Cargando…
The eRDS v6 Stereotest and the Vivid Vision Stereo Test: Two New Tests of Stereoscopic Vision
PURPOSE: To describe two new stereoacuity tests: the eRDS v6 stereotest, a global dynamic random dot stereogram (dRDS) test, and the Vivid Vision Stereo Test version 2 (VV), a local or “contour” stereotest for virtual reality (VR) headsets; and to evaluate the tests’ reliability, validity compared t...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Association for Research in Vision and Ophthalmology
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9987163/ https://www.ncbi.nlm.nih.gov/pubmed/36857068 http://dx.doi.org/10.1167/tvst.12.3.1 |
_version_ | 1784901324005965824 |
---|---|
author | Denkinger, Sylvie Antoniou, Maria-Paraskevi Tarello, Demetrio Levi, Dennis M. Backus, Benjamin T. Bavelier, Daphné Chopin, Adrien |
author_facet | Denkinger, Sylvie Antoniou, Maria-Paraskevi Tarello, Demetrio Levi, Dennis M. Backus, Benjamin T. Bavelier, Daphné Chopin, Adrien |
author_sort | Denkinger, Sylvie |
collection | PubMed |
description | PURPOSE: To describe two new stereoacuity tests: the eRDS v6 stereotest, a global dynamic random dot stereogram (dRDS) test, and the Vivid Vision Stereo Test version 2 (VV), a local or “contour” stereotest for virtual reality (VR) headsets; and to evaluate the tests’ reliability, validity compared to a dRDS standard, and learning effects. METHODS: Sixty-four subjects passed a battery of stereotests, including perceiving depth from RDS. Validity was evaluated relative to a tablet-based dRDS reference test, ASTEROID. Reliability and learning effects were assessed over six sessions. RESULTS: eRDS v6 was effective at measuring small thresholds (<10 arcsec) and had a moderate correlation (0.48) with ASTEROID. Across the six sessions, test-retest reliability was good, varying from 0.84 to 0.91, but learning occurred across the first three sessions. VV did not measure stereoacuities below 15 arcsec. It had a weak correlation with ASTEROID (0.27), and test-retest reliability was poor to moderate, varying from 0.35 to 0.74; however, no learning occurred between sessions. CONCLUSIONS: eRDS v6 is precise and reliable but shows learning effects. If repeated three times at baseline, this test is well suited as an outcome measure for testing interventions. VV is less precise, but it is easy and rapid and shows no learning. It may be useful for testing interventions in patients who have no global stereopsis. TRANSLATIONAL RELEVANCE: eRDS v6 is well suited as an outcome measure to evaluate treatments that improve adult stereodepth perception. VV can be considered for screening patient with compromised stereovision. |
format | Online Article Text |
id | pubmed-9987163 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | The Association for Research in Vision and Ophthalmology |
record_format | MEDLINE/PubMed |
spelling | pubmed-99871632023-03-07 The eRDS v6 Stereotest and the Vivid Vision Stereo Test: Two New Tests of Stereoscopic Vision Denkinger, Sylvie Antoniou, Maria-Paraskevi Tarello, Demetrio Levi, Dennis M. Backus, Benjamin T. Bavelier, Daphné Chopin, Adrien Transl Vis Sci Technol Pediatric Ophthalmology & Strabismus PURPOSE: To describe two new stereoacuity tests: the eRDS v6 stereotest, a global dynamic random dot stereogram (dRDS) test, and the Vivid Vision Stereo Test version 2 (VV), a local or “contour” stereotest for virtual reality (VR) headsets; and to evaluate the tests’ reliability, validity compared to a dRDS standard, and learning effects. METHODS: Sixty-four subjects passed a battery of stereotests, including perceiving depth from RDS. Validity was evaluated relative to a tablet-based dRDS reference test, ASTEROID. Reliability and learning effects were assessed over six sessions. RESULTS: eRDS v6 