Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait

This study characterized the absolute pressure measurement error and reliability of a new fully integrated (Kinetyx, SI) plantar-pressure measurement system (PPMS) versus an industry-standard PPMS (F-Scan, Tekscan) during an established benchtop testing protocol as well as via a research-grade, inst...

Descripción completa

Detalles Bibliográficos
Autores principales: Blades, Samuel, Jensen, Matt, Stellingwerff, Trent, Hundza, Sandra, Klimstra, Marc
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9963688/
https://www.ncbi.nlm.nih.gov/pubmed/36850951
http://dx.doi.org/10.3390/s23042352
_version_ 1784896315700805632
author Blades, Samuel
Jensen, Matt
Stellingwerff, Trent
Hundza, Sandra
Klimstra, Marc
author_facet Blades, Samuel
Jensen, Matt
Stellingwerff, Trent
Hundza, Sandra
Klimstra, Marc
author_sort Blades, Samuel
collection PubMed
description This study characterized the absolute pressure measurement error and reliability of a new fully integrated (Kinetyx, SI) plantar-pressure measurement system (PPMS) versus an industry-standard PPMS (F-Scan, Tekscan) during an established benchtop testing protocol as well as via a research-grade, instrumented treadmill (Bertec) during a running protocol. Benchtop testing results showed that both SI and F-Scan had strong positive linearity (Pearson’s correlation coefficient, PCC = 0.86–0.97, PCC = 0.87–0.92; RMSE = 15.96 ± 9.49) and mean root mean squared error RMSE (9.17 ± 2.02) compared to the F-Scan on a progressive loading step test. The SI and F-Scan had comparable results for linearity and hysteresis on a sinusoidal loading test (PCC = 0.92–0.99; 5.04 ± 1.41; PCC = 0.94–0.99; 6.15 ± 1.39, respectively). SI had less mean RMSE (6.19 ± 1.38) than the F-Scan (8.66 ±2.31) on the sinusoidal test and less absolute error (4.08 ± 3.26) than the F-Scan (16.38 ± 12.43) on a static test. Both the SI and F-Scan had near-perfect between-day reliability interclass correlation coefficient, ICC = 0.97–1.00) to the F-Scan (ICC = 0.96–1.00). During running, the SI pressure output had a near-perfect linearity and low RMSE compared to the force measurement from the Bertec treadmill. However, the SI pressure output had a mean hysteresis of 7.67% with a 28.47% maximum hysteresis, which may have implications for the accurate quantification of kinetic gait measures during running.
format Online
Article
Text
id pubmed-9963688
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-99636882023-02-26 Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait Blades, Samuel Jensen, Matt Stellingwerff, Trent Hundza, Sandra Klimstra, Marc Sensors (Basel) Article This study characterized the absolute pressure measurement error and reliability of a new fully integrated (Kinetyx, SI) plantar-pressure measurement system (PPMS) versus an industry-standard PPMS (F-Scan, Tekscan) during an established benchtop testing protocol as well as via a research-grade, instrumented treadmill (Bertec) during a running protocol. Benchtop testing results showed that both SI and F-Scan had strong positive linearity (Pearson’s correlation coefficient, PCC = 0.86–0.97, PCC = 0.87–0.92; RMSE = 15.96 ± 9.49) and mean root mean squared error RMSE (9.17 ± 2.02) compared to the F-Scan on a progressive loading step test. The SI and F-Scan had comparable results for linearity and hysteresis on a sinusoidal loading test (PCC = 0.92–0.99; 5.04 ± 1.41; PCC = 0.94–0.99; 6.15 ± 1.39, respectively). SI had less mean RMSE (6.19 ± 1.38) than the F-Scan (8.66 ±2.31) on the sinusoidal test and less absolute error (4.08 ± 3.26) than the F-Scan (16.38 ± 12.43) on a static test. Both the SI and F-Scan had near-perfect between-day reliability interclass correlation coefficient, ICC = 0.97–1.00) to the F-Scan (ICC = 0.96–1.00). During running, the SI pressure output had a near-perfect linearity and low RMSE compared to the force measurement from the Bertec treadmill. However, the SI pressure output had a mean hysteresis of 7.67% with a 28.47% maximum hysteresis, which may have implications for the accurate quantification of kinetic gait measures during running. MDPI 2023-02-20 /pmc/articles/PMC9963688/ /pubmed/36850951 http://dx.doi.org/10.3390/s23042352 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Blades, Samuel
Jensen, Matt
Stellingwerff, Trent
Hundza, Sandra
Klimstra, Marc
Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait
title Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait
title_full Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait
title_fullStr Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait
title_full_unstemmed Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait
title_short Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait
title_sort characterization of the kinetyx si wireless pressure-measuring insole during benchtop testing and running gait
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9963688/
https://www.ncbi.nlm.nih.gov/pubmed/36850951
http://dx.doi.org/10.3390/s23042352
work_keys_str_mv AT bladessamuel characterizationofthekinetyxsiwirelesspressuremeasuringinsoleduringbenchtoptestingandrunninggait
AT jensenmatt characterizationofthekinetyxsiwirelesspressuremeasuringinsoleduringbenchtoptestingandrunninggait
AT stellingwerfftrent characterizationofthekinetyxsiwirelesspressuremeasuringinsoleduringbenchtoptestingandrunninggait
AT hundzasandra characterizationofthekinetyxsiwirelesspressuremeasuringinsoleduringbenchtoptestingandrunninggait
AT klimstramarc characterizationofthekinetyxsiwirelesspressuremeasuringinsoleduringbenchtoptestingandrunninggait