Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait
This study characterized the absolute pressure measurement error and reliability of a new fully integrated (Kinetyx, SI) plantar-pressure measurement system (PPMS) versus an industry-standard PPMS (F-Scan, Tekscan) during an established benchtop testing protocol as well as via a research-grade, inst...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9963688/ https://www.ncbi.nlm.nih.gov/pubmed/36850951 http://dx.doi.org/10.3390/s23042352 |
_version_ | 1784896315700805632 |
---|---|
author | Blades, Samuel Jensen, Matt Stellingwerff, Trent Hundza, Sandra Klimstra, Marc |
author_facet | Blades, Samuel Jensen, Matt Stellingwerff, Trent Hundza, Sandra Klimstra, Marc |
author_sort | Blades, Samuel |
collection | PubMed |
description | This study characterized the absolute pressure measurement error and reliability of a new fully integrated (Kinetyx, SI) plantar-pressure measurement system (PPMS) versus an industry-standard PPMS (F-Scan, Tekscan) during an established benchtop testing protocol as well as via a research-grade, instrumented treadmill (Bertec) during a running protocol. Benchtop testing results showed that both SI and F-Scan had strong positive linearity (Pearson’s correlation coefficient, PCC = 0.86–0.97, PCC = 0.87–0.92; RMSE = 15.96 ± 9.49) and mean root mean squared error RMSE (9.17 ± 2.02) compared to the F-Scan on a progressive loading step test. The SI and F-Scan had comparable results for linearity and hysteresis on a sinusoidal loading test (PCC = 0.92–0.99; 5.04 ± 1.41; PCC = 0.94–0.99; 6.15 ± 1.39, respectively). SI had less mean RMSE (6.19 ± 1.38) than the F-Scan (8.66 ±2.31) on the sinusoidal test and less absolute error (4.08 ± 3.26) than the F-Scan (16.38 ± 12.43) on a static test. Both the SI and F-Scan had near-perfect between-day reliability interclass correlation coefficient, ICC = 0.97–1.00) to the F-Scan (ICC = 0.96–1.00). During running, the SI pressure output had a near-perfect linearity and low RMSE compared to the force measurement from the Bertec treadmill. However, the SI pressure output had a mean hysteresis of 7.67% with a 28.47% maximum hysteresis, which may have implications for the accurate quantification of kinetic gait measures during running. |
format | Online Article Text |
id | pubmed-9963688 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-99636882023-02-26 Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait Blades, Samuel Jensen, Matt Stellingwerff, Trent Hundza, Sandra Klimstra, Marc Sensors (Basel) Article This study characterized the absolute pressure measurement error and reliability of a new fully integrated (Kinetyx, SI) plantar-pressure measurement system (PPMS) versus an industry-standard PPMS (F-Scan, Tekscan) during an established benchtop testing protocol as well as via a research-grade, instrumented treadmill (Bertec) during a running protocol. Benchtop testing results showed that both SI and F-Scan had strong positive linearity (Pearson’s correlation coefficient, PCC = 0.86–0.97, PCC = 0.87–0.92; RMSE = 15.96 ± 9.49) and mean root mean squared error RMSE (9.17 ± 2.02) compared to the F-Scan on a progressive loading step test. The SI and F-Scan had comparable results for linearity and hysteresis on a sinusoidal loading test (PCC = 0.92–0.99; 5.04 ± 1.41; PCC = 0.94–0.99; 6.15 ± 1.39, respectively). SI had less mean RMSE (6.19 ± 1.38) than the F-Scan (8.66 ±2.31) on the sinusoidal test and less absolute error (4.08 ± 3.26) than the F-Scan (16.38 ± 12.43) on a static test. Both the SI and F-Scan had near-perfect between-day reliability interclass correlation coefficient, ICC = 0.97–1.00) to the F-Scan (ICC = 0.96–1.00). During running, the SI pressure output had a near-perfect linearity and low RMSE compared to the force measurement from the Bertec treadmill. However, the SI pressure output had a mean hysteresis of 7.67% with a 28.47% maximum hysteresis, which may have implications for the accurate quantification of kinetic gait measures during running. MDPI 2023-02-20 /pmc/articles/PMC9963688/ /pubmed/36850951 http://dx.doi.org/10.3390/s23042352 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Blades, Samuel Jensen, Matt Stellingwerff, Trent Hundza, Sandra Klimstra, Marc Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait |
title | Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait |
title_full | Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait |
title_fullStr | Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait |
title_full_unstemmed | Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait |
title_short | Characterization of the Kinetyx SI Wireless Pressure-Measuring Insole during Benchtop Testing and Running Gait |
title_sort | characterization of the kinetyx si wireless pressure-measuring insole during benchtop testing and running gait |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9963688/ https://www.ncbi.nlm.nih.gov/pubmed/36850951 http://dx.doi.org/10.3390/s23042352 |
work_keys_str_mv | AT bladessamuel characterizationofthekinetyxsiwirelesspressuremeasuringinsoleduringbenchtoptestingandrunninggait AT jensenmatt characterizationofthekinetyxsiwirelesspressuremeasuringinsoleduringbenchtoptestingandrunninggait AT stellingwerfftrent characterizationofthekinetyxsiwirelesspressuremeasuringinsoleduringbenchtoptestingandrunninggait AT hundzasandra characterizationofthekinetyxsiwirelesspressuremeasuringinsoleduringbenchtoptestingandrunninggait AT klimstramarc characterizationofthekinetyxsiwirelesspressuremeasuringinsoleduringbenchtoptestingandrunninggait |