Cargando…

An Accurate Isotope Dilution Liquid Chromatography-Tandem Mass Spectrometry Method for Serum C-Peptide and Its Use in Harmonization in China

BACKGROUND: Serum C-peptide results from various routine methods used in China are highly variable, warranting well-performing methods to serve as an accuracy base to improve the harmonization of C-peptide measurements in China. We developed an accurate isotope dilution liquid chromatography-tandem...

Descripción completa

Detalles Bibliográficos
Autores principales: Deng, Yuhang, Zhang, Chao, Wang, Jing, Zeng, Jie, Zhang, Jiangtao, Zhang, Tianjiao, Zhao, Haijian, Zhou, Weiyan, Zhang, Chuanbao
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Society for Laboratory Medicine 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9989529/
https://www.ncbi.nlm.nih.gov/pubmed/36843403
http://dx.doi.org/10.3343/alm.2023.43.4.345
Descripción
Sumario:BACKGROUND: Serum C-peptide results from various routine methods used in China are highly variable, warranting well-performing methods to serve as an accuracy base to improve the harmonization of C-peptide measurements in China. We developed an accurate isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC–MS/MS) method for serum C-peptide measurement and explored its use in harmonization. METHODS: After protein precipitation with Z(n)SO(4) solution, C-peptide was extracted from serum samples by anion-exchange solid-phase extraction and quantified by ID-LC–MS/MS in positive ion mode. The precision and analytical recovery of the ID-LC–MS/MS method were assessed. Seventy-six serum samples were analyzed using the ID-LC–MS/MS method and six routine immunoassays. Ordinary linear regression (OLR) and Bland-Altman (BA) analyses were conducted to evaluate the relationship between the ID-LC–MS/MS method and routine immunoassays. Five serum pool samples assigned using the ID-LC–MS/MS method were used to recalibrate the routine assays. OLR and BA analyses were re-conducted after recalibration. RESULTS: The within-run, between-run, and total precision for the ID-LC–MS/MS method at four concentrations were 1.0%–2.1%, 0.6%–1.2%, and 1.3%–2.2%, respectively. The analytical recoveries for the ID-LC–MS/MS method at three concentrations were 100.3%–100.7%, 100.4%–101.0%, and 99.6%–100.7%. The developed method and the immunoassays were strongly correlated, with all R(2) >0.98. The comparability among the immunoassays was substantially improved after recalibration. CONCLUSIONS: The performance of the ID-LC–MS/MS method was carefully validated, and this method can be used to improve the harmonization of serum C-peptide measurements in China.