Analytical Performance Evaluation of the New GEM(®) Premier™ 5000 in Comparison to the Epoc(®) Blood Gas Analyzer in Horses

SIMPLE SUMMARY: Blood gas analyzers are stationary or hand-held machines used to measure blood gases and electrolytes in a blood sample taken from a patient. No machine has been specifically designed for use in horses, and therefore every machine needs to be validated before being used in equine pra...

Descripción completa

Detalles Bibliográficos
Autores principales: Sandersen, Charlotte, Dmitrovic, Petra, Dupont, Julien, Cesarini, Carla, Guyot, Hugues, Serteyn, Didier, Kirsch, Katharina
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9966012/
https://www.ncbi.nlm.nih.gov/pubmed/36851418
http://dx.doi.org/10.3390/vetsci10020114
_version_ 1784896910238154752
author Sandersen, Charlotte
Dmitrovic, Petra
Dupont, Julien
Cesarini, Carla
Guyot, Hugues
Serteyn, Didier
Kirsch, Katharina
author_facet Sandersen, Charlotte
Dmitrovic, Petra
Dupont, Julien
Cesarini, Carla
Guyot, Hugues
Serteyn, Didier
Kirsch, Katharina
author_sort Sandersen, Charlotte
collection PubMed
description SIMPLE SUMMARY: Blood gas analyzers are stationary or hand-held machines used to measure blood gases and electrolytes in a blood sample taken from a patient. No machine has been specifically designed for use in horses, and therefore every machine needs to be validated before being used in equine practice. The aim of this study was to compare the newly marketed GEM5000 machine to the formerly validated epoc machine for blood gas analysis in horses. Blood samples taken from healthy and sick horses were run on both machines in alternate order, and various statistical tests were used to analyze if the two machines give similar results. Although the precision of the GEM5000 is good for most parameters, the agreement with the epoc machine is not always satisfactory. Therefore, data from different blood gas machines should not be used interchangeably. ABSTRACT: Different blood gas analyzers are used in equine practice. Every machine needs to be validated, as they have not been designed for use in horses. The aim of this study was to compare the newly marketed GEM5000 machine to the formerly validated epoc machine for blood gas analysis in horses. In this prospective, non-blinded, comparative laboratory analyzer study, 43 equine blood samples were analyzed on both analyzers and values were compared between the two machines via Lin’s concordance analysis, Passing–Bablok regression analysis and Bland–Altman plots. Duplicate measurements were conducted on the GEM5000 machine to evaluate precision. The GEM5000 failed to achieve the required precision for tHb, Hct and iCa(2+), but achieved acceptable precision for all other parameters. Concordance correlation analysis revealed poor correlation for Na(+), Cl(−), iCa(2+), K(+), Hct and tHb, while there was an at least moderate agreement for all other parameters. Passing–Bablok regression revealed significant constant bias for pCO(2), pO(2), Cl(−), and iCa(2+) and significant proportional bias for pCO(2), iCa(2+) and SO(2). Bland–Altman analysis revealed significant systematic bias for Na(+), Cl(−), iCa(2+), K(+), Hct, tHb and SO(2). This study shows that while precision of the GEM5000 is good, values should not be used interchangeably with data obtained from other blood gas analyzers.
format Online
Article
Text
id pubmed-9966012
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-99660122023-02-26 Analytical Performance Evaluation of the New GEM(®) Premier™ 5000 in Comparison to the Epoc(®) Blood Gas Analyzer in Horses Sandersen, Charlotte Dmitrovic, Petra Dupont, Julien Cesarini, Carla Guyot, Hugues Serteyn, Didier Kirsch, Katharina Vet Sci Article SIMPLE SUMMARY: Blood gas analyzers are stationary or hand-held machines used to measure blood gases and electrolytes in a blood sample taken from a patient. No machine has been specifically designed for use in horses, and therefore every machine needs to be validated before being used in equine practice. The aim of this study was to compare the newly