was effective at measuring small thresholds (<10 arcsec) and had a moderate correlation (0.48) with ASTEROID. Across the six sessions, test-retest reliability was good, varying from 0.84 to 0.91, but learning occurred across the first three sessions. VV did not measure stereoacuities below 15 arcsec. It had a weak correlation with ASTEROID (0.27), and test-retest reliability was poor to moderate, varying from 0.35 to 0.74; however, no learning occurred between sessions. CONCLUSIONS: eRDS v6 is precise and reliable but shows learning effects. If repeated three times at baseline, this test is well suited as an outcome measure for testing interventions. VV is less precise, but it is easy and rapid and shows no learning. It may be useful for testing interventions in patients who have no global stereopsis. TRANSLATIONAL RELEVANCE: eRDS v6 is well suited as an outcome measure to evaluate treatments that improve adult stereodepth perception. VV can be considered for screening patient with compromised stereovision. The Association for Research in Vision and Ophthalmology 2023-03-01 /pmc/articles/PMC9987163/ /pubmed/36857068 http://dx.doi.org/10.1167/tvst.12.3.1 Text en Copyright 2023 The Authors https://creativecommons.org/licenses/by-nc-nd/4.0/This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License. |
spellingShingle | Pediatric Ophthalmology & Strabismus Denkinger, Sylvie Antoniou, Maria-Paraskevi Tarello, Demetrio Levi, Dennis M. Backus, Benjamin T. Bavelier, Daphné Chopin, Adrien The eRDS v6 Stereotest and the Vivid Vision Stereo Test: Two New Tests of Stereoscopic Vision |
title | The eRDS v6 Stereotest and the Vivid Vision Stereo Test: Two New Tests of Stereoscopic Vision |
title_full | The eRDS v6 Stereotest and the Vivid Vision Stereo Test: Two New Tests of Stereoscopic Vision |
title_fullStr | The eRDS v6 Stereotest and the Vivid Vision Stereo Test: Two New Tests of Stereoscopic Vision |
title_full_unstemmed | The eRDS v6 Stereotest and the Vivid Vision Stereo Test: Two New Tests of Stereoscopic Vision |
title_short | The eRDS v6 Stereotest and the Vivid Vision Stereo Test: Two New Tests of Stereoscopic Vision |
title_sort | erds v6 stereotest and the vivid vision stereo test: two new tests of stereoscopic vision |
topic | Pediatric Ophthalmology & Strabismus |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9987163/ https://www.ncbi.nlm.nih.gov/pubmed/36857068 http://dx.doi.org/10.1167/tvst.12.3.1 |
work_keys_str_mv | AT denkingersylvie theerdsv6stereotestandthevividvisionstereotesttwonewtestsofstereoscopicvision AT antonioumariaparaskevi theerdsv6stereotestandthevividvisionstereotesttwonewtestsofstereoscopicvision AT tarellodemetrio theerdsv6stereotestandthevividvisionstereotesttwonewtestsofstereoscopicvision AT levidennism theerdsv6stereotestandthevividvisionstereotesttwonewtestsofstereoscopicvision AT backusbenjamint theerdsv6stereotestandthevividvisionstereotesttwonewtestsofstereoscopicvision AT bavelierdaphne theerdsv6stereotestandthevividvisionstereotesttwonewtestsofstereoscopicvision AT chopinadrien theerdsv6stereotestandthevividvisionstereotesttwonewtestsofstereoscopicvision AT denkingersylvie erdsv6stereotestandthevividvisionstereotesttwonewtestsofstereoscopicvision AT antonioumariaparaskevi erdsv6stereotestandthevividvisionstereotesttwonewtestsofstereoscopicvision AT tarellodemetrio erdsv6stereotestandthevividvisionstereotesttwonewtestsofstereoscopicvision AT levidennism erdsv6stereotestandthevividvisionstereotesttwonewtestsofstereoscopicvision AT backusbenjamint erdsv6stereotestandthevividvisionstereotesttwonewtestsofstereoscopicvision AT bavelierdaphne erdsv6stereotestandthevividvisionstereotesttwonewtestsofstereoscopicvision AT chopinadrien erdsv6stereotestandthevividvisionstereotesttwonewtestsofstereoscopicvision |