummer, Michael
Kubik-Huch, Rahel A.
Advanced Diffusion-Weighted Imaging Sequences for Breast MRI: Comprehensive Comparison of Improved Sequences and Ultra-High B-Values to Identify the Optimal Combination
title Advanced Diffusion-Weighted Imaging Sequences for Breast MRI: Comprehensive Comparison of Improved Sequences and Ultra-High B-Values to Identify the Optimal Combination
title_full Advanced Diffusion-Weighted Imaging Sequences for Breast MRI: Comprehensive Comparison of Improved Sequences and Ultra-High B-Values to Identify the Optimal Combination
title_fullStr Advanced Diffusion-Weighted Imaging Sequences for Breast MRI: Comprehensive Comparison of Improved Sequences and Ultra-High B-Values to Identify the Optimal Combination
title_full_unstemmed Advanced Diffusion-Weighted Imaging Sequences for Breast MRI: Comprehensive Comparison of Improved Sequences and Ultra-High B-Values to Identify the Optimal Combination
title_short Advanced Diffusion-Weighted Imaging Sequences for Breast MRI: Comprehensive Comparison of Improved Sequences and Ultra-High B-Values to Identify the Optimal Combination
title_sort advanced diffusion-weighted imaging sequences for breast mri: comprehensive comparison of improved sequences and ultra-high b-values to identify the optimal combination
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9955562/
https://www.ncbi.nlm.nih.gov/pubmed/36832095
http://dx.doi.org/10.3390/diagnostics13040607
work_keys_str_mv AT hausmanndaniel advanceddiffusionweightedimagingsequencesforbreastmricomprehensivecomparisonofimprovedsequencesandultrahighbvaluestoidentifytheoptimalcombination
AT todorskiinga advanceddiffusionweightedimagingsequencesforbreastmricomprehensivecomparisonofimprovedsequencesandultrahighbvaluestoidentifytheoptimalcombination
AT pinduralexandra advanceddiffusionweightedimagingsequencesforbreastmricomprehensivecomparisonofimprovedsequencesandultrahighbvaluestoidentifytheoptimalcombination
AT weilandelisabeth advanceddiffusionweightedimagingsequencesforbreastmricomprehensivecomparisonofimprovedsequencesandultrahighbvaluestoidentifytheoptimalcombination
AT benkertthomas advanceddiffusionweightedimagingsequencesforbreastmricomprehensivecomparisonofimprovedsequencesandultrahighbvaluestoidentifytheoptimalcombination
AT bosshardlars advanceddiffusionweightedimagingsequencesforbreastmricomprehensivecomparisonofimprovedsequencesandultrahighbvaluestoidentifytheoptimalcombination
AT prummermichael advanceddiffusionweightedimagingsequencesforbreastmricomprehensivecomparisonofimprovedsequencesandultrahighbvaluestoidentifytheoptimalcombination
AT kubikhuchrahela advanceddiffusionweightedimagingsequencesforbreastmricomprehensivecomparisonofimprovedsequencesandultrahighbvaluestoidentifytheoptimalcombination