marketed GEM5000 machine to the formerly validated epoc machine for blood gas analysis in horses. Blood samples taken from healthy and sick horses were run on both machines in alternate order, and various statistical tests were used to analyze if the two machines give similar results. Although the precision of the GEM5000 is good for most parameters, the agreement with the epoc machine is not always satisfactory. Therefore, data from different blood gas machines should not be used interchangeably. ABSTRACT: Different blood gas analyzers are used in equine practice. Every machine needs to be validated, as they have not been designed for use in horses. The aim of this study was to compare the newly marketed GEM5000 machine to the formerly validated epoc machine for blood gas analysis in horses. In this prospective, non-blinded, comparative laboratory analyzer study, 43 equine blood samples were analyzed on both analyzers and values were compared between the two machines via Lin’s concordance analysis, Passing–Bablok regression analysis and Bland–Altman plots. Duplicate measurements were conducted on the GEM5000 machine to evaluate precision. The GEM5000 failed to achieve the required precision for tHb, Hct and iCa(2+), but achieved acceptable precision for all other parameters. Concordance correlation analysis revealed poor correlation for Na(+), Cl(−), iCa(2+), K(+), Hct and tHb, while there was an at least moderate agreement for all other parameters. Passing–Bablok regression revealed significant constant bias for pCO(2), pO(2), Cl(−), and iCa(2+) and significant proportional bias for pCO(2), iCa(2+) and SO(2). Bland–Altman analysis revealed significant systematic bias for Na(+), Cl(−), iCa(2+), K(+), Hct, tHb and SO(2). This study shows that while precision of the GEM5000 is good, values should not be used interchangeably with data obtained from other blood gas analyzers. MDPI 2023-02-03 /pmc/articles/PMC9966012/ /pubmed/36851418 http://dx.doi.org/10.3390/vetsci10020114 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Sandersen, Charlotte
Dmitrovic, Petra
Dupont, Julien
Cesarini, Carla
Guyot, Hugues
Serteyn, Didier
Kirsch, Katharina
Analytical Performance Evaluation of the New GEM(®) Premier™ 5000 in Comparison to the Epoc(®) Blood Gas Analyzer in Horses
title Analytical Performance Evaluation of the New GEM(®) Premier™ 5000 in Comparison to the Epoc(®) Blood Gas Analyzer in Horses
title_full Analytical Performance Evaluation of the New GEM(®) Premier™ 5000 in Comparison to the Epoc(®) Blood Gas Analyzer in Horses
title_fullStr Analytical Performance Evaluation of the New GEM(®) Premier™ 5000 in Comparison to the Epoc(®) Blood Gas Analyzer in Horses
title_full_unstemmed Analytical Performance Evaluation of the New GEM(®) Premier™ 5000 in Comparison to the Epoc(®) Blood Gas Analyzer in Horses
title_short Analytical Performance Evaluation of the New GEM(®) Premier™ 5000 in Comparison to the Epoc(®) Blood Gas Analyzer in Horses
title_sort analytical performance evaluation of the new gem(®) premier™ 5000 in comparison to the epoc(®) blood gas analyzer in horses
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9966012/
https://www.ncbi.nlm.nih.gov/pubmed/36851418
http://dx.doi.org/10.3390/vetsci10020114
work_keys_str_mv AT sandersencharlotte analyticalperformanceevaluationofthenewgempremier5000incomparisontotheepocbloodgasanalyzerinhorses
AT dmitrovicpetra analyticalperformanceevaluationofthenewgempremier5000incomparisontotheepocbloodgasanalyzerinhorses
AT dupontjulien analyticalperformanceevaluationofthenewgempremier5000incomparisontotheepocbloodgasanalyzerinhorses
AT cesarinicarla analyticalperformanceevaluationofthenewgempremier5000incomparisontotheepocbloodgasanalyzerinhorses
AT guyothugues analyticalperformanceevaluationofthenewgempremier5000incomparisontotheepocbloodgasanalyzerinhorses
AT serteyndidier analyticalperformanceevaluationofthenewgempremier5000incomparisontotheepocbloodgasanalyzerinhorses
AT kirschkatharina analyticalperformanceevaluationofthenewgempremier5000incomparisontotheepocbloodgasanalyzerinhorses